Page 1
SLS1170 : 8 Litre Lab Pro Stirred Bath SLS1172: 14 Litre Lab Pro Stirred Bath SLS1174: 22 Litre Lab Pro Stirred Bath...
Page 2
Dear Customer Thank you for purchasing this piece of Lab Pro temperature control equipment. To get the best performance from your equipment and for your own safety please read these instructions carefully before use. GENERAL NOTES This product is designed for laboratory use only. Always follow good laboratory practice.
Page 3
LOCATION The product must be placed on a smooth, level and sturdy work surface. Use in a ventilated room. Suitable for use in ambient temperatures 5°C to 40°C with a maximum humidity 80% (temperature 31°C) decreasing to 50% (temperature 40°C). DO NOT block or restrict ventilation slots.
Page 4
POWER INTERRUPT DURING TIMER MODE If the power is interrupted during the timer mode, the display shows “P.OFF” when resumed. To clear press and hold the RUN key until display reverts to actual temperature. Timer mode will then continue. To deactivate the timer press and hold the RUN key. OVER TEMPERATURE ALARM The over temperature alarm is automatically set 4˚C above set temperature.
Page 5
OPERATING INSTRUCTIONS Switch the unit on using the mains switch located on the rear. The switch will illuminate and the controller will perform a self test. When the unit is switched off all time and temperature values remain in memory. SETTING TEMPERATURE Press and hold the DOWN arrow, or the FUNCTION key to display SP1: the set temperature.
Page 6
CARE AND MAINTENANCE Please ensure that the washing agent and sanitizing agent are BSI accredited and approved by the H&S department for use on laboratory equipment and stainless steel within your laboratory. DISCONNECT THE BATH FROM THE POWER SUPPLY PRIOR TO CLEANING BASIC CLEANING The water bath should be cleaned at regular intervals by wiping external surfaces with...
Page 7
EC Declaration of Conformity We herewith confirm the following product, designed and manufactured in the UK: LAB PRO RANGE STIRRED WATER BATHS Conforms with the requirements outlined by the following European Directives: Low Voltage Directive 2014/35/EU EMC Directive 2014/30/EU RoHS Directive 2011/65/EU Conforms with the requirements of the following standards: BS EN 61010-1: 2010 BS EN 61010-2-010: 2014...
Page 8
FINAL INSPECTION AND ELECTRICAL SAFETY TEST REPORT BX1084 Issue 2: March 2019...
Need help?
Do you have a question about the Lab Pro SLS1170 and is the answer not in the manual?
Questions and answers