Priming The Vacuum Pump; Priming The Vacuum Oven; Setting Of Temperature; Start / Stop Operation - Holland Green Science Hephaestus V094 User Manual

Vacuum ovens
Table of Contents

Advertisement

7.3 Priming the Vacuum Pump

Priming of the vacuum pump requires that the connecting tube provided is securely
(airtight sealed) placed between the vacuum pump and the vacuum oven.
The oven door should be securely shut and the door handle secured in a closed position
(this will prevent air from entering the chamber by causing the hole in the rubber plug
and the hole in the pipe core to be 180 ° out of phase).
Power to the vacuum pump should be turned on and then the vacuum valve should be
closed by turning the handle clockwise 90°during processing.
To stop the process, reduce the vacuum in the chamber -0.1Mpa before turning off the
vacuum pump. This will ensure that no engine oil from the vacuum pump will be drawn
into the chamber.

7.4 Priming the Vacuum Oven

Turn on power to the vacuum oven (a power-on indicator light will confirm this) and after
a 5 second self-test is completed, the temperature controller will automatically enter the
"run" mode. This will allow the PV screen to display the process temperature and the SV
screen display the set temperature.
Once the process begins the AT indicator light will turn on and if process temperature is
less than the set temperature heating lamps will turn on.

7.5 Setting of Temperature

When the instrument is switched on, press the
temperature value" will start to blink. Set the desired temperature value (in Celsius
degrees) by pressing keys.
Use the
button to move between the digits.
Confirm the set value by pressing the
Note:
Pressing the '
Press and hold the keys changes the value rapidly.

7.6 Start / Stop Operation

After setting the operating parameters, press
heating process. When the word "STOP" appears on the upper right corner of the
screen, the system is in the stopped state. When the message "RUN" appear in the
Version 1.0
' key once changes the value by 0.1.
or
User Manual
button one time, and the "set
button.
button for 4-5 seconds to start the
17