Operation; Memorizing The Memory Positions - Hausmann 1396 Operation And Service Manual

Table of Contents

Advertisement

OPERATION

(See hand control switch at top of page 7.)
Unit is pre-wired and pre-tested at factory ready for use. Set the platform in place and plug
power cord into 110V wall outlet.
IMPORTANT
Follow these instructions exactly upon initial use, as follows:
Press
"Up" Button and elevate both handrails approximately 2" (not more) - the LED
display will read-out the height in inches.
Press
"Down" Button and lower both handrails until they are at lowest height
(bottomed out). The unit will stop automatically at lowest level.
Press
"Down" Button again for 5 seconds to make sure all (4) columns have
bottomed out to their lowest height.
Press
"Up" Button and elevate hand rails to maximum highest height of 44".
Press
"Up" Button again for a second or two to make sure all (4) lift motors are at
maximum height.
Press
"Down" Button to lower to normal height level.
YOU ARE NOW READY TO OPERATE FOR NORMAL USE.

MEMORIZING THE MEMORY POSITIONS

As the height adjustment is variable and random from patient to patient, we normally do not
recommend use of memory control, but it is an option if needed.
With the memory capacity, you can specify a given handrail height. Up to four different heights
can be memorized. These memorized positions can be later re-specified.
S
Press
. An S appears in the LED display.
Press one of the position keys
After the position key has been pressed, the number of the position key appears in the LED
display next to S. After the memorization procedure, a double-click is heard, and after
approximately 2 seconds the current height appears again in the LED display.
2004 Hausmann Industries, Inc.
11/3/04
1
2
3
,
,
4
or
.
3

Advertisement

Table of Contents
loading

Related Products for Hausmann 1396

This manual is also suitable for:

1398

Table of Contents