Bair Hugger Model 750 Temperature Management Unit Operator's Manual
Electrical Characteristics
heating element
leakage current
blower motor
power consumption
power cord
device ratings
fuses
certifications
classification
diagnostics
1400 W Resistive
Meets UL 60601-1 and IEC 60601-1 requirements�
Operating speed: approximately 4,000 rpm
Airflow: up to 48 cfm or 23 L/s
Peak: 1550 W
Average: 800 W
15 ft�, SJT, 3 cond�, 13 A
15 ft�, SJT, 3 cond�, 15 A
4�6 m, HAR, 3 cond�, 10 A
110-120 VAC, 50/60 Hz, 11�7 A, or
220-240 VAC, 50/60 Hz, 7�2 A, or
100 VAC, 50/60 Hz, 15 A
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
IEC 60601-1; EN 60601-1-2; UL 60601-1;
CAN/CSA-C22�2, No� 601�1, EN 55011;
ASTM F2196-02
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type
BF, Ordinary equipment, Continuous operation�
Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous
oxide� Classified by Underwriters Laboratories Inc�
with respect to electric shock, fire, and mechanical
hazards only, in accordance with UL 60601-1, ASTM
F2196-02 and in accordance with Canadian/CSA
C22�2, No� 601�1� Classified under the Medical
Device as a Class IIb device�
A qualified service technician can perform over-
temperature detection system testing, temperature
output testing, operating temperature calibration, and
fault code troubleshooting�
19