13. TROUBLESHOOTING..............................9 14. CLEANING................................10 15. IMPORTANT SAFETY PRECAUTIONS..........................10 16. MATERIALS DISPOSAL............................10 17. EU / DECLARATION OF CONFORMITY ........................10 18. EMFIT LIMITED WARRANTY STATEMENT ........................11 19. TECHNICAL SPECIFICATIONS ..........................12 20. ELECTROMAGNETIC CONDITIONS..........................13 21. MANUFACTURER´S CONTACT INFORMATION......................15 DANGER NOTES THE FOLLOWING NOTES ARE PROVIDED BOTH FOR YOUR PERSONAL SAFETY AND TO PROTECT THE DESCRIBED PRODUCT OR ATTACHED DEVICES FROM DAMAGE.
Intended Use The Emfit Nocturnal Epileptic Tonic-Clonic Seizure Monitor (device model D-1090-2G and sensor model L-4060SL) is intended to be used for monitoring a person with epilepsy in a bed or on a chair and notifying caregiver of an over preset continued tonic- clonic seizure, with muscle jerkings.
Table of DIP Switch Settings Switch # OFF (down) ON (up) Movement continuation time delay setting, see next Movement continuation time delay setting, see next chapter after volume settings chapter after volume settings Movement continuation time delay setting, see next Movement continuation time delay setting, see next chapter after volume settings chapter after volume settings...
THE SENSOR MUST ALWAYS BE PLACED BENEATH A MATTRESS OR MATTRESS PAD. SENSOR MUST NOT BE IN DIRECT CONTACT WITH A PERSON! THE EMFIT BED SENSOR IS DESIGNED TO WITHSTAND USE BELOW THE MATTRESS, AGAINST A HARD BED BOTTOM FOR 4 – 5 YEARS OF USE. MANUFACTURER RECOMMENDS REPLACING IT AFTER 6 YEARS FROM MANUFACTURING DATE OR AFTER 5 YEARS OF USE, WHICHEVER COMES FIRST.
Connect the under mattress sensor, optional AC adapter and possible connection wire to other system as follows: Connect the sensor wire plug to If you need to connect the dry- Connect the optional AC adapter the control unit X3 marked contact output to another system, cord to the X1 marked connector connector.
4 pcs sure it stays in place THE EMFIT EPILEPTIC TONIC-CLONIC SEIZURE ALARM IS DESIGNED AND TESTED TO BE USED ONLY WITH GLOBTEK INC. POWER SUPPLY MODEL: NO GTM41060-1505 AND P/N:WR9QA3000LCP-N-MNK. ANY OTHER TYPE OF AC ADAPTER MAY AFFECT THE PRODUCT’S SAFETY.
10. LED-LIGHT INDICATORS 1. Green LED / Presence - The green light starts blinking slowly (half the speed of blue light) when the device notices micro or other movements (person goes to bed) - The green light starts blinking same speed with blue light when the device 1.
Tests When Taking into Use and Every Month Do the following tests at least once a month and when putting into use, to ensure the system’s proper operation! Testing the “no movements” notification Check to see if system is turned on. If disabled, set on the “no movement” notification function (DIP switch 5 at OFF position).
17. EU / DECLARATION OF CONFORMITY Emfit Epileptic Seizure Alarm (device model D-1090-2G and bed sensor model L-4060SL) complies with the essential requirements of EMC directive 2004/108/EC, CE mark directive 93/68/EEC and Medical Device Directive...
- treatment resulting in physical, cosmetic or surface damage or changes to the product - failure to install or use the product for its normal purpose or in accordance with Emfit’s instructions on installation or use - failure to maintain the product in accordance with Emfit’s instructions on proper maintenance...
You may assert any rights you have at your sole discretion. 19. TECHNICAL SPECIFICATIONS Control Unit Model: D-1090-2G Operating voltage: 5 V DC Input / output connectors: Power, AUX and Sensor Dry-contact outputs: Max.
Note! RF communications equipment can effect medical electrical equipment! Electromagnetic emissions The Emfit Emfit Nocturnal Epileptic Tonic-Clonic Seizure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment.
Page 14
Portable and mobile RF communications equipment should be used no closer to any part of the The Emfit Nocturnal Epileptic Tonic-Clonic Seizure Monitor , including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
EQUIPMENT SYSTEM The Emfit Nocturnal Epileptic Tonic-Clonic Seizure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Emfit epileptic seizure alarm can help prevent electromagnetic interference...
Need help?
Do you have a question about the D-1090-2G and is the answer not in the manual?
Questions and answers