Manual Calibration - KC N901 User Manual

Smart helmet
Table of Contents

Advertisement

temperature.
Step 1: Open
helmet through bluetooth;
Step 2: The calibration assisting person stands 1 meter in front
of the thermal imager.
Step 3: Adjust the center of the screen to target the ears or
forehead of the assisting person, and click
Calibration
"
Calibration is complete
Step 4: Click "Save" to save the Calibration parameter

4.4. Manual Calibration

Note:
⚫ he thermal imager can be parameterized manually to
obtain higher measurement accuracy;
⚫ Temperature measurement can be started 5~10 minutes
after boot when the measurement accuracy stabilizes.
⚫ The calibration parameter has a great influence on the
measurement precision, so the calibration parameter
should be reset when the precision has a large error during
the use;
⚫ This work should be carried out with the assistance of a
person of normal body temperature.
Step 1: Let the assisting staff stand 1 meter from the thermal
imager and measure his temperature of forehead with the
thermal imager;
Step 2: If the temperature reading is within 36.5~36.7
calibration
Otherwise if the reading is out of the range, the calibration
parameter should be adjusted until the measurement error
meets the requirements. The temperature reading will rise as
increasing the calibration parameter and vice versa. The
calibration parameter should be within the range of -3~+3.
Step 3: Let the assisting staff stand 0.5 meter from the thermal
"
KC Smart Helmet
"
. The quick calibration is completed as
"
parameter
needs
13
"
App and connect the
prompts on the screen.
no
further
"
One-Click
, the
modification.