Rosenberger 74CK011-170 User Manual

Rosenberger 74CK011-170 User Manual

Calibration kit

Advertisement

Quick Links

User Manual
F
Calibration Kit
75 Ω
Industrial Version
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260
D-84526 Tittmoning
www.rosenberger.com
74CK011-170 (Industrial Version)
Tel.
Germany
Email : info@rosenberger.com
74CK011-170
: +49 8684 18-0
Page
1 / 15

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 74CK011-170 and is the answer not in the manual?

Questions and answers

Summary of Contents for Rosenberger 74CK011-170

  • Page 1 User Manual Calibration Kit 74CK011-170 75 Ω Industrial Version 74CK011-170 (Industrial Version) Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com Email : info@rosenberger.com 1 / 15...
  • Page 2: Table Of Contents

    Extent of Rosenberger Re-Calibration Services 6. Use, Maintenance and Care Connector Care Interface Dimensions and Gauging Mechanical Specifications Handling and Storage 7. Connections Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com...
  • Page 3: General Information

    VNAs usually offer several techniques for correcting systematic measurement errors. The standards in this calibration kit allow you to perform one-port OSL (OSM) or two-port OSLT (TOSM) and TRM (TMR) calibrations from DC up to 6 GHz for measurements of Rosenberger components with F 75 Ω connectors.
  • Page 4: Kit Contents

    User Manual Calibration Kit 74CK011-170 75 Ω Industrial Version 1.3 Kit Contents Table 1-1 Parts of the 74CK011-170 Calibration Kit Device Part number Remarks Quantity Open circuit plug 74S12L-001S3 74K12L-001S3 Open circuit jack Short circuit plug 74S12S-001S3 Short circuit jack...
  • Page 5: Kit Documentation

    Each calibration kit comes with a calibration certificate. Measurement results and uncertainties for all calibration standards are included. Standard scope of delivery is the Factory Calibration. Accredited Calibration is not available. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 6: Specifications

    These standard definitions are used in the calibration of the VNA to determine the systematic errors of the measurement system. Please always use the most recent source of information (Standard Definitions Cards or VNA specific files). Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 7: Installation Of Standard Definitions

    PNA family and compatible ones ‘calkit’ file can be used on Rohde&Schwarz VNAs ZVA, ZVB, ZVBT, ZNB families and compatible ones Example file name: 74CK011-170 (BH003)_mb_2020-05-30.calkit Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 8: Vna Calibration

    Note: This Calibration Kit does not include in-series calibration adapters. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 9: Rosenberger Hochfrequenztechnik Gmbh & Co. Kg Tel. : +49 8684

    This should be done from time to time to identify defects. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 10 5. Calibration Kit Service 5.1 Calibration interval Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the wear and the actual requirements a different calibration interval might be suitable. This lies in the responsibility of the user.
  • Page 11 Small particles, usually of metal, are often found on the inside connector mating planes and on the threads. They should be removed to prevent damage to the connector surfaces. The items required for cleaning connectors and the procedure to be followed is described below. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 12 6.2.1 Gauging connectors A connector should be gauged before it is used for the very first time or if someone else has used the device to which it is connected. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 13 This will result in poor reflection and possibly unreliable contact. 6.2.2 Gauging There are no special Rosenberger gauges available for F 75 Ω connectors. It is recommended to use a Caliper (e.g. Mitutoyo ID-F150H) Checking devices for protrusion or excessive recession only a single measurement is sufficient.
  • Page 14 Hold the handle in the middle and turn it until the torque wrench “breaks”. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
  • Page 15 Rev. Engineering change number Name Date Herbert Babinger 24/04/12 Markus Hantschel 21/12/20 20-1058 Lars Ramtke 21/12/20 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com Email : info@rosenberger.com 15 / 15...