VNAs usually offer several techniques for correcting systematic measurement errors. The standards in this calibration kit allow you to perform one-port OSL (OSM) or two-port OSLT (TOSM) and TRM (TMR) calibrations from DC up to 6 GHz for measurements of Rosenberger components with F 75 Ω connectors.
User Manual Calibration Kit 74CK011-170 75 Ω Industrial Version 1.3 Kit Contents Table 1-1 Parts of the 74CK011-170 Calibration Kit Device Part number Remarks Quantity Open circuit plug 74S12L-001S3 74K12L-001S3 Open circuit jack Short circuit plug 74S12S-001S3 Short circuit jack...
Each calibration kit comes with a calibration certificate. Measurement results and uncertainties for all calibration standards are included. Standard scope of delivery is the Factory Calibration. Accredited Calibration is not available. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
These standard definitions are used in the calibration of the VNA to determine the systematic errors of the measurement system. Please always use the most recent source of information (Standard Definitions Cards or VNA specific files). Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
PNA family and compatible ones ‘calkit’ file can be used on Rohde&Schwarz VNAs ZVA, ZVB, ZVBT, ZNB families and compatible ones Example file name: 74CK011-170 (BH003)_mb_2020-05-30.calkit Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Note: This Calibration Kit does not include in-series calibration adapters. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
This should be done from time to time to identify defects. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 10
5. Calibration Kit Service 5.1 Calibration interval Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the wear and the actual requirements a different calibration interval might be suitable. This lies in the responsibility of the user.
Page 11
Small particles, usually of metal, are often found on the inside connector mating planes and on the threads. They should be removed to prevent damage to the connector surfaces. The items required for cleaning connectors and the procedure to be followed is described below. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 12
6.2.1 Gauging connectors A connector should be gauged before it is used for the very first time or if someone else has used the device to which it is connected. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 13
This will result in poor reflection and possibly unreliable contact. 6.2.2 Gauging There are no special Rosenberger gauges available for F 75 Ω connectors. It is recommended to use a Caliper (e.g. Mitutoyo ID-F150H) Checking devices for protrusion or excessive recession only a single measurement is sufficient.
Page 14
Hold the handle in the middle and turn it until the torque wrench “breaks”. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 15
Rev. Engineering change number Name Date Herbert Babinger 24/04/12 Markus Hantschel 21/12/20 20-1058 Lars Ramtke 21/12/20 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com Email : info@rosenberger.com 15 / 15...