N 75 Ω before connecting them to any calibration standard of the mentioned Calibration Kits or a test port of a Vector Network Analyzers (VNA). We recommend to use Rosenberger Gauge Kit P5GK0KS-010 or comparable ones. For customer specific Calibration Kits all VNA calibration methods may not be available.
Never connect RPC-N 75 Ω connectors to RPC-N 50 Ω or other type N 50 Ω connectors. Specially the RPC-N 75 Ω female center conductor will be be destroyed by an 50 Ω male center conductor. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
User Manual RPC-N 75 Ω Calibration Kit P5CK14A-170 P5CK14A-170 and P5CK010-170 1.3 Kit Contents Table 1-2 Parts of the P5CK14A-170 Calibration Kit Device Part number Remarks Quantity Open circuit plug P5S12L-00AS3 Open circuit jack P5K12L-00AS3 Short circuit plug P5S12S-00AS3 Short circuit jack...
Optional scope of delivery is the Accredited Calibration. In this case the calibration kit is calibrated in the Rosenberger calibration laboratory, which is accredited by the German accreditation body DAkkS (Deutsche Akkreditierungsstelle) according to DIN EN ISO 17025. Calibration Certificates issued by an accredited calibration laboratory are accepted worldwide to demonstrate that measurement results are traceable to national / international standards.
These standard definitions are used in the calibration of the VNA to determine the systematic errors of the measurement system. Please always use the most recent source of information (Standard Definitions Cards or VNA specific files). Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
PNA family and compatible ones ‘calkit’ file can be used on Rohde&Schwarz VNAs ZVA, ZVB, ZVBT, ZNB families and compatible ones Example file name: P5CK010-170 (BH003)_mb_2020-05-30.calkit Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
OSLT is the most commonly used two-port calibration method. The Thru connections are performed by using well-known, precision calibration adapters in male-male or female-female configuration or as a direct connection of the male and the female test ports. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
This should be done from time to time to identify defects. Air Lines, Precision Adapters and Mismatch Standards can also be used as a faster method for more frequent checks. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
5. Calibration Kit Service 5.1 Calibration interval Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the wear and the actual requirements a different calibration interval might be suitable. This lies in the responsibility of the user.
It is important to take all necessary safety precautions when using chemicals or solvents. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
This may result in buckling of the socket contacts or damage to the internal structure of a device due to the axial forces generated. NOTE: At no time should the pin depth of the connector be protruding. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 14
This will result in poor reflection and possibly unreliable contact. 6.2.2 Gauges The Rosenberger gauges are of screw-on type in plug and jack version. They can be used to gauge calibration standards as well as test ports, test port cables, adapters and DUTs.
Page 15
The Gauge Blocks are set to the reference gauge (Pin Depth) for the RPC-N 75 Ω connector. Please keep in mind that there might be different values for other N 75 Ω connector types. NOTE: further Information see user manual 05GK0KS-010 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
• Keep connectors clean. • Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
10. Use a torque wrench to make a final connection. Table 4-2 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...