Customer specific RPC-1.85 Calibration Kits, excluding Air Line Kits The Industrial Version does not contain gauges. So all gauge-related chapters are only applicable if additional gauges are available, e.g. Rosenberger 08GK0KS-010 or comparable ones. For customer specific Calibration Kits all VNA calibration methods may not be available.
User Manual RPC-1.85 Calibration Kit 08CK001-150 08CK001-150 and 08CK010-150 1.3 Kit Contents Table 1-1 Parts of the 08CK001-150 Calibration Kit Device Part number Remarks Quantity Open circuit plug 08S12L-000S3 Open circuit jack 08K12L-000S3 Short circuit plug 08S12S-000S3 Short circuit jack...
DAkkS (Deutsche Akkreditierungsstelle) according to DIN EN ISO 17025. Calibration Certificates issued by an accredited calibration laboratory are accepted worldwide to demonstrate that measurement results are traceable to national / international standards. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
These standard definitions are used in the calibration of the VNA to determine the systematic errors of the measurement system. Please always use the most recent source of information (Standard Definitions Cards or VNA specific files). Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
The file extension defines the VNA types the file can be used on. For further Information see the Application Note AN001. KitType (SN)_StdDef_CalDate.Ext • KitType 08CK001-150, 08CK010-150 or customized type • (SN) Serial number (5 character) enclosed by brackets • StdDef ‘mb’...
OSLT is the most commonly used two-port calibration method. The Thru connections are performed by using well-known, precision calibration adapters in male-male or female-female configuration or as a direct connection of the male and the female test ports. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
This should be done from time to time to identify defects. Air Lines, Precision Adapters and Mismatch Standards can also be used as a faster method for more frequent checks. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 10
5. Calibration Kit Service 5.1 Calibration interval Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the wear and the actual requirements a different calibration interval might be suitable. This lies in the responsibility of the user.
Page 11
It is important to take all necessary safety precautions when using chemicals or solvents. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 12
There are some dimensions that are critical for the mechanical integrity, non-destructive mating and electrical performance of the connector. The mechanical gauging of coaxial connectors will detect and prevent the following problems: Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 13
This will result in poor reflection and possibly unreliable contact. 6.2.2 Gauges The Rosenberger gauges are of screw-on type in plug and jack version. They can be used to gauge calibration standards as well as test ports, test port cables, adapters and DUTs.
Page 14
Figure 6-1 shows two sets of gauges and gauge blocks for the 1.85 mm / 2.4 mm connector. Figure 6-1 Connector gauges and gauge blocks for the 1.85 mm / 2.4 mm connector NOTE: For further Information see user manual 08GK0KS-010 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 15
• Keep connectors clean. • Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 16
10. Use a torque wrench to make a final connection. Table 4-2 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260...
Page 17
Rev. Engineering change number Name Date Herbert Babinger 24/04/12 Markus Hantschel 21/12/20 20-1058 Lars Ramtke 21/12/20 Rosenberger Hochfrequenztechnik GmbH & Co. KG Tel. : +49 8684 18-0 Page P.O.Box 1260 D-84526 Tittmoning Germany www.rosenberger.com Email : info@rosenberger.com 17 / 17...