GE TONOPORT VI Operator's Manual page 17

Ambulatory blood pressure system
Table of Contents

Advertisement

Before using TONOPORT VI on a patient
1.
clear the memory
2.
check date and time and adjust if required
3.
select a measurement protocol
4.
enable or disable the audio signal.
Note
When using TONOPORT VI in conjunction with
CASE/CardioSoft, it is recommended to perform
the first three steps at the PC.
Clearing the Memory
The symbol M on the display indicates that the memory
holds BP data. If these data still need to be analyzed, refer
to chapter 5 "Data Output" for details on data evaluation.
If you do not need the data any more, delete it as follows:
 Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
INFO
 Push
: the display indicates "H 1".
 Push
: the display indicates "LLLL".
START
STOP
 To delete the data, push
indicates "0000", followed by the time (if you do not
wish to clear the memory, turn off the BP monitor
instead of pushing
START
STOP
2001589-312 Revision B
again: the display
START
STOP
).
TONOPORT VI
Setup
Selecting the Measurement Method
 Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
INFO
 Push
9 times: the display indicates "H 9".
 Push
: the display indicates "0000" if the
START
STOP
selected method is the deflation measurement
method, or "1111" if the selected method is the
inflation measurement method.
 Either confirm with
INFO
tion with
, then confirm with
or switch to the other op-
START
STOP
.
START
STOP
17