Before using TONOPORT VI on a patient
1.
clear the memory
2.
check date and time and adjust if required
3.
select a measurement protocol
4.
enable or disable the audio signal.
Note
When using TONOPORT VI in conjunction with
CASE/CardioSoft, it is recommended to perform
the first three steps at the PC.
Clearing the Memory
The symbol M on the display indicates that the memory
holds BP data. If these data still need to be analyzed, refer
to chapter 5 "Data Output" for details on data evaluation.
If you do not need the data any more, delete it as follows:
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
INFO
Push
: the display indicates "H 1".
Push
: the display indicates "LLLL".
START
STOP
To delete the data, push
indicates "0000", followed by the time (if you do not
wish to clear the memory, turn off the BP monitor
instead of pushing
START
STOP
2001589-312 Revision B
again: the display
START
STOP
).
TONOPORT VI
Setup
Selecting the Measurement Method
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
INFO
Push
9 times: the display indicates "H 9".
Push
: the display indicates "0000" if the
START
STOP
selected method is the deflation measurement
method, or "1111" if the selected method is the
inflation measurement method.
Either confirm with
INFO
tion with
, then confirm with
or switch to the other op-
START
STOP
.
START
STOP
17