Page 1
Telephone: +49 (0) 5431/9416- Fax: +49 (0) 5431/9416- 66 Web: www.rebotec.de BA-Nr.: 02.09.200/2021-08-A01 Email: info@rebotec.de User Manual Patient Hoist Arnold 150 Arnold 125 320.00.10 310.00.10 150 kg 125 kg Figure similar Read this User Manual before using the device for the first time and keep it at hand for the user as well as for future use and subsequent sales.
The product is not intended for prolonged moving and transfer. Specification: • Maximum user weight for Arnold 125: 125kg • Maximum user weight for Arnold 150: 150kg • Electric lifting mechanism • 2 swivel castors, lockable • Indoor use Indication •...
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / 2.5. Packaging Symbol Meaning Symbol Meaning Degrees of • We recommend you keep the packaging for a later protection Protection transport. according class II • Dimensions : 1240mm x 350mm x 690mm to DIN EN 60529 2.6.
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / • Sudden movements and pushing as well as swing- Letters: Let- Meaning ing the patient to be lifted must be avoided due to the increased risk of accidents. Protected against access to hazardous • Never leave the patient unsupervised when using active parts with the back of a hand.
1x fuse set for charging cable ranty of 12 months. Compliance with the general 1x manual switch terms and conditions(www.rebotec.de/agbs), as • 1x user manual (02.09.200) well as the intended use, are a basic prerequisite. The statutory warranty provisions shall apply.
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / • Use the two M6x16 screws (1) (included in the de- Bring the pedal in the centre position and then ac- tuate the brakes of the rear castors (see section livery) and an Allen key (size 5) to screw the con- 11.).
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / 9.5. Connecting the electrical system • Insert the plug of the linear drive in the jack (1). Please pay attention to the positioning of the re- cess. Then insert the manual switch in the jack (2). •...
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / After the power plug has been removed, the 10.1. Connecting & charging function of the control unit will be blocked for a few seconds. Observe the safety instructions 10.2. Charge level & control display • Check the cable or plug to be sure that they are not damaged.
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / • By pressing the • EMERGENCY OFF switch, it locks Press down into position. The drive unit is interrupted and can lever in the front area with no longer be operated with the manual switch. It is your foot.
• We reserve the right to make technical develop- ments and improvements with respect to the de- sign shown in this User Manual. • See also 14. Parameters Arnold 150 Arnold 125 Load Max. 150 kg Max. 125kg *Device weight...
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / Possible Fault Correction Parameters Arnold 150 Arnold 125 cause Approx. 25 lifting cycles (depend- Greatly reduced Put drive out of op- Lifting cycles ing on the intensity of use and Motor, gear unit de- speed, loud...
20.2. Maintenance of the sling hanger versely affect the safety and functioning of the de- • The bolt and the brass washer must be checked vice. In such cases, REBOTEC will not assume any product liability. for abrasion during the annual maintenance. By pushing up the sling hanger, the bolt (A) with the •...
REBOTEC ® Rehabilitationsmittel GmbH DE-49610 Quakenbrück, Artlandstr. 57- 59 Telephone: +49 (0) 5431/9416- Fax: +49 (0) 5431/9416- 66 Web: www.rebotec.de BA-Nr.: 02.09.200/2021-08-A01 Email: info@rebotec.de 21. Spare parts & accessories 5a/b Read this User Manual before using the device for the first time and keep it at hand for the user as well as for future use and subsequent sales.
Page 15
Hoist scales for A150 with sling mounting bracket (*) *Installation height of the sling mounting bracket + hoist scales = 204mm from CSP (See 21). Use only REBOTEC spare parts and accessories to ensure the proper and safe functioning of the device.
Page 17
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / Arnold 150 Dimensions Designation in mm Highest point Highest point at highest reach Lowest point Highest reach at a height of 600 mm (reference height) Highest reach from the base frame Reach from the base frame when expanding the leg supports to 700 mm...
Page 20
BA-Nr.: 02.09.200/2021-08-A01 EBOTEC / ® REBOTEC Rehabilitationsmittel GmbH D-49610 Quakenbrück, Artlandstr. 57- 59 Telephone: +49 (0) 5431/9416-0 Fax: +49 (0) 5431/9416- 66 Web: www.rebotec.de Email: info@rebotec.de Read this User Manual before using the device for the first time and keep it at hand for the...
Need help?
Do you have a question about the Arnold 150 and is the answer not in the manual?
Questions and answers