Digital RTD Thermometer
Part# RD0370
Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RD0370 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for scigiene RD0370

  • Page 1 Digital RTD Thermometer Part# RD0370 Instruction Manual...
  • Page 2: Table Of Contents

    Table of Contents Features..................3 Specifications..................3-4 Instrument Description..............5-6 Operating Instructions..............7-9 Temperature Measurement............. 10 Battery Replacement.................11 Calibration Procedure................12 Temperature Meter Calibration Setup..........12-14 Product Maintenance.................15 Temperature Probe................15...
  • Page 3: Features

    Features This digital RTD thermometer is to be used with platinum-type temperature sensors. The temperature indication follows the IEC751 temperature table for PT-type sensors. • IP67 waterproof housing • Platinum resistor temperature sensor requires less temperature compensation than thermocouple units •...
  • Page 4 Environmental: • Operating Temperature and Humidity: 0°C ~50°C (32°F - 122°F); 0 - 80% RH • Storage Temperature and Humidity: -10°C to 60°C (14°F - 140°F); 0 - 80% RH • Altitude up to 2000 meters, Accuracy (at 23 ± 5°C): Range Accuracy -100°C ~ 300°C...
  • Page 5: Instrument Description

    Instrument Description 1 — Pt type temperature sensor connector 2 — LCD display 3 — Power ON/OFF button 4 — HOLD button 5 — MAX/MIN/Average control button 6 — Back light button 7 — Relative readout button...
  • Page 6 Display Symbol definitions and button locations This indicates that the minus temperature is sensed Celsius and Fahrenheit indication °C/°F Platinum Type Indication Pt xxx This indicates that the display data is being held HOLD The Maximum value is being displayed The Minimum value is being displayed The Average value is being displayed The reading is under Relative Mode...
  • Page 7: Operating Instructions

    Operating Instructions Turning the meter on Press the button to turn the thermometer ON or OFF. Connecting the temperature probe For measurement, plug the temperature probe into the input connectors. Selecting the Temperature Scale When the meter is first turned on, the default scale setting is set to the Celsius (°C) scale.
  • Page 8 Relative Operation When the “ REL” button is pressed, the meter will memorize the present reading and the difference between the new reading and the memorized reading and the result will be shown on the display. Press the “ REL” button again to exit the relative operation.
  • Page 9 Auto Power Off By default, the meter is in auto power off mode. The meter will power itself off after 30 minutes without operation. To disable the auto power off, press and hold the “HOLD” button and turn the meter on. There will be two successive beeps to indicate that the auto power off has been disabled.
  • Page 10: Temperature Measurement

    Temperature Measurement Correct Measurement Method The temperature sensor is located at the end of the metal sheath of the sheath type temperature probe. To accurately test internal temperatures, insert the probe to a distance of at least 15 times the diameter of the sheath directly into the item you want to measure.
  • Page 11: Battery Replacement

    Battery Replacement 1) Remove all of the back cover screws to be able to remove the cover. 2) Verify the polarity of the batteries and install the new LR03 (AAA size) Alkaline batteries into the instrument. 3) Replace cover properly onto the meter and tighten screws. The unit’s back cover is fitted with rubber rings.
  • Page 12: Calibration Procedure

    Calibration Procedure Step 1) Make sure you have a certified reference thermometer that is at least twice the accuracy of the thermometer to be recalibrated. Eg. RD0370R https://www.scigiene.com/Scigiene-Reference-Thermometer-NIST-Traceable Step 2) Insert temperature probe into https://www.scigiene.com/SCCAL-2- Calibrator https://www.scigiene.com/SCCAL-9-Calibrator https://www.scigiene.com/TempCalibrator140 at 0°C in order to calibrate precisely, and follow the steps in the calibration setup instructions.
  • Page 13 Pt type selection Press and hold “Power” + “Light” buttons for 2 seconds to enter setup mode Press “Hold” button to enter PT selection mode Press “Max” button to select between PT probe type (Pt100, Pt500, Pt1000 is circulate) Press “°C/°F” button to confirm selection O°C Calibration Press and hold “Power”...
  • Page 14 Press “REL” button to confirm present calibration value. Press “°C/°F” button to confirm PT selection Recall Default factory setting value Press and hold “Power” + “Light” buttons for 2 seconds to enter setup mode. Press “REL” button to enter calibration mode. Enter PT selection first.
  • Page 15: Product Maintenance

    In order to ensure the accuracy of the thermometer for a long period of time it should be calibrated once a year. For service (repairs or calibration) or additional information on this or any other products, contact Scigiene at 416-261-4865 or visit our website at: www.scigiene.com.