FI51P Thermometer Operating Instructions
The thermometer is a non-contact infrared thermometer, also with Probe thermometer. You can
select only one Mode at the same time but can change the Mode at will. Please remember to keep
away from baby and children and don't use it for safety related applications.
In Non-contact Infrared Thermometer function (IRT Mode)
# The laser will automatically turns on while the Scan button is pressed.
Distance:Spot (FOV)= 12:1
Simply aim the thermometer at the measure target with "Infrared Lens" and
press Meas.(infrared) key to display the surface temperature. The distance
Emissivity = 0.1~1 Step.01
to target ratio is 12:1.Please make sure the target size is within the field of
Wave Length = 8um-14um
view
While scanning, the newest temperature will be updated on the LCD and the measurement will continue as long as
the Meas.(infrared) key is depressed. When the Meas.(infrared) key is released, icon "Hold" will appear on the display
and the last measurement will remain visible for 15 seconds before the display goes blank.
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE
MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE'S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.
Mode Selection
MIN  MAX  LOCK 
MINIMUM OR MAXIMUM MODE
The thermometer will display the minimum or maximum reading during the measurement period only until the Mode
key is pressed.
To utilize the minimum mode, please press Meas.(infrared) key  Mode key  Meas.(infrared) key. And keep
pressing Meas.(infrared) key for measurement.
To utilize the maximum mode, please press Meas.(infrared) key  Mode key *twice Meas.(infrared) key. And keep
pressing Meas.(infrared) key for measurement.
o
C/
o
F  EMIS
LOCK MODE
The lock mode is particularly useful for continuous monitoring of temperatures. The thermometer will continuously
display the temperature for up to 60 minutes or until the Meas.(infrared) key button is pressed.
To utilize the lock mode, please press Meas.(infrared) key  Mode key *three times Meas.(infrared) key.
o
o
C OR
F MODE
o
o
To change the '
C' or '
F' mode, please press Meas.(infrared) key  Mode key *four times Meas.(infrared) key.
Same steps can be taken when switching from
EMISSIVITY
The infrared thermometer is supplied with a default emissivity of 0.95. The emissivity can be changed from 0.10 (10E)
to 1 (100E). Changes should only be carried out by experienced personnel. For information relating to the emissivity
of specific materials, please contact the nearest retailer. Note: non-contact infrared thermometers are not
recommended for use in measuring the temperature of shiny or polished metals.
To change the emissivity, please Meas.(infrared) key  Mode key *five times Meas.(infrared) key for each 0.01
(1E) adjustment Mode key.
In Contact Thermocouple Probe function (COT Mode)
Connect the thermocouple with Thermocouple socket and put the probe in/on the target. Press Scan key to
continuously display the temperature for up to 4 minutes. After that the device will automatically shut off to extend the
battery life. Press Scan key will interrupt the scanning to display the last temperature with a 'Hold' wording. To reenter
scanning just press Scan key again.
After measure high temp, the probe may remain HOT for a while.
The measure range is for thermometer only. User should choose proper probe types for different kinds
of application. Please make sure the target to be measured will not exceed the temperature range of
the probe to avoid permanent damage of the thermocouple probe.
To avoid electric shock and thermometer damage, do not measure live circuit where voltage exceeding
24V AC RMS or 60V DC with the thermocouple probe.
HACCP check
The "HACCP CHECK" feature is incorporated in our thermometer
temperature to graphically indicate critical temperature zone. The icons
and LED indicators located above the display indicate a food product
stays in a safe or unsafe HACCP " Danger Zone" temperature. The
green and red LED light will always be lit before power off.
A Green LED appears with icon "
" indicates a safe cool or frozen
condition below 4
o
C(40
o
F) or appears with icon "
holding temperature above 60
o
C(140
o
F).
When temperature is between 4
o
C and 60
o
C, the red LED with icon "
" will appear and indicate that the temperature is fallen within the HACCP "Danger Zone" from 4
.
(40~140
o
F)
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens
should be kept clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab
with water or medical alcohol, allowing the lens to fully dry before using the thermometer. Do not submerge any part
of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
'Hi' or 'Lo' is displayed when the temperature being measured is outside of the
measurement range.
」 「
o
o
F to
C.
" indicates a safe
o
C to 60
o
C

Advertisement

loading

Summary of Contents for scigiene FI51P

  • Page 1 LOCK MODE FI51P Thermometer Operating Instructions The lock mode is particularly useful for continuous monitoring of temperatures. The thermometer will continuously display the temperature for up to 60 minutes or until the Meas.(infrared) key button is pressed. The thermometer is a non-contact infrared thermometer, also with Probe thermometer. You can To utilize the lock mode, please press Meas.(infrared) key ...
  • Page 2 ‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed when the ambient temperature exceeds 「 」 「 」 0 C (32 F) or +50 C (122 F). The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.