Pollution degree:
Storage temperature:
Altitude:
Temperature coefficient:
Sensing:
Safety:
Overload protections:
TERMINALS (TO COM) MEASUREMENT CATEGORY:
TRANSIENT PROTECTION:
E.M.C.:
In an RF field of 3V/m:
Power supply:
Power consumption:
Low battery:
Apo timing:
Apo consumption:
2
-20°C to 60°C (-4°F to 140°F), < 80% R.H.
(With battery removed)
Operating below 2000m
Nominal 0.15 x (specified accuracy)/ °C @ (0°C -
18°C/32°F – 64.4°F) or 28°C - 45°C/82.4°F - 113°F),
or otherwise specified
AC, AC+DC true RMS
Double insulation per iec61010-1 2nd ed.,
En61010-1 2nd ed., Ul61010-1 2nd ed. & Can/csa
c22.2 No. 61010.1-0.92 To category III 1000v AC
& DC and category IV 600V AC & DC
µA & mA: 1A/600V, IR 10kA or better, F fuse
A: 10A/600V, IR 100kA or better, F fuse
V: 1050VRMS, 1450VPEAK
mV, , & others: 600 VDC & VAC RMS
V: CATEGORY III 1000V AC & DC AND CATEGORY
IV 600V AC & DC
A / mAµA: CATEGORY III AND CATEGORY IV 600
VAC AND 300 VDC.
8KV (1.2/50µS SURGE)
Meets en61326-1:2006 (EN55022, EN61000-3-
2, EN61000-3-3, EN61000- 4-2, EN61000-4-3,
EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4- 8, EN61000-4-11)
Capacitance function is not specified
Single standard 9V battery NEDA1604, JIS006P
OR IEC6F22
4.3mA typical
Below approx. 7V
Idle for 17 minutes
50µA typical
14