CODE 080120-158 Flow Calibrator OPERATION MANUAL Read this operation manual before using the product. Thank you for purchasing this product. • This operation manual describes precautions that are important for preventing accidents as well as procedures used to handle the product. •...
Table of Contents Before Use .................... 4 Safety Precautions ................5 1 Product Overview ................9 2 Features ..................9 3 Names of Parts ................10 4 Measurement ................12 Measurement Site ................. 12 Checking the Power Supply ..............12 Confirming the Flowrate Conversion Temperature Setting ....
Page 3
7 Storage ..................26 8 Maintenance ................. 27 9 Troubleshooting ................27 10 Summary of Error Screens............28 11 Specifications ................29 12 Warranty and Repairs ..............30 13 Product Disposal ................30 14 Inquiries ..................30 Problem Report Form ................. 31...
• Every effort has been made to ensure that the information in this operation manual is correct. If you discover any errors or omissions, however, please contact your Sibata representative. • The copyright of this operation manual belongs to Sibata Scientific Technology Ltd.
Items to Be Provided by the User For calibrating MP-ΣNII series and MP-W5P mini pumps, provide a buffer tank, either currently available or purchased separately. Buffer tank (Sold separately: Item code 080880-1000) Performing Automatic Calibration It requires a set of two types of communication cables for MP-W5P/MP-ΣNII. However, a commercial A-mini B USB cable can be used for automatic calibration of MP-ΣNII series mini pumps.
Page 6
OFF, remove the batteries or unplug the AC adapter from the power outlet, and then contact a Sibata representative. Never operate the instrument if it is not functioning normally. Also, never allow anyone other than a qualified service engineer to disassemble or modify the product.
Page 7
Do not insert screws or other foreign objects into instrument gaps. Doing so could cause a malfunction. If a foreign object accidentally enters the instrument, immediately switch the power OFF and remove the batteries. Then contact your Sibata representative or the distributor where you purchased the product.
Page 8
Caution Labels ■ Caution labels are affixed to the back of this product. During use, be sure to comply with the cautions noted. AVOID SHOCK Do not subject the product to strong impacts or drops. Doing so could cause a malfunction or accident. KEEP DRY Do not touch the outlet or the AC adapter plug with wet hands.
1 Product Overview The FC-M1 flow calibrator is able to measure flowrates between 0.050 and 5.000 L/min. The compact and easily portable design makes it easy to carry the unit to measurement sites for calibrating the flowrate of mini pump.
4 Measurement Measurement Site For precise measurements, perform measurements in a room (or thermostatic chamber) where temperature and humidity are controlled to constant levels. Note that the following types of sites are not appropriate for measurements. (1) Sites with a sloped floor (2) Sites with vibration (3) Sites exposed to direct sunlight (4) Sites where condensation could form...
Starting Measurements After confirming the items indicated in 4-1 to 4-5, connect the tube between the air intake port on the mini pump and the exhaust port on the FC-M1 flow calibrator. Then start operating the instrument. Confirm the flowrate value indicated on the LCD display.
LCD Screen Flashing 5 Menus Status (steady) The following describes the parameters that can be set or confirmed in respective flow calibrator menus. Main Screens The current air pressure and temperature can be confirmed by pressing the [MODE/ESC] key in the flowrate measurement screen. Flowrate Measurement Screen Pressure Sensor Display [MENU] Screen...
[MENU] Screen In the [MENU] screen, press the [ENTER] key to open the menu sub-screens. Screen Sequence Within Dashed Line Area Below Clockwise Counterclockwise (2) Exit to main screen (1) Flowrate conversion (3) Accumulation time temperature setting setting screen (6) Software version (5) Pressure display (4) Temperature display display screen...
Accumulation Time Setting The accumulation time setting can be configured. → (No time setting configured) Manual setting (default setting) → (Time setting configured) Timer setting : No accumulation time (Hours setting) (Minutes setting) setting configured Press the [ENTER] key, which causes the hours display (hours) to flash. Then use the [UP] or [DOWN] key to set the time (hours).
Press the [ENTER] key, which causes the display to flash, then use the [UP] or [DOWN] key to select auto or manual mode. In the manual mode, press the [ENTER] key again and then the [UP] or [DOWN] key to change the temperature value. Press the [ENTER] key to confirm the setting.
Displaying the Software Version The software version is displayed as shown to the right (Version 1.00 in the example shown). Returning from a [MENU] Sub-Screen to the Main Screen In the [EXIT] screen, press the [ENTER] or [MODE/ESC] key. In the [MENU] screen, press the [MODE/ESC] key.
LCD Screen Flashing 6 Calibration Status (steady) Calibrating the Pressure Sensor The pressure sensor display value can be calibrated as follows. With a main screen displayed, press the [MODE/ESC] key to display the pressure, temperature, or flowrate. Then hold the [CAL] key down for two seconds to illuminate the [CAL] indicator on the mode bar.
Calibrating the Temperature Sensor The indicated temperature sensor value can be calibrated in the same manner as the pressure sensor value, as follows. With a main screen displayed, press the [MODE/ESC] key to display the pressure or temperature. Then hold the [CAL] key down for two seconds to illuminate the [CAL] indicator on the mode bar.
Determining SPAN and ZERO Values by Calibration The SPAN and ZERO values are determined using the following formulas. Reference Measurement Value ZERO = Reference Measurement Value - Display Value SPAN = Display Value To calibrate the values based on a constantly changing value, such as temperature or atmospheric pressure, change the ZERO value.
The following describes the procedure for the MP-W5P. Switch the power ON for the mini pump. Switch the power ON for the FC-M1 flow calibrator. Warm up the FC-M1 unit by operating it for at least 10 minutes and then press the [A.ZERO] key.
Page 23
[START/STOP] key on the MP-W5P unit before pressing the [ENTER] key on the FC-M1 unit. (4) Connect tubing between the buffer tank and the exhaust port on the FC-M1 flow calibrator and between the buffer tank and the air intake port on the mini pump.
Page 24
Using an MP-W5P Mini Pump to Measure Low Flowrates (0.200 L/min or less) If the collection tube actually used is not available when using the FC-M1 flow calibrator to verify the MP-W5P mini pump flowrate, calibrate the flowrate by connecting the low-flowrate orifice adapter included with the product in front of the MP-W5P mini pump (as shown below).
Calibrate flowrate with the collection tube to be used connected to the pump. All operations are performed on the FC-M1 flow calibrator unit. Select the flowrate to calibrate using the [UP] and [DOWN] keys on the FC-M1 flow calibrator. Then press the [ENTER] key to confirm the setting.
7 Storage • Always store the FC-M1 flow calibrator with the rubber feet oriented downward. • Store the unit within a temperature range of 0 to 40 °C and humidity range of 10 to 90 % RH.
In the following cases, repair or inspection by the manufacturer is required. Contact your Sibata representative or the distributor where you purchased the product. • Over one year has elapsed since the flowrate was calibrated by Sibata. • Rubber parts have started hardening or cracking.
Over Mark Display Indicates a communication error occurred. Press the [MODE/ESC] key or restart the FC-M1. Then USB Error connect the dedicated communication cable or USB cable. Indicates automatic calibration failed. Auto-Calibration Failure Error...
MP-ΣNII series or MP-W5P mini pumps. MP-W5P/MP-ΣNII Required for calibrating flowrate on 080880-1000 Buffer tank MP-ΣNII series or MP-W5P mini pumps. Options Please purchase these separately, based on intended applications. Item Code Product Name 080120-1511 Soft case for FC-M1 080120-1571 AC adapter, SMI18-5-V-P5...
12 Warranty and Repairs If a Sibata product fails within one year from date of purchase, it will be repaired free of charge. To request repairs, contact the distributor where you purchased the product. Consumables (rubber parts) included with Sibata products are not covered by the warranty.
Problem Report Form with your shipment when requesting repairs. FC-M1 Flow Calibrator Problem Report Form If a problem occurs, please copy and fill out this Problem Report Form and contact your Sibata representative or the distributor where you purchased the product. Date completed: (yyyy/ mm/dd)