Quincy lab Digital Series Operating Manual

Quincy lab Digital Series Operating Manual

Advertisement

uincy
ab
SPECIFICATIONS
INTERIOR DIMENSIONS
EXTERIOR DIMENSIONS
TEMPERATURE RANGE
CONTROL STABILITY
STANDARD ELECTRICAL
* Standard models voltage only, optional 220 voltage available. For power requirements, see electrical info on the label located on the back of the oven.
Common Unit Construction
Exterior: Powder Coated Steel
Insulation: Fiberglass
Thermo-control: P.I.D. Microprocessor
Issue Rev 06.22
Copyright Quincy Lab Inc. 2022
MODEL
10GCE
10GCE-LT
12x10x10
INCHES W x H x D
31x25x25
(CM) W x H x D
14x17x12
INCHES W x H x D
36x43x31
(CM) W x H x D
450 / 232
210°/ 99°
@ 75°C ±°F / °C
1.0° / 0.5°
1.0° / 0.5°
115/5.2*
VOLTS / AMPS
115 / 2.60*
WATTS
5-15P*
PLUG / NEMA
WEIGHT (lbs)
SHIPPING
STAND ALONE
Quincy Lab Inc. 109 Shore Dr.
Digital Series Lab Oven
Standard Contents
(1) GCE or GCE-LT Series Lab Oven
(2) Adjustable nickel-plated wire shelf
(4) Shelf brackets
Approvals:
Underwriter's UL / CUL, United States/
Canadian for laboratory equipment.
Compliance: UL Standard 61010-1
IEC 61010-1.
MODEL
20GCE
20GCE-LT
13x13x13
33x33x33
15x21x15
38x53x38
450 / 232
210°/ 99°
1.0° / 0.5°
1.0° / 0.5°
115/6.3*
115 / 2.60*
600
750
360
360
5-15P*
44
61
38
54
Interior: Aluminized Steel
Motor:
Heater: Resistive-Tubular Incoloy
Burr Ridge, Illinois 60527
Operating Manual
For models GCE and GCE-LT
NOT FOR USE WITH FLAMMABLE
LIQUIDS OR GASES
MODEL
MODEL
30GCE
40GCE
30GCE-LT
40GCE-LT
18x16x12
18x21x14
46x41x31
46x53x36
20x25x14
20x30x16
51x64x36
51x76x41
450 / 232
450 / 232
210°/ 99°
210°/ 99°
1.0° / 0.5°
1.0° / 0.5°
1.0° / 0.5°
1.0° / 0.5°
115/10.5*
115/12.5*
115 / 3.10*
115 / 3.10*
1200
1500
720
720
5-15P*
5-15P*
78
94
70
85
N/A
1800-482-4328
PAGE 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Digital Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Quincy lab Digital Series

  • Page 1 Common Unit Construction Exterior: Powder Coated Steel Interior: Aluminized Steel Insulation: Fiberglass Motor: Thermo-control: P.I.D. Microprocessor Heater: Resistive-Tubular Incoloy Issue Rev 06.22 Quincy Lab Inc. 109 Shore Dr. Burr Ridge, Illinois 60527 1800-482-4328 Copyright Quincy Lab Inc. 2022 PAGE 1...
  • Page 2: Setup And Installation

    WARNING IMPORTANT WARNING Refers to an important note in the usage of the unit. Warns of a possible electrical shock Warns of a possible risk of fire CAUTION WARNING Cautions of HOT exterior surface during operation. Warns of possible injury or muscle strains, use assistance when moving or liftiing. Safety Precautions Read Operating Instructions Thoroughly Prior to Operation Read Operating Instructions thoroughly prior to operation and observe the following safety precautions:...
  • Page 3 Menu Level Functions Guide To navigate the controller's menu, please refer to the easy-to-use Menu Guide below: Menu setting changes can be done quickly with our 4-button digital microprocessor. Your incubator's controller allows you to perform the following: Auto-tune your Incubator for maximum efficiency Set the operating set-point temperature Lock the set-temperature Select between degrees Fahrenheit or Centigrade...
  • Page 4 AUTO-TUNING THE UNIT Your unit’s controller has been tuned from our factory to Once the desired set-point temperature has been set, achieve set-point temperatures at a faster rate. However, you will need to input the factory set passcode to access you can auto-tune your unit’s controller to work best with the tuning parameter.
  • Page 5 Chamber Loading Understanding the unit's thermal convection and "load-effect" are necessary to optimize oven’s performance. Article or media processing times and/or uniformity are largely dependent on load density and positioning. Important guidelines to chamber loading and processing: Load the oven so that heated air circulation within the chamber is not impaired. Leave a space between articles on the shelf.
  • Page 6: Troubleshooting

    (see pg. 3 for more information). Technical Support If you have any ques�ons or need technical assistance, contact Quincy Lab technical support for further assistance, or visit us on the www.quincylab.com web at Email: information@quincylab.com...