Avantes AVASPHERE-50-LS-HAL-12V Operation And Installation Manual

Light sources

Advertisement

Quick Links

AVASPHERE-50-LS-HAL-12V
LIGHT SOURCES
Operation and Installation Manual
NEED TECHNICAL SUPPORT?
Scan the QR-code or visit www.avantes.com/support
We are happy to help you!

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AVASPHERE-50-LS-HAL-12V and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Avantes AVASPHERE-50-LS-HAL-12V

  • Page 1 AVASPHERE-50-LS-HAL-12V LIGHT SOURCES Operation and Installation Manual NEED TECHNICAL SUPPORT? Scan the QR-code or visit www.avantes.com/support We are happy to help you!
  • Page 2 This document is protected by copyright. Do not copy or publish this document or parts of it without written permission of Avantes BV. Product specifications and descriptions in this manual are subject to change without notice. Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages in connection with the use of this device or information.
  • Page 3: Table Of Contents

    Settings for disable ..................6 Changing the light source bulb ................7 Technical support....................10 Getting Help ....................10 AvaSoft Updates ..................10 Specifications....................11 5.1 Ordering information ................11 Appendix A ......................12 Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...
  • Page 4: Introduction

    It is also very useful for measuring gem stones. The AvaSphere-50-LS-HAL-12V has an internal diameter of 50mm, a sample port of 10mm and a SMA terminated reference port. The 5W halogen bulb is stabilized and cooled with forced airflow.
  • Page 5: Quick Start

    2. Quick start SMA-Sample port ON-OFF switch Power LED I/O connector 12VDC Power Connector Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...
  • Page 6 ON-OFF SMA-Reference port Sample port SMA-Reference port Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...
  • Page 7: Settings For Disable

    When the light source is used in combination with an AvaSpec spectrometer, interface cable IC- DB26-AvaSphere-0.5 and AvaSoft, the output can be disabled or enabled by AvaSoft. See appendix A. Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...
  • Page 8: Changing The Light Source Bulb

    2) Disconnect the interface cable from the connector (if connected). 3) Remove the 3 torx screws on top with a torx screwdriver (Torx T10). 4) Remove the top from the AvaSphere-50-LS-HAL-12V. 5) Remove the 2 torx screws located next to the bulb with a torx screwdriver (Torx T10).
  • Page 9 8) Remove the bulb clamp from the removed bulb. 9) Replace the bulb clamp in the same direction onto the new bulb. 10) Place the new bulb inside the AvaSphere-50-LS-HAL-12V. NOTE: Do not touch the glass of the halogen bulb with bare hands! 11) Reconnect the connector of the bulb into the connector on the PCB.
  • Page 10 12) Replace the 2 torx screws aside the bulb. 13) Replace the top back onto the AvaSphere-50-LS-HAL-12V. (avoid cables getting stuck in- between) 14) Replace the 3 torx screws on top. 15) Reconnect the power connector and interface connector (if applicable).
  • Page 11: Technical Support

    4. Technical support 4.1 Getting Help If you have any questions, comments or requests concerning your AvaSphere-50-LS-HAL-12V or the AvaSoft software, please go to: http://www.avantes.com In case you have any questions or need support, please send us an email via support@avantes.com...
  • Page 12: Specifications

    Integrating sphere 50 mm for reflection (360-2500 nm), including 5W halogen light source, sample port 10mm diam., 2 SMA ports (reflection and reference) and 12V power supply included. AvaLigth-HAL-B-5W Replacement 5W halogen bulb for AvaSphere-50-LS-HAL-12V IC-DB26-AvaSphere-0,5 Interface cable to control switch line with AvaSpec AvaSphere-50-HOLD...
  • Page 13 Appendix A Start AvaSoft and press the settings button (see below). First press settings The menu as shown below wills pop-up. Press the light source control tab. Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...
  • Page 14 Please execute the following steps 1 and 2. Than press ok. The intensity mode will not work for the AvaSphere-50-LS-HAL-12V. Revision: 2-00 © Avantes BV www.avantes.com info@avantes.com...

Table of Contents