Ametek Chandler Engineering 3260 Instruction Manual page 15

Constant speed mixer
Table of Contents

Advertisement

7. To save the new settings and exit configuration mode, press and hold the Time +/-
keys simultaneously.
The available parameters for adjustment are listed below, along with their factory default
values.
• Speed Pb Value (Proportional band for PID control loop) – 29500
• Speed Ti Value (Integral/reset time for PID control loop) – 3300
• Speed Td Value (Derivative time for PID control loop) – 3300
• ACCELERATION – 5,000 RPM/sec
• DECELERATION – 5,000 RPM/sec
To restore Factory Defaults (this will restore default programs and PID values):
Unplug the mixer
Hold down the P1 and P4 buttons
Plug in the mixer. The mixer screen will show "STUCK KEY" and "POWER
OFF" and the backlight will strobe at 1Hz (turn off and on every second).
Release P1 and P4; the backlight will stop strobing.
Unplug the mixer and then plug it back in.
SECTION 3 – MAINTENANCE