Section 2 – Operating Instructions

Operation

The control enclosure includes four program buttons (P1-P4) located just below the display.
All program buttons are preset at 4,000 RPM for 15 seconds followed by 12,000 RPM for 35
seconds. These buttons may be programmed for up to 8 speed / time combinations (1,000 to
20,000 RPM and 0 to 999 seconds). Refer to the Maintenance section of this manual for
additional information on how to reprogram the functionality of these buttons.
Alternatively, starting the mixer via the Start/Stop button allows the speed to be manually
controlled via the Speed +/- buttons for continuous operation of the motor (up to 9999
seconds).

Front Panel Description

1. LCD Display – Displays current speed, mixing time and allows setup, programming,
etc.
2. Program Buttons – Allows the mixer to operate at preset speeds for preset time
intervals. Factory default settings for all Program Buttons are as follows:
a. 4,000 RPM for 15 seconds
b. 12,000 RPM for 35 seconds
3. Power On/Off (Standby) Button – Turns the power on/off.
4. Manual Speed Control Buttons – Controls the mixer speed. If idle, the Start/Stop
button must be pressed to run at the prescribed manual speed.
5. Time +/- Buttons – Used for configuring the mixer and setting program parameters.
6. Start/Stop Button – Stops the mixer if it is currently running. Starts the mixer in
manual speed control mode if it is currently stopped.
SECTION 2 – OPERATING INSTRUCTIONS