Operator Manual
POLYMETRON 9125
Conductivity/resistivity measurement
По вопросам продаж и поддержки обращайтесь:
Астана +7(77172)727-132, Волгоград (844)278-03-48, Воронеж (473)204-51-73, Екатеринбург (343)384-55-89,
Казань (843)206-01-48, Краснодар (861)203-40-90, Красноярск (391)204-63-61, Москва (495)268-04-70,
Нижний Новгород (831)429-08-12, Новосибирск (383)227-86-73, Ростов-на-Дону (863)308-18-15,
Самара (846)206-03-16, Санкт-Петербург (812)309-46-40, Саратов (845)249-38-78, Уфа (347)229-48-12
221=191=025 - Revision L - 23/02/2007
Единый адрес: hca@nt-rt.ru
Веб-сайт: hach.nt-rt.ru

Advertisement

Table of Contents
loading

Summary of Contents for Hach Ultra POLYMETRON 9125

  • Page 1 All manuals and user guides at all-guides.com Operator Manual POLYMETRON 9125 Conductivity/resistivity measurement По вопросам продаж и поддержки обращайтесь: Астана +7(77172)727-132, Волгоград (844)278-03-48, Воронеж (473)204-51-73, Екатеринбург (343)384-55-89, Казань (843)206-01-48, Краснодар (861)203-40-90, Красноярск (391)204-63-61, Москва (495)268-04-70, Нижний Новгород (831)429-08-12, Новосибирск (383)227-86-73, Ростов-на-Дону (863)308-18-15, Самара...
  • Page 2 Producer for disposal at no charge to the user. Note : For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment for proper disposal. Important document. Retain with product records. 621=191=025 Issue L Hach Ultra Analytics 11/2005...
  • Page 3: Table Of Contents

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Table of contents 1. General presentation of the conductivity / resistivity measure system ....................3 Presentation of the 9125 transmitter ................3 Introduction........................4 Conductivity measurement ....................4 Characteristics........................
  • Page 4 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement 1 point calibration ................. 32 Execution of a 2 point calibration ............33 TDS measure calibration .................. 34 Temperature electric calibration / Resistor adjust ........ 35 Process calibration ................
  • Page 5: General Presentation Of The Conductivity / Resistivity Measure System

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement General presentation of the conductivity / resistivity measure system Presentation of the 9125 transmitter The 9125 transmitter and associated measuring sensors has been designed for measuring and continuous control of conductivity and resistivity (with possibility of temperature measurement) in industrial process.
  • Page 6: Introduction

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Introduction The 9125 transmitter is a user-friendly instrument (installation, programming), equipped with microprocessor it can be configured to correspond to any application in the following sectors : drinking water, waste water, process (chemistry, paper mills, sugar mills...),...
  • Page 7 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Mobility of these species in an electric field depends naturally on their size, weight, transport charge, viscosity of the field. The greater the concentration of the species is, the greater the interactions between these ionic species is.
  • Page 8: Characteristics

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Characteristics The 9125 is equipped with an input measurement channel : a conductivity probe, 2 electrodes and inductive may be connected as well as a temperature Pt100 or Pt1000 probe.
  • Page 9 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Connections 2,5 mm screw terminals Fuse 5 x 20 mm Cartridge T2AL - 250 V Consumption 25 VA European standards EN 61326-1997 and EN61326 A1-1998 (Industrial level for immunity) EN- 61010-1 UL and CSA agreement...
  • Page 10 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Inductive probe Cell constant Resistivity range 1 cm 1 .cm...10 k .cm 2,35 cm 0,5 .cm...5 k .cm 10 cm 0,1 .cm...1 k .cm Ambient temperature -20°C...+200°C (-4…392 °F) range Display resolution...
  • Page 11 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Temperature calibration 20 °C ( 36 °F) ANALOGUE OUTPUT Output signals 2 galvanicly outputs insulated Allocation Conductivity/resistivity/temperature 0 ... 20 mA Type 4 ... 20 mA Mode Linear Dual...
  • Page 12 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement PROGRAMMING Language French English German Italian Spanish Dutch Display Icones + graphic zone (80*64 pixels) Protection codes Calibration Programming Service Page 10...
  • Page 13: Installation Of The Transmitter

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Installation of the transmitter Unpacking of the transmitter 9125 Inspect the package at the reception to detect an eventual damage due to the transport. Make sure the package contents are not damaged.
  • Page 14: Dimensions

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Dimensions (Dimensions are in mm [inches]). Fig. 2.1 Transmitter 9125 dimension Page 12...
  • Page 15: Mounting Types

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Mounting types 3 possibilities to mount the instrument (use of the red clamping bow) : The transmitter housing conforms to norm DIN 43700. Panel mounting : Panel cutting : 138 x 138 mm Front panel dimensions : 144 x 144 mm 2 screws...
  • Page 16 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Wall mounting : 2 screws 4 mm lg 60 flat head (not provided)/ 80 mm center distance Fig. 2.3 Wall mounting Page 14...
  • Page 17 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Pipe mounting : 2" maximum - 2 screws 4 mm lg 60 (provided) Fig. 2.4 Vertical mounting Fig. 2.5 Horizontal mounting Page 15...
  • Page 18 All manuals and user guides at all-guides.com...
  • Page 19: Electric Connections

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Electric connections Electronic board lay-out in the 9125 transmitter enclosure ________________________________________________________________________ Fig. 3.1 Electronic board lay-out Terminal block 4-20 mA CPU board RS485 board (option) Relay board (option) Choice between an inductive (all 4 switches on position I) or a 2 electrode (Kohlraush) probe (all 4 switches on position K) Programmed EEPROM...
  • Page 20 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement ________________________________________________________________________ Fig. 3.2 9125 shielded plate Electric connections are realised on the terminal inside the housing. Put the cables into the appropriated openings. The main supply and relay cables should be dispatched via the openings behind the shielded plate.
  • Page 21 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement ________________________________________________________________________ Fig. 3.3 Power and relays connections 1. Terminal block connections Analogue outputs 4-20 mA 2. Terminal block connections "option RS 485" 3. Terminal block connections of the conductivity sensors 4-20 mA 4.
  • Page 22: Main Connection

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Main connection Electrical connection should be performed only by qualified personnel. For the base model, the power supply accepts 100-240 VAC 10 %, (50/60 Hz) without changes in configuration. Before switching the transmitter, make sure the site voltage corresponds to the instrument voltage indicated on the identification plate.
  • Page 23: Relays Connections

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Relays connections The 9125 is equipped with 4 relays. Relay S4may be configured as a temporized relay. The nominal value of the cutoff current of each relay is 2A for 250 Vca or 0,5 A for 100 Vcc.
  • Page 24: Sensor Connections

    TEMP- black yellow white white brown internal internal shield shield (X2) external external shield shield Use only Hach Ultra Analytics supplied cables. Using other types of cables does not ensure the conformity to the electromagnetic compatibility standards. Page 22...
  • Page 25: Using The 9125 Transmitter

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Using the 9125 transmitter Utilization rules for the menus The user interface of the 9125 transmitter is made of a display screen and 4 keys. The (Esc) key is used to go back to the previous menu.
  • Page 26: Modification Of A Value

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Modification of a value The highlighted digit may be modified with the key . Each digit can be validated by pressing ENTER. Repeat both operations for each digit. Note : If you do not use the display for at least 10 minutes,...
  • Page 27: Main Display

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Main display 26.8 26.8 mS/cm : conductivity measurement 25.0 °C : temperature measurement S1...S4 : alarm status (invisible if alarm inactive). mS/cm 25.0°C Disp2 Menu Display 2 S1 : °C >...
  • Page 28: Display Options

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Display options Choice of the language English is the default language. You can choose an MENU other language available (French, German, Italian, Spanish or Dutch) by following the procedure below : CALIBRATION MAINTENANCE Use the right function key MENU.
  • Page 29: Unit Choice

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement UNIT choice As the language used, it’s possible to choose the units of the measures in which they will be display. Unit. allows to choose one of the units in DISPLAY conductivity or resistivity in which values will be S/cm...
  • Page 30: Concentration Measure (Tds)

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Concentration measure (TDS) The Total Dissolved Solid (TDS) measure is a concentration measure in the limits of measure where, for a part given, the concentration of the solution is proportional to its conductivity.
  • Page 31: Programming The Transmitter

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Programming the transmitter Main menu The main menu gives access to 4 main functions of MENU the instrument : CALIBRATION The CALIBRATION menu enables to adjust the MAINTENANCE instrument measurement...
  • Page 32: Calibration Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Calibration Menu MENU Note : CALIBRATION ETALONNAGE Before any calibration launching, check the MAINTENANCE parameters of the MEASURE menu (probe type, PROGRAMMING temp. Comp.) Are correctly configured. SERVICE Select (See page 25).
  • Page 33: Type Of Calibration

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Type of calibration COND. CALIB. With a 2 electrode probe TYPE : Elect. For the first point, remove the probe from liquid or R = 200 unscrew the connector from the probe.
  • Page 34: Point Calibration

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement COND. CALIB. 2 point calibration For the first point, remove the probe from liquid or 2 pts TYPE : unscrew the connector from the probe (2 electrode probe) in a known concentration solution.
  • Page 35: Execution Of A 2 Point Calibration

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement EXECUTION COND. CALIB. EXECUTION EXECUTION Execution of a 2 point calibration PROGRAMMING Program the manual 2 point calibration and execute it as follows : Select Remove the electrode from liquid. When the measure is stable, the instrument goes automatically to next step.
  • Page 36: Tds Measure Calibration

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement TDS measure calibration This option is available only if you have selected the measurement TDS in the menu DISPLAY. Select with the function key Choice CALIB TDS in the menu CALIBRATION and press ENTER.
  • Page 37: Temperature Electric Calibration / Resistor Adjust

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Temperature electric calibration / Resistor adjust TEMP. CALIB This calibration is factory realised. Elect. TYPE : Used to realise an electric calibration for the Pt100 RES1 : 100 measurement.
  • Page 38: Parameters Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement PARAMETERS Menu This menu displays the calibration parameters of the PARAMETERS conductivity measurement. DATE : 01/01/01 Date of last calibration. This date is the date the user SLOPE : 100% has entered after a conductivity calibration ( electric, 2 COEFF : 0.50...
  • Page 39: Maintenance Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement MAINTENANCE Menu MAINTENANCE When changing or cleaning a probe or servicing the instrument, the transmitter continues to display 6.98 measures. mS/cm 26.4°C The analogue output value is the value programmed in the mA menu.
  • Page 40: Programming Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement PROGRAMMING Menu PROGRAMMATION PROGRAMMING Note : MEASURE MESURE MESURE MESURE Warning ! An access code may be required if SORTIES mA mA OUTPUTS programmed. ALARMES ALARMS RS485 RS485 This menu enables the configuration of the instrument...
  • Page 41: S/Measure Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement MEASURE S/MEASURE Menu PROBE The PROBE menu allows to configurate the utilized probe type, TEMP. COMP. temperature TEMP. COMP. compensation, TDS the TDS measure coefficient 0.50 TDS : when it is actived.
  • Page 42: Temperature Compensation

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement TEMPERATURE COMPENSATION MEASURE : - No Choice temperature TEMP. COMP. measurement with without - Pt 100 Pt100 MEASURE : Pt100 / Pt1000. - Pt1000 TYPE : Manual TYPE : - No Choice between no temperature...
  • Page 43: S/Ma Outputs Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/mA OUTPUTS Menu mA OUPUTS This menu allows to adjust analogue outputs. OUTPUT 1 OUTPUT 2 SPECIAL PROG. TEST Select Page 41...
  • Page 44: Output 1/2

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement OUTPUT 1/2 OUTPUT 1/2 AFFECT. : AFFECT. : choice analogue output TYPE : 0/20 mA allocated to measure or temperature. MODE : Dual LOWER : 10.0 mS °C/°F MIDD : 15.0 mS UPPER : 20.0 mS...
  • Page 45: S/Alarms Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/ALARMS menu ALARMS This sub-menu allows to reach the configuration of ALARM 1 alarms 1 to 4. ALARM 2 ALARM 3 ALARM 4 Select allows to choose the operating The MODE parameter ALARM 3 mode of the 4 alarms :...
  • Page 46: Alarm 1/2 (Limit)

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement ALARM 1/2 (limit) AFFECT. : choice of a limit on the measurement ALARM 1 or on the temperature. °C/°F AFFECT. : Non LIMIT : 20.3°C DIR. : Up °C/°F DELAY : 30s HYST.
  • Page 47: Usp Mode

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement The TAB menu is used to configure the stored USP curve: W.F.I. (Water For Injection) or P.W. (Pure Water). USP mode USP is a standard used in the pharmaceutical industry. It recommends the use of a non temperature- compensated conductivity measurements and sets the upper limits of acceptable conductivity for USP-...
  • Page 48 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Example: if the measured temperature is equal to 22°C, the threshold is of 1.1 S/cm and shall remain at this value as long as the temperature stays at 22°C. If the temperature varies, rising to 25°C, the threshold automatically switches to 1.3 S/cm.
  • Page 49: Alarm 3 (Alarm System)

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement ALARM 3 (Alarm system) ALARM 3 In case of an alarm 3, choice between a limit System MODE : Syst alarm or alarm system function. ACCEPT : Manu MODE : RELAY : NO Limit...
  • Page 50: S/Special Prog. Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement SPECIAL PROG. S/SPECIAL PROG. Menu MAINTENANCE This screen allows to adjust 4..20 mA outputs states in special events : CALIBRATION SYST. ALARM MAINTENANCE CALIBRATION Select SYST. ALARM MODE : choice preset...
  • Page 51: S/Rs485 Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/RS485 Menu This option requires the RS485 kit. RS485 N° Monec number (0...32) N° : 4 BAUD : 9600 BAUD 300/600/1200/2400/4800/9600/19200 Transmission speed in bauds PARIT. : odd BIT STOP : 1 PARIT.
  • Page 52: Service Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement SERVICE Menu SERVICE Note : POLARIZATION An access code may be required if it has been AVERAGE programmed. DISPLAY CODE This screen allows to reach the 9125 transmitter SOFT ISSUE configuration screens.
  • Page 53: S/Polarization Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/POLARIZATION Menu POLARIZATION This menu is displayed only if you use a 2 electrode CABLE CAP. CAP. CABLE probe. TEST Select CABLE CAP. CABLE CAP. CABLE measure the electrode cable capacity. CAPACITY It is mandatory to disconnect the probe connector to measure the capacity or to make an electric...
  • Page 54: S/Average Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/AVERAGE Menu AVERAGE S/AVERAGE menu program a moving average on the AVERAGE : 9 measurement. TEST : TEST : Select AVERAGE : define the number of measures to calculate the average.
  • Page 55: S/Code Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/CODE Menu CODE 0000 CALIB. : access code for temperature and CALIB. : conductivity calibrations menu. PROG. : 0000 SERVICE : 0000 PROG. : access code for "Programming" menu. Select SERVICE : access code for "service"...
  • Page 56: S/Soft Issue Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/SOFT ISSUE Menu SOFT ISSUE The transmitter displays the type of instrument and the MONEC 9125 software version installed. Cond X.XX S/DEFAULT/VALUES Menu DEFAULT/VAL. Loading Note : values If you press YES, you load the default values and default ?
  • Page 57: S/Ma Adjust Menu

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement S/mA ADJUST Menu mA ADJUST This menu adjustment of the analogue outputs to OUTPUT 1 20 mA with internal coefficient between -9999...9999. OUTPUT 2 Select OUTPUT 1/2 VALUE : 0000 S/FACTORY Menu Factory code necessary.
  • Page 58 All manuals and user guides at all-guides.com...
  • Page 59: Polarization

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Polarization Electric representation of the probe and its cable In a conductivity measurement with a 2 electrode probe the measurement current is transmitted via the electrodes. The current is ensured by electrons in the electrodes and by ionic migration in the solution measured.
  • Page 60: Frequency Adjustment According To The Conductivity Measurement

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement where : The dipole Rp, Cp represents the active and reactive part of the energy necessary to the electron exchange between the sample and the electrodes. and : Cc : the cable capacity Frequency adjustment according to the conductivity...
  • Page 61: Automatic Adjustment Of The Frequency

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement measuring range 100 μS/cm to 1 mS/cm : In this measuring range the cable capacity or the polarization are negligible and do not perturb conductivity measurement. measuring range 1 mS/cm to 20 mS/cm : When the measuring frequency is low, the polarization becomes important and induces an...
  • Page 62 All manuals and user guides at all-guides.com...
  • Page 63: Error Messages

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Error messages Note : In manual acquittal, in order to suppress an error message press ENTER after correcting the default. ERROR MESSAGE DESCRIPTION/POSSIBLE CAUSE 10.3 Sensor not correctly connected Temperature sensor damaged Replace it if necessary.
  • Page 64 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement ERROR MESSAGES DURING A CALIBRATION Note : Press ESC to leave the menu and calibrate again. COND. CALIB. The electronic zero shift is superior to the limit programmed.
  • Page 65: Appendix A : Default Values

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Appendix A : Default values CALIBRATION COND. CALIB. PARAMETERS TYPE : Electr. DATE : 01/01/01 SLOPE : 100.0% RES. : 1000 : 0.0°C TEMP. CALIB. TYPE : Process PROGRAMMING MEASURE...
  • Page 66 All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement SPECIAL PROG. MAINTENANCE CALIBRATION MODE : Last MODE : Last TIMER SYST. ALARM MODE : Last MODE : Last RS485 BAUD : 19200 PARITY : No STOP BIT : 1 SWAP WORD SERVICE...
  • Page 67: Appendix B : Spare Parts List

    All manuals and user guides at all-guides.com Transmitter 9125 – Conductivity / resistivity measurement Appendix B : Spare parts list No other spare parts except those below in the table should be replaced in the instrument. Part number Description 09125=A=1001 9125 equipped CPU board 09125=A=1500 9125 complete conductivity module...
  • Page 68 All manuals and user guides at all-guides.com По вопросам продаж и поддержки обращайтесь: Астана +7(77172)727-132, Волгоград (844)278-03-48, Воронеж (473)204-51-73, Екатеринбург (343)384-55-89, Казань (843)206-01-48, Краснодар (861)203-40-90, Красноярск (391)204-63-61, Москва (495)268-04-70, Нижний Новгород (831)429-08-12, Новосибирск (383)227-86-73, Ростов-на-Дону (863)308-18-15, Самара (846)206-03-16, Санкт-Петербург (812)309-46-40, Саратов (845)249-38-78, Уфа (347)229-48-12 Единый...