ABB ADS420 Operating Instruction page 8

Dissolved oxygen probe
Hide thumbs Also See for ADS420:
Table of Contents

Advertisement

8
...3 Calibration
...1-Point calibration (water-saturated air)
9 If 1-Point Cal is displayed, go to step 12.
If 1-Point Cal is not displayed, press the
(below the Edit prompt).
The Calibration Type selection page is displayed:
Calibration Type
1-Point Cal
2-Point Cal
Cancel
10 Use the
/
keys to select 1-Point Cal and press the
key (below the OK prompt).
The Calibration Type page is displayed:
S1:
DO
Calibration Type
1-Point Cal
Next
11 Press the
key below the Next prompt).
The Calibrate/Start Span Cal? page is displayed:
Calibrate
Start Span Cal?
Place in 100 % Sat
Abort
Continue
12 Press the
key (below the Continue prompt).
The Calibrate page is displayed:
Calibrate
PV
98.26 %
Tmp
25.22 °C
Setting-Please Wait
Abort
A D S 4 2 0 | D I S S O LV E D OX YG E N P R O B E | O I/A D S 4 2 0 - E N R E V. A
key
OK
Edit
13 When span calibration is complete, a results page is
displayed:
Calibrate
PV
98.98 %
Tmp
24.90 °C
Offset
0.045
Slope
1.054
Exit
In the event of a calibration failure, the message Calibration
Failed is displayed with an indication of the reason for
failure, for example 'Result Out Of Bounds'.
14 Press the
key (below the Exit prompt).
The Calibrate page is displayed:
Calibrate
S1:
DO
Back
15 Press the
key (below the Back prompt).
The Calibrate page is displayed:
Menu
Calibrate
Exit
16 Press the
key (below the Exit prompt) to return to the
Operator page.