10. Please put the AED back into its protected casing and put it back in either
the wall mounted bracket or the alarmed security box. Please hold on to
the old set of pads so we may dispose of them. You may also recycle the
plastic cartridge and dispose of the pads yourself. Thank you.
How to install the new battery (Item # M5070A):
1. Take Onsite AED out of the red carry case.
2. Remove battery from AED
3. Remove new battery from plastic and insert battery into AED. Put the
battery, bottom end first, into the battery compartment on the back of the
AED.
4. Press the top end down until the battery clicks into place.
5. AED will begin to speak. It will do a 1-minute diagnostic self-test.
Momentarily, the AED will ask you to push the flashing orange button.
Push hard on the button for 1 second when it prompts you.
6. Once the test is complete, the AED will shut down automatically and the
green LED light (Top right corner of AED), will begin to flicker every few
seconds.
7. Servicing completed!
8. Please put the AED back into its protected casing and put it back in either
the wall mounted bracket or the alarmed security box. You may also
recycle the battery where facilities exist. Thank you.