How to install the new Smart Pads cartridge (Item # M5071A):
1. Take Onsite AED out of the red carry case.
2. You may remove battery from AED
3. Turn AED over and push the release button (to the right) at the top of the
AED and take out expired pads cartridge and put to the side.
4. Open new pads cartridge (Tear open foil covering), insert cartridge into
AED and snap pads into place.
5. Insert battery back into AED. Put the battery, bottom end first, into the
battery compartment on the back of the AED
6. Press the top end down until the battery clicks into place.
7. AED will begin to speak. It will do a 1-minute diagnostic self-test.
Momentarily, the AED will ask you to push the flashing orange button. Do
so for 1 second when it prompts you.
8. Once the test is complete, the AED will shut down automatically and the
green LED light (top right corner of AED), will begin to flicker every few
seconds.
9. Servicing completed!