Sunoptic Surgical Titan 400 Service Manual

Sunoptic Surgical Titan 400 Service Manual

Illuminator

Advertisement

S400T
Titan 400 Illuminator
Service Manual
6018 Bowdendale Avenue
Jacksonville, FL 32216
Customer Service: 904 737 7611 FAX 904 733 0012
Toll Free 877 677 2832
LIT0129Sunoptics Surgical
®
Rev. A
(English)
Page 1 of 8

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Titan 400 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sunoptic Surgical Titan 400

  • Page 1 S400T Titan 400 Illuminator Service Manual 6018 Bowdendale Avenue Jacksonville, FL 32216 Customer Service: 904 737 7611 FAX 904 733 0012 Toll Free 877 677 2832 LIT0129Sunoptics Surgical ® Rev. A (English) Page 1 of 8...
  • Page 2 Titan 400 Xenon Illuminator Service Manual TABLE OF CONTENTS Page INTRODUCTION………………………………………………………………….. 3 TECHNICAL SUPPORT SERVICES…………………………………………..… 3 GENERAL THEORY OF OPERATION ………………………………………… 4-5 CIRCUIT BREAKER REPLACEMENT……………………………………………5 SHUTTER REPLACEMENT………………………… …………………………… 5 I.R. FILTER ASSEMBLY REPLACEMENT……………………………………… 6 POWER SUPPLY REPLACEMENT …………………………………………… 6 COOLING FAN REPLACEMENT………………………………………………….7 REPLACEMENT PARTS …………………………………………………………...
  • Page 3 Titan 400 Xenon Illuminator Service Manual INTRODUCTION This manual has been prepared to aid in the repair and maintenance of the Titan 400 Xenon Illuminators. The procedures and instructions contained in this document are to be used by qualified technical personnel only. Some procedures may have live exposed circuitry and wiring which could be hazardous if contacted with.
  • Page 4 Titan 400 Xenon Illuminator Service Manual GENERAL THEORY OF OPERATION The Titan 400 illuminator general operation is as follows. Please consult Fig. 1 for wiring information. A.C. POWER DISTRIBUTION The Titan 400 Xenon Illuminators are based around a universal input power supply.
  • Page 5 Titan 400 Xenon Illuminator Service Manual COOLING SYSTEM Cooling is provided by a 86CFM, 12 VDC fan. Air is drawn through the sides of the unit, across the power supply and lamp, and exhausted through the rear mounted exhaust louver. An infrared filter which is located between the lamp lens and the lamp housing provides additional cooling of the light.
  • Page 6 Titan 400 Xenon Illuminator Service Manual I.R. FILTER REPLACEMENT CAUTION: Before performing this procedure, be sure the unit has cooled to room temperature. The lamp cartridge and IR filters operate at very high temperatures. Disconnect the illuminator from the power source.
  • Page 7 Titan 400 Xenon Illuminator Service Manual COOLING FAN REPLACEMENT Disconnect the illuminator from the power source. Remove the 11 cover screws in the sides of the unit. Open lamp access door and lift the top cover straight up and off the unit.
  • Page 8 Titan 400 Xenon Illuminator Service Manual FIGURE 1 Code Part number Description I4009400 Input module power switch CB1/CB2 I4009403 Circuit breaker I4009402 EMI filter I4009405 Power supply I1104000 Limit Switch I2000041 Reset switch I4009404 Lamp switch I4009410 Fan, 12 VDC...