PRODUCT OPERATING MANUAL 
AIRSCAN MOBILE CO MONITOR 
Manual Number: ZVP‐PC‐0070‐00 
 
TM
PANBLAST
 
 
 
 
 
 
 
 
 

Advertisement

Table of Contents
loading

Summary of Contents for Panblast ZVP-PC-0080-00

  • Page 1   PRODUCT OPERATING MANUAL  PANBLAST     AIRSCAN MOBILE CO MONITOR  Manual Number: ZVP‐PC‐0070‐00               ...
  • Page 2 SECTION                GENERAL INFORMATION             INTRODUCTION             INSTALLATION             OPERATION             INITIAL SET UP  ...
  • Page 3: General Information

    Any modification of the equipment or use of 1.1.1 All products and equipment designed and non-genuine PanBlast™ replacement parts will manufactured by Panblast are intended for use void warranty. by experienced users of abrasive blasting equipment and its associated operations and 1.3.8...
  • Page 4: Installation

    Respirators THE AIRSCAN MOBILE CO MONITOR IS NOT FOR USE IN EXPLOSIVE AREAS INSTALLATION The PanBlast™ AirScan Mobile CO Monitor is designed to operate in a portable or fixed application. In all applications it is important to consider: Availability of electrical power – 220VAC or 110VAC or 12VDC.
  • Page 5: Initial Setup

    Indicates four possible Confirm that users have read this manual and conditions: - are familiar with the operation of the PanBlast • Warm-up period two minutes (No Audio). AirScan Mobile CO Monitor. • RUN/CAL switch in CAL position (10 second beep).
  • Page 6: Operating Mode

    DO NOT CHECK THE PANBLAST AIRSCAN MOBILE CO MONITOR UNTIL IT HAS BEEN IN In normal operation the lid of the instrument OPERATION ON COMPRESSED AIR FOR AT LEAST should be closed except during the calibration 60 MINUTES TO STABILZE THE SENSOR.
  • Page 7 The following instructions cover the calibration Monitor the flow bar graph display to ensure of the PanBlast AirScan Mobile CO Monitor. GREEN condition and allow the impurity free gas to flow for at least one minute. Check the Preparing the Calibration Connector air quality indicator to ensure it is GREEN.
  • Page 8 (Reference Figure 5) Mode Selector Switches addition 12VDC Molex output connectors the PanBlast AirScan Mobile CO The PanBlast AirScan Mobile CO Monitor mode Monitor has one relay connector (J2) shown in select dip-switches have four (4) functional Figure 5. The relay connector is a switch that...
  • Page 9: Sensor Replacement

    Sensor failure will be indicated by increasing failure frequency of user response test. The following procedure should be observed in replacing the sensor. To access the PanBlast AirScan Mobile CO Monitor Sensor, loosen the screw #8, Figure 3 and open the faceplate. When the enclosure is open follow the directions listed below: 1)...
  • Page 10: Troubleshooting Guide

    TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION Check flow bar graph and ensure green Adjust regulator until green LED is lit LEDs are lit. middle scale. Unit beeps every 10 seconds. RUN/CAL switch is in the CAL position. Put RUN/CAL switch in the RUN position. Restricted internal plumbing.
  • Page 11 10.0 AirScan Mobile CO Monitor and Alarm - Assembly Stock Code Description Weight BAC-AF-0268-00 AirScan Mobile CO Monitor 2.3 Kg (5.07 lbs) 10.1 Accessories / Replacement Parts Stock Code Description BAC-AF-0270-00 CO Test Gas BAC-AF-0271-00 Impurity Free Air Test Gas BAC-AF-0269-00 Calibration Valve Assembly BAC-AF-0341-00...