Polar Instruments Ltd. T1000A Garenne Park St. Sampson SERVICE Guernsey Channel Islands MANUAL GY2 4AF ENGLAND Fax: +44 (0)1481 52476 MAN158–9604...
Page 3
INSTRUMENTS LTD and not having been altered in any way or subject to misuse, negligence or damage, and not having been repaired or attempted to be repaired by any other than POLAR INSTRUMENTS LTD or its authorized distributors. EXCEPT FOR THE FOREGOING EXPRESS WARRANTY OF...
This product is in conformity with the protection requirements of EC Council Directive 89/336/EEC on the approximation of the laws of the Member States relating to electromagnetic compatibility. A declaration of conformity with the requirements of the Directive has been signed by POLAR INSTRUMENTS (UK) LTD 11 College Place London Road Southampton...
T1000A SERVICE MANUAL SAFETY WARNING The service instructions contained in this manual are for use by qualified electronic service personnel only. WARNING The LIVE and NEUTRAL lines on this unit are BOTH fused. When the unit is connected to its supply, the opening of covers or removal of panels is likely to expose dangerous voltages.
The unit is designed for use indoors in an electrical workshop environment at a stable work station comprising a bench or similar work surface. Use only the accessories (e.g. test probes and clips) provided by Polar Instruments. The instrument must be maintained and repaired by a skilled electronics technician in accordance with the manufacturer’s instructions.
T1000A SERVICE MANUAL SPECIFICATIONS ASA Test Ranges Open Circuit Voltage Short Circuit Current Junction 500µA Logic 150mA Hi-Cap (1Hz fixed) 150mA Low-Cap (2kHz fixed) 20µA ASA Test Frequencies 95Hz Medium 500Hz High 2kHz Curve Tracing Collector drive Peak Voltage Peak Current...
160mAT SYMBOLS REFER TO MANUAL These sockets are for connecting only Polar Instruments probes and connectors for use as described in the Operator Manual. To prevent damage to this product and to ensure its safe use observe the specifications given in this manual when connecting to terminals marked with this symbol.
T1000A SERVICE MANUAL ACCESSORIES Standard Accessories Oscilloscope cable assembly ACC150 Probe set MMP159 Test clip set ACC110(Red) ACC111(Black) Operator manual MAN148...
Connect to Channel B socket and confirm a similar display. Set the T1000A to LOW range and check for a similar display from both sockets. Set the T1000A to MED range Check that the waveform at A and B is now approximately 8 divisions pk to pk Set the oscilloscope to 0.1mS/div.
Page 13
Set the T1000A STEP AMPLITUDE to fully CW and + steps of Volts. Check for a +ve going step waveform of about 1.5 divisions total amplitude. Set the oscilloscope to XY mode and 0.1V/div. and connect the T1000A by its oscilloscope cable (red to X).
R11. Note that the AND gates are turned off in ASA to stop the step generator. The output is inverted by U202A and drives the output through buffer U202B. Switch S9 (refer to T1000A Switching schematic) selects the polarity of the drive to the buffer. SINE Amplifier and Power Supply...
Note: The schematic diagram is arranged to correspond exactly with the circuit board layout as an aid to troubleshooting. The mechanical switch bank is wired to interlock many of the T1000A functions. U200B is a simple logic inverter of the control signal “JR”.
Page 16
CIRCUIT DESCRIPTION CMOS changeover switch U10 is located in the virtual earth of the output op amps. It switches between channels at a rate dependant on the ALT_CLK signal, which is once per cycle of sine wave or every two cycles depending on range. Position offset is added to the A signal at this point.
SECTION 3 – DISMANTLING INSTRUCTIONS DISMANTLING THE T1000A Removing the cover • Disconnect the power cord. • To remove the cover, undo the two screws at the rear and lift the rear of the cover first. Removing the circuit boards •...
SECTION 4 – LINE VOLTAGE AND FUSE CHANGING WARNING: There are hazardous voltages inside the instrument when connected to its power supply. REMOVE THE POWER CORD before touching any part of the line input circuit. Note that high voltages may continue to be present for 2 minutes after power is removed until internal capacitors discharge.
SECTION 5 – FAULT DIAGNOSIS WARNING: Hazardous voltages are exposed on the PCB when the cover is removed. These procedures should only be performed by a technically qualified person aware of the hazards and taking all reasonable care. TROUBLESHOOTING HINTS •...
If it becomes necessary to repair the instrument, in the first instance contact the Polar Instruments distributor in your country. In case of difficulty contact Polar Instruments Ltd. at the address at the front of this manual. Do not send the...
Note: Parts marked with an asterisk (*) have been subject to modification in later instruments. Safety critical parts (listed in bold type) must be replaced with parts obtained from Polar Instruments Ltd or your Polar Instruments distributor to ensure continued safe operation.
REPLACEMENT PARTS MECHANICAL PARTS Part Nº Quantity Description MCA138 Case Top MCA162 Front Moulding MCA164 Case Bottom (Drilled) MKB127 Knob Cap MKB128 Knob MKB129 Button caps MKB153 Power Switch Button MNS107 K30 x 6 Self Tap MNS117 M3 x 10 Pzd Screw MNS163 M3 x 12 Csk Head Screw MNS166...