Manual
BlueConnect Modules
Version of this manual: 3.0 en
www.go-sys.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Blue Connect 486 CS00-4 and is the answer not in the manual?

Questions and answers

Summary of Contents for GO Systemelektronik Blue Connect 486 CS00-4

  • Page 1 Manual BlueConnect Modules Version of this manual: 3.0 en www.go-sys.de...
  • Page 2 All rights reserved in the event of patent, utility model or design registration." Changes GO Systemelektronik GmbH retains the right to modify the contents of the manual without prior notice. Liability exclusion GO Systemelektronik GmbH takes no responsibility for correct system operation under all possible operating conditions.
  • Page 3: Table Of Contents

    Appendix B – Old Article Numbers ............................27 Appendix C – EU Declaration of Conformity Sensor Module ....................28 Appendix D – EU Declaration of Conformity I/O Module ......................29 Page 3 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 4: Introduction

    The old article numbers are listed in Appendix B - Old Article Numbers. The products of GO Systemelektronik are constantly being developed, therefore deviations between this manual and the delivered product can result. Please understand that no legal claims can be derived from the contents of this manual.
  • Page 5: Description Of The Blueconnect Modules

    Connection of analog sensors to a PLC system Connection of analog sensors and Modbus sensors to a BlueBox System Connection of analog sensors with additional power supply to a BlueBox System Page 5 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 6: Technical Data And Connections

    Open housing cover to the left. Loosen the screws (Torx T20). Sticker with pin assignment and termination see Appendix A – Interior Cover Stickers Page 6 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 7: Cable Connections, Switch Positions And Leds

    Earth the module. This is the only way to ensure trouble-free measuring operation. Sensor Module O2, pH, ISE, ORP Bus Module Current Input Module 2x 4 – 20 mA Page 7 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 8 COM4 COM5 COM6 Setting the COM Port with the DIP switches Factory setting: COM2 (COM Port 2) Relay Module Page 8 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de info@go-sys.de...
  • Page 9 CAN-H and CAN-L on slot X4. Modbus termination by means of a resistor of approx. 120 Ω on the open terminals for TX/RX+ and TX/RX- on slot X3. Example CAN bus Page 9 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 10: Pin Assignment

    CAN-L +24 VDC GND 24 V Mainboard of the BlueBox R1 – Slot X07 CAN bus 1 2 3 4 Page 10 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 11: Configuring The Blueconnect Modules With The Program Modbus Tool.exe

    This chapter describes the operation of the BlueConnect configuration program Modbus Tool.exe of GO Systemelektronik with the article number 420 6500 in software version 1.07. For example, you can use it (depending on the type of module and sensor) to read out sensor information, assign a Modbus address, calibrate the sensor and display measurement values.
  • Page 12: Title Bar And Menu Bar

    COM 1 selected The default Modbus Slave ID of a BlueConnect Sensor Module is 1 and ∗ does not need to be changed. ∗ In special cases contact GO Systemelektronik. Page 12 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 13: The Info Window

    Calibration table Calibration coefficients ∗ A raw sensor value is the uncalibrated sensor measurement value or the uncalibrated current input value. Page 13 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 14: The Calibration Table

    °C Data Processing Data Display of the current measurement values The measurement values are updated every second. COM 1 selected Page 14 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 15: The Measurement Value Recording Window

    At the bottom right of the program window this appears: D ata Logger r unning S top Click on <Stop> stops the data recording. Page 15 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 16: Configuring The Sensor Modules

    Note: Decimal separator is the comma; if a dot is entered, an error message appears. ∗ In this case, the internal temperature sensor of the O2 Sensor. Page 16 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 17: The Calibration Window O2

    The temperature adjustment can be accelerated by immersing the sensor in water, which has approximately the temperature of the ambient air, before performing the saturation calibration. Moreover, abrupt temperature changes (e.g., by direct exposure to the sun) must be avoided. Page 17 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 18: Configuring The Current Input Module

    Settings that have not yet been saved are marked in red. Note: Decimal separator is the comma; if a dot is entered, an error message appears. Page 18 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 19: Configuring The Current Output Module

    Resetting to the operating state is done by disconnecting the module from the supply voltage. Note: Decimal separator is the comma; if a dot is entered, an error message appears. Page 19 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 20: Configuring Older Bus Modules

    Writes the selected sensor identifier into the module memory. write Settings that have not yet been saved are marked in red. Page 20 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de...
  • Page 21: Modbus Address Overview Of The Sensor Modules

    32 Bit Float Note on 32 bit Float data (MSB = 0xByte4, LSB = 0xByte1) , The Receiving sequence of the values (Hex) is: 0x [Byte2] [Byte1] [Byte4] [Byte3] Page 21 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 22 32 Bit Float Note on 32 bit Float data (MSB = 0xByte4, LSB = 0xByte1) , The Receiving sequence of the values (Hex) is: 0x [Byte2] [Byte1] [Byte4] [Byte3] Page 22 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 23 32 Bit Float Note on 32 bit Float data (MSB = 0xByte4, LSB = 0xByte1) , The Receiving sequence of the values (Hex) is: 0x [Byte2] [Byte1] [Byte4] [Byte3] Page 23 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 24 32 Bit Float Note on 32 bit Float data (MSB = 0xByte4, LSB = 0xByte1) , The Receiving sequence of the values (Hex) is: 0x [Byte2] [Byte1] [Byte4] [Byte3] Page 24 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 25: Appendix A - Interior Cover Stickers

    B l u e C o n n e c t – I n t e r i o r C o v e r S t i c k e r s Appendix A – Interior Cover Stickers Page 25 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 26 B l u e C o n n e c t – I n t e r i o r C o v e r S t i c k e r s Page 26 / 29 GO Systemelektronik GmbH Faluner Weg 1...
  • Page 27: Appendix B - Old Article Numbers

    486 C000-RS05 Current Output 486 C000-mAO Output Voltage 12 V 486 C000-RS12 Relay Module Article No. old 486 C000-REL Page 27 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0 Fax: -58080-11 www.go-sys.de info@go-sys.de...
  • Page 28: Appendix C - Eu Declaration Of Conformity Sensor Module

    B l u e C o n n e c t – E U D e c l a r a t i o n s o f C o n f o r m i t y Appendix C – EU Declaration of Conformity Sensor Module Page 28 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...
  • Page 29: Appendix D - Eu Declaration Of Conformity I/O Module

    B l u e C o n n e c t – E U D e c l a r a t i o n s o f C o n f o r m i t y Appendix D – EU Declaration of Conformity I/O Module Page 29 / 29 GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0...