HoMedics HL158VA Instruction Manual page 39

Automatic wrist blood pressure monitor
Table of Contents

Advertisement

Appendix
Test specifications for enclosure port immunity to RF wireless communications
equipment.
Test frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
NOTE:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m, the 1 m test distance is permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
Modulation
Pulse modulation 18 Hz
FM ± 5 kHz deviation 1kHz sine
Pulse modulation 217 Hz
Pulse modulation 18 Hz
Pulse modulation 217 Hz
Pulse modulation 217 Hz
Pulse modulation 217 Hz
39
IMMUNITY
TEST LEVEL
(V/m)
27
28
9
28
28
28
9