CON TEN TS GEN ERAL CH ARACTERI STI CS ..........................3 D ESCRI PTI ON OF TH E FUN CTI ON S ....................... 6 TH E PROGRAM M I N G M EN U ..........................9 PROBES AN D M EASUREM EN TS ........................10 ...........................
GEN ERAL CH ARACTERI STI CS The RTD Therm om et er Model H D 2 3 0 7 .0 is a port able inst rum ent t hat m easures t he t em perat ure. I t is fit t ed w it h a lar ge LCD display for excellent visualizat ion of t he m easured dat a. The RTD Therm om et er Model H D 2 3 0 7 .0 m easures t he t em perat ure using im m ersion, penet rat ion, cont act or air probes.
Page 4
RTD The r m om e t e r H D 2 3 0 7 .0 - 4 - HD2307 V1.7...
Page 5
H D 2 3 0 7 .0 1. I nput for probes, 8- pole DI N45326 connect or. 2. Bat t ery sym bol: displays t he bat t ery charge level. 3. Funct ion indicat ors. 4. Secondary display line. 5.
D ESCRI PTI ON OF TH E FUN CTI ON S The keyboard of t he RTD Therm om et er Model H D 2 3 0 7 .0 is com posed of double- funct ion keys. The funct ion on t he key is t he " m ain funct ion" , w hile t he one above t he key is t he "...
Page 7
D isa blin g of t h e a u t om a t ic t u r n in g off The inst rum ent has an Aut oPowerOff funct ion t hat aut om at ically t urns t he inst rum ent off aft er about 8 m inut es if no key is pressed dur ing t he int ervening t im e.
Page 8
• : once t he MENU has been opened wit h t he D ATA+ UN I T keys, t he key will allow t o increase t he value of t he select ed param et er. Pressed t oget her wit h t he ON / OFF key, during t urn on, t he Aut oPowerOff funct ion is disabled ( see t he descript ion of t he ON/ OFF key) .
TH E PROGRAM M I N G M EN U To access t o t he m enu press sim ult aneously t he follow ing keys: The it em s t o be set are list ed in t his order: 1 .
PROBES AN D M EASUREM EN TS The inst rum ent works wit h t em perat ure probes fit t ed wit h SI CRAM m odule ( wit h a Platinum Pt100 sensor with 100Ω resistance) or with direct 4 wire Pt100, 3 wire Pt 100, or 2 w ire Pt 1000 sensor.
a r e ch e ck e d for con for m it y w it h cla ss A t ole r a n ce according t o norm I EC751 - BS1904 - DI N43760. The inst rum ent is provided w it h t he FACT ( fact ory) calibrat ion. The user is also able t o perform a USER ca libr a t ion of inst rum ent + probe.
TP4 7 1 0 0 ON N ECTI N G TH E CON N ECTOR FOR TH E AN D W I RE AN D 1 0 0 0 W I RE PROBES Where present , ext end t he t elescopic rod t o t he necessary lengt h paying at t ent ion t o t he cable so t hat All probes produced by Delt a OHM are provided wit h a connect or.
Page 13
6. Weld t he w ires as shown in t he t able: Se nsor TP4 7 Boa r d conne ct ion Jum pe r t o close Pt 100 4 Pt100 None 4 wires wires Pt 100 3 Pt100 3 wires wires...
IRECT CONNECTION OF WIRE SENSORS 4 w ir e Pt 1 0 0 se nsor s can be soldered Sensor Direct soldering to the connector direct ly t o t he pins of t he flying fem ale connect or wit hout m aking use of t he TP47 board.
W ARN I N GS AN D OPERATI N G I N STRUCTI ON S 1. Do not expose t he probes t o gases or liquids t hat could corrode t he m at er ial of t he sensor or t he probe it self.
I N STRUM EN T SI GN ALS AN D FAULTS The follow ing t able list s all error indicat ions and inform at ion displayed by t he inst rum ent and supplied t o t he user in different operat ing sit uat ions: D ispla y in dica t ion s Ex pla n a t ion Calibrat ion m ode...
LOW BATTERY W ARN I N G AN D BATTERY REPLACEM EN T The bat t ery sym bol on t he display const ant ly shows t he bat t er y charge st at us. To t he ext ent t hat bat t er ies have discharged, t he sym bol "...
N OTES ABOUT W ORKI N G AN D OPERATI VE SAFETY Au t hor ize d use The t echnical specificat ions as given in chapt er TECHNI CAL CHARACTERI STI CS m ust be observed. Only t he operat ion and running of t he m easur ing inst rum ent according t o t he inst ruct ions given in t his operat ing m anual is aut hor ized.
I N STRUM EN T TECH N I CAL CH ARACTERI STI CS RTD T ECH N I CAL I N FORM ATI ON ON TH E H ERM OM ETER I nst rum ent Dim ensions ( Lengt h x Widt h x Height ) 140 x 88 x 38 m m Weight 160 g ( com plet e w it h bat t eries)
ECH N I CAL D ATA OF PROBES AN D M OD ULES I N LI N E W I TH TH E I N STRUM EN T 1 0 0 SI CRAM EM PERATURE PROBES SEN SOR USI N G M OD ULE Applica t ion M ode l...
H D 2 3 0 7 .0 The kit is com posed of t he HD2307.0, 3 1.5V alkaline bat t eries, operat ing m anual, and case. The pr obe s m ust be or de r e d se pa r a t e ly.
Page 22
SI CRAM EM PERATURE PROBES W I TH OUT M OD ULE TP4 7 .1 0 0 .O I m m ersion probe, sensor Pt 100 direct 4 wires. Probe's st em Ø3 m m , lengt h 230 m m . 4- wire connect ion cable wit h connect or, lengt h 2 m et res.
Page 24
Delta OHM repairs the product s that show defect s of const ruction in accordance with the t erm s and conditions of warrant y included in the m anual of t he product . For any dispute, the com pet ent court is the Court of Padua.