IMMUNITY
IEC 60601
Compliance
test
test level
Level
Surge
±1 kV
±1 kV
differential
differential
IEC 61000-4-5
mode
mode
±2 kV common
±2 kV common
mode
mode
Voltage
<5 % U
<5 % U
T
dips, short
(>95 % dip in
(>95 % dip in
U
U
interruptions and
) for 0.5 cycle
) for 0.5 cycle
T
T
voltage variations
40 % U
40 % U
T
on power supply
(60 % dip in U
)
(60 % dip in U
T
input lines
for 5 cycles
for 5 cycles
IEC 61000-4-11
70 % U
70 % U
T
(30 % dip in U
)
(30 % dip in U
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
(>95 % dip
(>95 % dip
in U
) for 5
in U
) for 5
T
T
seconds
seconds
163
DOC40703_rev-K_manual.indd 163-164
DOC40703_rev-K_manual.indd 163-164
Electromagnetic
environment – guidance
Mains power quality should be that
of a typical domestic, commercial,
or hospital environment.
Mains power quality should be that
T
of a typical domestic, commercial,
or hospital environment. If the user
of the System requires continued
T
operation during power mains
)
T
interruptions, it is recommended
that the System be powered from
T
an uninterruptible power supply or
)
T
a battery.
T
IMMUNITY
IEC 60601
Compliance
test
test level
Level
Power frequency
30 A/m
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level
T
Electromagnetic
environment – guidance
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
domestic, commercial, or hospital
environment.
164
3/19/21 9:13 AM
3/19/21 9:13 AM