Sensor Please use sensors manufactured by jensprima; Damages to the instrument caused by the use of sensors not manufactured by jensprima within the warranty period will not be covered by the warranty. Only technicians authorized by jensprima can modify and update the instrument.
Page 5
6800 Monitor Manual The innoCon 6800O intelligent controller is designed for water treatment and industrial process monitoring. The innoSens series DO electrode can measure dissolved oxygen and temperature (ppb/ppm )in water, with automatic temperature compensation, a variety of installation methods, standard 4-20mA/RS485 output and a variety of control function output.
Page 6
Please refer to the treatment, aquaculture, etc. detailed technical documents and innoSens 425 DO Electrode choose appropriate Suitable for the measuring dissolved oxygen in water from 0 to 200.0 g/L (ppb), such as sensor boiler feed water, condensed water, etc. www.jensprima.com Page 4...
Page 7
Make a hole of 138 x 138mm on the instrument case, put the instrument into the case from the front of the panel of the case, insert screws from the back cover, and then lock the fastening block to the fastening bar from the back of the panel. 138mm Cut size www.jensprima.com Page 5...
Page 8
6800 Monitor Manual JENSPRIMA has membrane dissolved oxygen electrodes in ppm and ppb grade, ppm level(innoSens 420: 0~20.00ppm) PRIMA_pH_1BC6051 O ring 7.00 Water-proof cable 25.0 connector NORMAL WASH O ring Fixed bracket Molded shell electrode(3/4”thread) Recommend to wrap the thread with tape...
Page 9
90-260Vac) sensor -- live wire of power supply relay1 ANODE -- neutral wire of power supply CATHODE relay2 Analog current output SHIELD -- Temperature wash relay Dissolved oxygen Temperature 1 Temperature 2 RS485 output www.jensprima.com www.jensprima.com Page 7 Page 7...
Arrows key Entry into the setup mode is protected by password. The default password is 0022. to the previous menu. You can return to the measurement mode by pressing this www.jensprima.com Page 8...
Page 11
→ Relay 3 → Current 1 4.00mA → Current 2 4.00mA Entry into the setup mode is protected by password. The default password is 0022. to the previous menu. You can return to the measurement mode by pressing this www.jensprima.com Page 9...
Serial Number ▲ 4BH0001 3. Press “Menu” to return to previous menu. Restore 2. Select “Restore” and press “Enter”. Restore → YES → NO 3. Press upward and downward arrows to select YES or NO, and press “Enter”. www.jensprima.com...
1. Enter the configuration menu, select “Analysis System”and press “Enter”. 2. Select “Salinity” and press “Enter”. Salinity ▲ Salinity : 0.0 ppt Optional range: 0 ~ 99.9 ppt 3. After completion of setting, press “Enter”to save the setting, and then reture to previous menu. www.jensprima.com Page 11...
2. Select “Relay 1” and press “Enter”. Relay 1 ▲ ON/OFF: Optional: ON, OFF Action: Optional: HI, LO Set point: 100.0% Optional range:0~400.0%、0~40.00ppm、0~200.0ppb Hysteresis: 0.1% Optional range:0~50.0%、0~5.00ppm、0~50.0ppb 3. After completion of setting, press “Enter”to save the setting, and then reture to previous menu. www.jensprima.com Page 12...
Press “Enter” to test whether Relay3 is closed or open. → Current 1 4.00mA Press “Enter” to output 4mA, 12mA , 20mA compulsorily. → Current 2 4.00mA Press “Enter” to output 4mA, 12mA , 20mA compulsorily. 2. Press “Config” to return to previous menu. www.jensprima.com Page 13...
Zero Calibration ▲ Current: ﹢0.01nA Current nA value of the electrode Input Temp.: 25.0℃ Current temperature of standard solution Waiting for reading stable and then press ENTER key to exit Store calibration value → YES → NO www.jensprima.com Page 14...
→ NO Concentration Calibration Concentration Calibration Current nA value of the electrode Current : +80.0nA Input Temp.: 25.0℃ Input Conc.: 100% Input Stand Saturation value and Press ENTER Please refer to " Restore Restore → YES → NO www.jensprima.com Page 15...
12 hours. When water is stopped for a long time Close the water sample; Turn off the power; Unplug the electrode cable, remove the electrode and clean the electrode surface, and put on the protective cap. www.jensprima.com Page 16...
Page 19
Data addresses: (01) 0x01 DO measurement value Reading value: x 0.1/ x 0.01 (02) 0x02 DO current Reading value: x 0.01 (03) 0x03 Temperature value Reading value: x 0.1 (04) 0x04 Temperature current Reading value: x 0.01 www.jensprima.com Page 17...