Olympus USG-400 Troubleshooting Manual

Olympus USG-400 Troubleshooting Manual

Ultrasonic
Hide thumbs Also See for USG-400:

Advertisement

Quick Links

Disclaimer: Please refer to the complete USG-400 IFU Chapter 8, which is supplied with each item of equipment and should be thoroughly
read and understood prior to use of the equipment.
W A R N I N G
Never use the ultrasonic generator on a patient when any irregularity is observed with the equipment. Otherwise, the
system may not work properly and serious injury to the patient, surgical staff and/or surgeon may result.
When withdrawing the instruments from the body cavity due to irregular observation, gently hold the control handle and
make sure that the grasping section is closed. Otherwise, the probe tip may break and drop inside the body cavity.
If the grasping section or probe tip detaches, stop using the instruments immediately and retrieve by appropriate means.
If any irregularity is observed with the equipment, follow the below steps.

8.1 Troubleshooting

1
Immediately stop the procedure and withdraw any instrument from inside the body cavity.
2
Check that the probe tip and the grasping section are intact.
3
Check if any error window is displayed.
4
If an error window is displayed, first read Section 8.2, "Error screen" and Section 8.3, "Troubleshooting guide"
and take proper remedial actions according to "What to do when an error code is displayed" on page 140.
5
If no error window is displayed, take proper remedial actions according to "What to do when no error code is
displayed" on page 134.

8.2 Error screen

The error window is configured as shown below in Figure 8.1
Remedial actions
Next button
Previous button
Issue 1
27 May 2020
USG-400, SONICBEAT & THUNDERBEAT
Error code
Error title
Troubleshooting Guide
.
USG-400 IFU chapter 8
Symbol
Instrument in error
KMF 08808
NCDR 02576

Advertisement

loading

Summary of Contents for Olympus USG-400

  • Page 1 USG-400, SONICBEAT & THUNDERBEAT Troubleshooting Guide Disclaimer: Please refer to the complete USG-400 IFU Chapter 8, which is supplied with each item of equipment and should be thoroughly read and understood prior to use of the equipment. W A R N I N G •...
  • Page 2 USG-400, SONICBEAT & THUNDERBEAT Troubleshooting Guide Replace instrument: USG-400 IFU page 128-129 Without turning off the ultrasonic generator or disconnecting the transducer plug from the transducer socket of the ultrasonic generator, disconnect the handle from the transducer and replace it with a new THUNDERBEAT or SONICBEAT handle.
  • Page 3 USG-400, SONICBEAT & THUNDERBEAT Troubleshooting Guide Probe check result When the confirmation window (Activation OK) is displayed, tap the “OK” button. When the activation is completed: Information window (Checking) Confirmation window (Activation OK) The confirmation window (Activation OK) disappears and the “All” screen or “Set” screen is displayed.
  • Page 4 USG-400, SONICBEAT & THUNDERBEAT Troubleshooting Guide Error Code Irregularity Description Possible Causes Remedial Actions U509 Ultrasonic level 1 Reduce the force applied to the distal end or error to the probe tip. Excessive load is applied to the distal end.
  • Page 5 USG-400, SONICBEAT & THUNDERBEAT Troubleshooting Guide Error Code Irregularity Description Possible Causes Remedial Actions U511 SEAL short circuit Output is generated while a 1 Make sure that no metallic object is grasped. error metallic object is touched. Output in body fluids.