Table of Contents ············· Information on registrant/manufacturer ······················ Product introduction ······················· Indication for use ····················· Structural composition ····················· Name of components ·················· Correct application methods ··················· 1.Preventive inspections ························· 2.Unpacking ·············· 3.Battery installation or replacement ························ 4.How to use ····· Contraindications, precautions, warnings and prompts ·····················...
This Forehead Thermometer takes the body temperature by ( ) measuring the infrared energy emitted from the forehead skin Registrant/Manufacturer: Comper Chuangxiang Beijing above the eyebrows. The Forehead Thermometer adopts gentle- Technology Co., Ltd touch, user friendly and non-invasive measurement design. It Address: Building 1, Unit 4, Room 102, 103 1st Floor, No.1...
Name of components As shown in Figure 1, the product is composed of four Declaration of replaceable parts components: 1.Use 2 AAA alkaline dry batteries only, otherwise device failure 1.LED Display may occur. 2.Power/Measurre Button 3.Temperature Sensor 2.Install the batteries according to the positive and negative poles 4.Measuring Probe indication.
Page 5
± ° ± ° Product name Infrared Forehead Thermometer than 10 C (50 Model IR-FT Three short beeps are emitted to indicate high Type of temperature Infrared, Forehead environment temperature: When the temperature is higher ± ° ± ° measurements...
Correct application methods The patient is an intended operator and can perform the The device uses D.C. 3 V 2 AAA alkaline dry batteries. When the “ ” maintenance of the equipment. display shows , it indicates low battery and the batteries should be replaced as soon as possible to ensure measurement accuracy.
Page 7
How to work: 3.Sleep/Wakeup If there is no further operation, the Forehead Thermometer will 1.Measurement enter the sleep state 10s after displaying the measurement result. Power on It is feasible to wake up the device by lightly pressing the Power/ Press the button, it will start self-inspection and be ready for Measure button.
Page 8
Display icon definition Function definition Icon Status description The battery is extremely low and the product is not Display and flash working properly. Replace the Battery prompt batteries immediately. No display Full battery ° ° Temperature unit Degrees Celsius/ Degrees Fahrenheit Displayed value Measured temperature value Low temperature...
Contraindications, precautions, warnings and prompts Contraindications Warnings · It is very dangerous to make self-judgment on the Currently none measurement result or carry out treatment. Please follow the doctor's instructions. Precautions Self-judgment may lead to illness exacerbation. 1.The product is an infrared forehead thermometer for professional measurement of body temperature.
Page 10
Maintenance Before use of product, we recommend thorough cleaning of Product repair the device according to the method below, and if the device is Do not repair the product by yourself. In case of any quality issue determined not to be visually clean at the end of the cleaning step, or any doubt about the measurement accuracy of thermometer, the user should either repeat the relevant previous cleaning steps please contact the dealer or the manufacturer.
List of accessories Troubleshooting In case of any inconsistency between the physical object and the Failure/issue Possible cause Measures that can be taken following information, please contact the manufacturer. Make measurement immediately after taking Make measurement 30 min the forehead thermometer later after placing the forehead Accessory Quantity...
Clinical accuracy Statement The normal body temperatures are within a certain range. The The product is intended to measure heat radiation from forehead to body temperatures measured at different sites will be different. See display body temperature. It has been verified for clinical accuracy. the specific differences in the following table: The performance verification method for the product can be accessed by contacting out Customer Service Department.
Guidance and Manufacturer's Statement Warnings · IR-FT Infrared Forehead Thermometer shall not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary, it shall be observed to verify normal operation in the configuration in which it will be used;...
Page 14
IR-FT Infrared Forehead Thermometer is intended for use in the electromagnetic environment as specified below. The buyer or user should assure that it is used in environment as specified below. The buyer or user of IR-FT Infrared Forehead such an environment:...
Page 15
If the measured field strength in part of IR-FT Infrared Forehead the location in which the IR-FT Infrared Forehead Thermometer is used exceeds the Thermometer, including cables, applicable RF compliance level above, the IR-FT Infrared Forehead Thermometer than the recommended should be observed to verify normal operation.
IR-FT Infrared Forehead Thermometer Commission (FCC) Statement IR-FT Infrared Forehead Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The buyer or user can help prevent electromagnetic interference by maintaining a minimum distance This device complies with part 15 of the FCC Rules.
Page 17
Beijing China) take full responsibility for the conformance of the thermometer to the standard. Comper Limited Product Warranty The IR-FT Infrared Forehead Thermometer conforms to the following standards: IEC 60601-1:2005+A1:2012 Medical electrical equipment – Part 1: From the date of purchase, you can enjoy a free one-year warranty General requirements for basic safety and essential performance.