1. Information on registrant/manufacturer Name of registrant: Comper Chuangxiang(Beijing)Technology Co., Ltd Address of registrant: 102,103 - F1, Room 203,213 - F2, Building 4, No. 1 Kangding Street, Beijing Economic-Technological Development Area, Beijing Contact information of registrant: +86 10-57480968 Name of manufacturer: Comper Chuangxiang(Beijing)Technology Co., Ltd Address of manufacturer: 102,103 - F1, Room 203,213 - F2, Building 4, No.
2. Product introduction This Forehead Thermometer takes the body temperature by measuring the infrared energy emitted from the forehead skin above the eyebrows. The Forehead Thermometer adopts gentle-touch, user friendly and non-invasive measurement design. It instantly displays the accurate reading of the body temperature with just a gentle forehead touch.
Page 5
Please dispose of the replaced components (e.g. dry batteries) and the device reaching the end of service life according to the local laws and regulations. Technical indicators Product name Infrared Forehead Thermometer Model/specification IR-FT The type of temperature Infrared, Forehead measurements Temperature display 32.0℃~43.0℃...
Four short beeps are emitted to signal low battery when the actual supply voltage of the product is ≤ 2.51 V. Three short beeps are emitted to signal low environment temperature: When the temperature is lower than 10±1℃ (50± 1℉); Three short beeps are emitted to signal high environment temperature: When the temperature is higher than 40±1℃...
Check to make sure the batteries have enough power. The packaged product can be transported via common vehicles. However, it is necessary to prevent the product from rain, damp, extrusion and mechanical collision. 2) Unpacking: Open the individual-packed color box, take out the package with Forehead Thermometer and remove the Forehead Thermometer from the plastic tray.
emissivity. The default emissivity of this product is 1.0. The measured temperature will deviate from the actual value for different emissivities. 4) Operation procedures 4.1 Measurement Power On/Off: In the Off state, lightly press the “Power/Temperature Measurement Button” for to start up. The device will start self-inspection and the LED screen will be fully on. It will be ready for measurement after the self-inspection.
Page 9
After 1 min of inactivity, the Forehead Thermometer will automatically shut down. Be sure to read the IFU carefully prior to use. The ambient temperature applicable to this product is 10℃ (50℉)to 40℃ (104℉), where 25℃ (77℉) is optimal. Please do not take this product too close to a charged object to avoid electric shock. Please do not use this product in an environment with a relative humidity greater than 95%.
High temperature In body temperature mode, if the measured temperature is higher than 43.0℃ (109.4℉) warning When the ambient temperature range is beyond 10℃~40℃ (50℉-104℉) Error prompt ErrH Hardware fault ErrE Storage hardware fault 7. Contraindications, precautions, warnings and prompts 1) Contraindications Currently none 2) Precautions...
● It may lead to illness exacerbation. If the device is to be left idle for a long time (over 3 months), remove the batteries and keep out of the reach of children. ● Otherwise, it may cause ingestion by children. Keep the product out of the reach of children.
● Places with high temperature, moist, direct sunlight exposure, dust and being salty. ● If the product is to be left idle for a long time (over 3 months), please remove the batteries for safekeeping. 3) About safekeeping ● When not in use, store it in a locker/box. 4) Maintenance period ●...
Type BF applied part Keep dry Low battery prompt Avoid sunlight Straight up Fragile; Handle with care Discarded electronic or Radio transmitting equipment electrical equipment Serial number Date of manufacture Manufacturer 11. Troubleshooting Failure/issue Possible cause Measures that can be taken Make measurement immediately after Make measurement 30 min later after taking the forehead thermometer into a...
measurement. 12. Clinical accuracy: The normal body temperatures are within a certain range. The body temperatures measured at different sites will be different. See the specific differences in the following table: Measurement site Normal body temperature Deviation 0.3℃~0.5℃ (0.5-0.9℉) 34.7℃~37.8℃ Forehead lower than the oral (94.5℉-100℉)
Guidance and manufacturer's statement - Electromagnetic emission IR-FT Infrared Forehead Thermometer is intended for use in the electromagnetic environment as specified below. The buyer or user of IR-FT Infrared Forehead Thermometer should assure that it is used in such an environment: Electromagnetic environment - –Guidance...
Page 16
Table 2: Electromagnetic immunity 1 Guidance and manufacturer's statement - Electromagnetic immunity IR-FT Infrared Forehead Thermometer is intended for use in the electromagnetic environment as specified below. The buyer or user should assure that it is used in such an environment:...
Page 17
Guidance and manufacturer's statement - Electromagnetic immunity IR-FT Infrared Forehead Thermometer is intended for use in the electromagnetic environment as specified below. The buyer or user of IR-FT Infrared Forehead Thermometer should assure that it is used in such an environment: Electromagnetic environment - –...
Page 18
If the measured field strength in the location in which the IR-FT Infrared Forehead Thermometer is used exceeds the applicable RF compliance level above, the IR-FT Infrared Forehead Thermometer should be observed to verify normal operation.
Declaration of Conformity The IR-FT Infrared Forehead Thermometer meets requirements established in ASTM E 1965 for the thermometer system. And Comper Chuangxiang (Beijing)technology Co., Ltd (Add: Room 102,103 1st Floor, Unit 4,No.1 Kangding Street, Beijing Economic Technology Development Area,...
Page 20
Federal Communications Commission (FCC) Statement 1) This device may not cause harmful interference and 2) This device must accept any interference received, including interference that may cause undesired operation. 15.21 Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
Page 21
Copyright Statement The copyright is reserved by Comper Chuangxiang(Beijing)Technology Co., Ltd, which enjoys and retains all the exclusive rights of copyright according to law. No part of this document may be modified, excerpted, reproduced, or otherwise, in any form without prior written permission of the company.