Table of Contents Information of Manufacturer ..........................1 Product introduction .............................. 2 Intended use ................................2 Product composition .............................. 2 Specifications ................................ 3 How to use ................................4 Features ................................. 5 Unit Conversion ............................. 5 Power off ............................... 5 Bluetooth connection ..........................5 Dual measurement modes ........................
Product introduction ThermArt is an infrared forehead and ear thermometer which is hand-held, non-sterile, reusable and battery-powered. It is designed to measure human body temperature via one of two body sites: the ear canal or skin of the forehead. The device uses infrared technology to measure the natural thermal infrared energy emitted from the surface of the skin of the forehead or from the eardrum tissue.
How to use Please follow the steps below to achieve the best measurement results. 1. Remove the forehead probe, open the battery compartment, make sure the poles are in right direction and install the batteries. 2. Press“Power/Measure”button to turn the thermometer on. 3.
it is critical to know the individual’s normal body temperature when he/she is well. Take multiple readings to obtain the normal body temperature range and note the specific body site measured, for example: forehead or eardrum temperature. Body Site Normal Temperature Range Forehead 94.5℉-99.1℉...
4) When the probe needs to be cleaned, gently wipe the probe surface with 70% medical alcohol. 5) Please use the ear probe cover assigned by our company, or the measurement may be inaccurate. Please contact service@comper.com for assistance if you don’t kown where to purchase the ear probe cover.
17) DO NOT submerge the thermometer in any type of liquid, it is not waterproof. When wet due to contact with steam, wait until dry or gently wipe with a soft dry cloth. 18) DO NOT use the thermometer under an ambient temperature higher than 40°C (104°F) or lower than 15°C (59°F), which is beyond the operating temperature range of the thermometer.
Precautions for battery installation or replacement 1. When opening the battery cover to replace the batteries, pay attention to the positive and negative poles. Incorrect mounting will cause damage to the product. 2. In long period storage, please remove the batteries to prevent damage to the thermometer by fluid leakage.
safekeeping. 3) Safekeeping When not in use, store the thermometer in a locker or box. 4) Maintenance period It is recommended to clean and maintain the thermometer after each use. Product repair Do not repair the product by yourself. In case of any quality issue or any doubt about the measurement accuracy of the thermometer, please contact the dealer or the manufacturer.
Troubleshootings Symptom Possible cause Solution The thermometer is used within 30 Remove the thermometer from the minutes after taking from a cold situation and wait for 30 minutes in High a steady-state room condition before The thermometer probe is exposed measurement taking a temperature.
⚫ Users should install and operate the equipment according to the EMC information provided in the accompanying documents; ⚫ Portable and mobile RF communication equipment may affect the performance of IR-EFT Infrared Forehead and Ear Thermometer. Therefore during use, the product should be kept away from electromagnetic interference resources, such as radio station and microwave ovens, etc.;...
Page 14
⚫ See details of guidance and manufacturer's statement in the subsequent “Guidance and Manufacturer's Statement”. Warnings: ⚫ IR-EFT ThermArt shall not be used adjacent to or stacked with other equipment, and if adjacent or stacked use is necessary, it shall be verified the device can work in normal normal operation in the configuration in which it will be used;...
Page 15
Guidance and manufacturer's statement - Electromagnetic immunity IR-EFT ThermArt is intended for use in the electromagnetic environment as specified below. The buyer or user of IR-EFT ThermArt should assure that it is used in such an environment: Compliance Immunity test...
Page 16
RF transmitters, an electromagnetic site survey shall be considered. If the measured field strength in the location in which the IR-EFT ThermArt is used exceeds the applicable RF compliance level above, the IR-EFT ThermArt should be observed to verify normal operation.
Declaration of Conformity The IR-EFT ThermArt meets requirements established in ASTM E 1965 for the thermometer system. And Comper Chuangxiang (Beijing) Technology Co., Ltd (Add: Room 102,103 1st Floor, Building 4,No.1 Kangding Street, Beijing Economic Technology Development Area, Beijing, 100176, China) take full responsibility for the conformance of the thermometer to the standard.
4. Damage or failure caused by careless drop in the process of handling 5. Warranty does not extend to ear probe covers and batteries. If you have any questions or comments on our product, please email to service@comper.com. Manufacturered by: Comper Chuangxiang(Beijing)Technology Co., Ltd Address : Room 102,103 - F1, Room 203,213 - F2, Building 4, No.
Page 19
Federal Communications Commission (FCC) Statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: 1) This device may not cause harmful interference and 2) This device must accept any interference received, including interference that may cause undesired operation.
Page 20
Copyright Statement The copyright is reserved by Comper Chuangxiang(Beijing)Technology Co., Ltd, which enjoys and retains all the exclusive rights of copyright according to law. No part of this document may be modified, excerpted, reproduced, or otherwise, in any form without prior written permission of the company.