Universal Analyzers SCP Instruction Manual page 15

Table of Contents

Advertisement

Maintenance
MAINTENANCE SCHEDULE
The cooler heat sink is used to dissipate heat away from the heat transfer block/ Peltier elements. Over time in an
industrial environment, dust/ debris can build up between the fins on the back side of the heat sink. This build will reduce
the efficiency of the cooler and can cause premature failure of the Peltier elements.
INSPECT HEAT SINK FINS
Using a flash light (or other light source), shine a light through the heat sink fins. If the fins are unobstructed, you should
be able to see the aluminum blower fins. If the blower is not visible or partial obstruction exists, clean the heat sink fins.
CLEAN HEAT SINK FINS
Access to the left side of the cooler is required.
First, remove the side panel from the unit. Remove the screws securing the side panel. Once the panel is loose,
disconnect all electrical connectors from the power supply board.
Looking in from the left side, the blower assembly should be visible at the rear of the enclosure. Immediately in front
of the blower is the inlet side of the heat sink. Using a soft bristled brush, gently remove to debris from the heat sink.
Alternatively, a computer safe aerosol cleaner can be used to remove the debris.
Clean any loose debris from the enclosure and blower motor using a vacuum or compressed air. Replace all
connectors, reference electrical drawings below for details. Reinstall the side pane.
CLEANING INSTRUCTIONS
Should the Heat Exchangers require cleaning, disconnect the tubing and remove the Exchangers from the Heat Transfer
Blocks. Wipe off the white heat sink compound with a clean rag. Disassemble the Exchanger. Refer to Drawing Numbers
P0147 through P0149 for an illustration on the assembly of the exchanger. Wipe off all surfaces with a clean rag.
STAINLESS STEEL EXCHANGER (IF REQUIRED)
a) Soak in soap and water solution, or
b) Soak in Solvent such as MEK, Acetone, etc. or
c) Soak in a 10% HNO
PV (GLASS & KYNAR) (IF REQUIRED)
a) Soak in Alcohol to remove hydrocarbons
DO NOT USE MEK, ACETONE OR SIMILAR SOLVENTS ON PV EXCHANGERS
Re-assemble Exchangers with new O-Ring(s). Apply an even coat of Heat Sink Paste to the exterior of the exchanger
tube only. Re-install into Heat Transfer Block and reconnect tubing.
solution
3
Page 15 of 43
MAN.SCPSCU.REVD.03022015