O P E R A T I N G M A N U A L M I C R O T O M E C U T 5 0 6 2 SEMI-AUTOMATIC ROTARY MICROTOME D ES I GN & M A N UFACT UR I NG M A DE IN GE R M A NY...
1 INTENDED USE The rotary microtome CUT 5062 is intended for cutting thin sections of soft paraffin-embedded and harder specimens for professional use in routine and research laboratories in the fields of biology, medicine and industry. The system of specimen advance operates very reliable from 0.5 to 100 µm. The quality of cutting of this microtome is increased by the automatic retraction during the upstroke of the specimen, which avoids rubbing on the disposable blades or microtome knives.
MAGNIFIER WITH LIGHTING (OPTIONAL) Risk of glare and injury! Never look into the sun or any other bright light source with optical devices! Do not look directly into the light source. Fire hazard! Lenses in optical devices can cause considerable damage due to the "burning glass effect"...
Page 7
3.1.2 HANDWHEEL STOP WITH BRAKE ON THE BASE (OPTIONAL) Always use the finger protection guard with the knife- / blade holder and put hand wheel in stop position • before working with the knife or specimen, • before changing the specimen, •...
3.2 FINGER PROTECTION GUARD Use the finger protection guard for the blade and knife holder as well as the handwheel stop always Finger protection guard released • before you start any work with the knife or spec- imen, • before changing the specimen, •...
3.3 ELECTRICAL CONNECTIONS Before connecting any electrical units, the device must be switched off. Electrical connections are attached at the backside of the microtome. Further on the name plate is located here (pos. 1). Following connections are available depending on the microtome type: 2: External Control Panel 3: Mains cable...
3.4 WORKING WITH KNIVES AND DISPOSABLE BLADES Microtome knives and disposable blades have extremely sharp edges and this can lead to injuries. Please be extremely careful when handling microtome knives and disposable blades. Do not place microtome knives or disposable blades at unsecured areas. Never position micro- tome knives or disposable blades with the sharp edge pointing upwards.
4 COMPONENTS Slee rotary microtomes are provided with the following standard components: CUT 5062, #10070001 Specimen orientation • Universal cassette clamp, orientable, alternatively standard specimen clamp, orientable • 3-component knife holder, can be used for low profile blades as well as high profile blades •...
4.1 MAGNIFIER WITH ILLUMINATION, DIMMABLE (OPTIONAL) Mounting: The ring magnifier with illumination can easily be at- tached to the side of the microtome's base plate. 2 M5x12 countersunk screws (supplied) are screwed into the base plate from below. Manual_CUT_5062_2021-10-007.00_EN.docx...
Page 13
Switching on / off: The light can be switched on and off via the sensor switch on the light head. Switching on: Touch the sensor for a brief moment -> the light will be switched on (switch-on value is the last dim- ming value).
Page 14
Specifications magnifier with illumination: Protection class Protection type IP 20 Operating mode Continuous operation Technical safety check EN 60 598-1 Light head ca. Ø 122 x 13 mm Magnifier 6.0 dpt. approx. Ø 72 Rated voltage 100 - 240 V AC Frequency range 50/60 Hz Power consumption...
5 SPECIFICATIONS Nominal supply voltage 100 - 240 V + / -10 % Nominal frequency 50/60 Hz Power draw 40 VA Protective class (1) Power fuses 2 x T 1,6 A Pollution degree (1) Overvoltage installation category Operating temperature range +10 to +35 °C Operating humidity max.
6 UNPACKING AND INSTALLATION 6.1 UNPACKING THE DEVICE Remove the upper wooden cover. Remove the upper supporting foams. Lift the device out of the wooden transportation case. Grasp the device underneath the base plate from the rear and front as shown in the figure below. Do not transport the device by holding it on the hand wheel shaft, the object head, the cover or the tray.
7 OPERATION OF CUT 5062 7.1 INSERTION OF SPECIMEN IN STANDARD SPECIMEN CLAMP Turn the handwheel to the highest position and activate the handwheel lock (see chapter 3.1) and the finger guard. Loosen the standard specimen clamp by turning the fixing screw counterclockwise.
7.2 INSERTION OF SPECIMEN IN UNIVERSAL CASSETTE CLAMP Turn hand wheel to its highest position and activate the hand wheel lock. Activate finger protection guard. Open cassette fixation by pulling the fixation lever. Insert / remove cassette. Object fixation will go back into fixation position automati- cally.
7.3 SPECIMEN ORIENTATION ADJUSTMENT Turn hand wheel to its highest position and activate the hand wheel stop (see chapter 3.1). For orientation of the specimen, open the fixing lever on the right side of the orientation. For upward and downward orientation use the upper ori- entation wheel (see fig.).
7.4 EXCHANGING THE CLAMP TYPE For exchanging the universal cassette clamp with specimen orientation with the standard specimen clamp (fixed), please proceed as follows: Move the universal cassette clamp to the upper position and fix the hand wheel by pressing the brake lever. Pull out the right handle of the specimen orientation to re- lease the universal cassette clamp out of the system.
Page 21
Disassemble the specimen orientation. Disassemble the standard specimen clamp to get access to the center screw – M8 x 50 mm. Install the basic part of the standard specimen clamp and fix it with the Allen screw. Manual_CUT_5062_2021-10-007.00_EN.docx...
Page 22
Install the spindle, the washer and the knob. Manual_CUT_5062_2021-10-007.00_EN.docx...
Page 23
To replace the universal cassette clamp with specimen orientation with the standard specimen clamp with specimen orien- tation, please proceed as follows: Move the universal cassette clamp to the upper position and fix the hand wheel by pressing the brake lever. Pull out the right handle of the specimen orientation to release the universal cassette clamp out of the system.
Page 24
Install the right handle of the specimen orientation to fix the standard clamp. Manual_CUT_5062_2021-10-007.00_EN.docx...
7.5 INSERTION AND ORIENTATION OF MICROTOME KNIVES Turn the handwheel to the highest position and activate the handwheel lock (see chapter 3.1). Remove finger protection guard on the right side of the knife holder. Loosen the knife fixation by turning the two black screws counter clockwise (see fig.).
7.6 INSERTION AND ORIENTATION OF DISPOSABLE BLADES Turn the handwheel to the highest position and activate the handwheel lock (see chapter 3.1). Release the finger guard. Release the blade clamping lever on the left side of the blade holder by turning it counterclockwise. Remove the old blade or insert a new blade (see fig.).
Page 27
Horizontal orientation of the blade For adjustment of cutting area of the blade, the blade holder can be moved to the left or right. To do so, release the lever on the lower right side of the swivel piece by turning it counterclockwise (see fig.). Move the knife holder to the left or right.
Page 28
Adjusting the sledge clamping Remove the screw with Allen key size 2,5 (pos. Removing washers makes clamping tighter. Add- ing washers loosens clamping. Tighten the screw (pos. 1) and test the clamping by assembling the blade holder. Adjusting the blade clamping Remove away the blade clamping plate (pos.
Page 29
Modification from low to high blades Before any manipulation of the blade holder remove the blade! The first picture shows the low-profile configuration, un- derneath the low-profile configuration without the blade and the third picture shows the high-profile configuration. To change from low profile to high profile configuration the two screws (pos.
Page 30
Changing the lever to the other side - blade clamp Loosen the screw (located on bottom side of lat- eral sledge) with Allen key (see fig., marked with red circle no. 1). Take the lever out of the sledge and put it into the hole from the other side.
7.7 APPROACH TO SPECIMEN Turn the handwheel to the highest position and acti- vate the handwheel lock (see chapter 3.1). Loosen the knife or blade holder base by turning the left lever counter clockwise (see fig.). Loosening of knife or blade holder base fixation and move- ment towards specimen Move the blade holder base inside the guiding rail to- wards or away from the specimen (see fig.).
7.8 SPECIMEN FEED MEMORY CUT 5062 is equipped with a specimen feed memory that stores two independent positions of the specimen. The LED on the upper right of the control panel will indicate which memory position is currently active. To select the other memory position press twice quickly.
7.11 MANUAL CUTTING Release hand wheel stop (see chapter 3.1). Remove the finger guard if necessary. Turn the handwheel on the right side of the device evenly clockwise. A counter clockwise movement of the hand wheel will also result in cutting and the specimen is gradually brought closer to the blade.
7.12 ADJUSTMENT OF TRIMMING 30 µm TRIM R 5 µ The display indicates the current setting for the trimming mode (e. g. 30 µm in this example). 0000 n To change the trimming settings, press continuously and press increase or decrease the value. To return to the cutting mode, please press the key again.
7.15 SECTION COUNTER / SECTION SUM / REMAINING TRAVEL INDICATION 7 µm R 5 µ The model CUT 5062 allows optional display of different parameters in the lower left display area. These are the number of the performed sections, the section sum 0015 n or the remaining travel.
7.17 OVERVIEW KEY COMBINATIONS 7.17.1 Operation via keyboard(s) Keyboard shortcuts at power-on (for configuration) TRIM/CUT TRIM/CUT switchover latching / momentary SLOW Setting feed rate for SLOW Retraction limitation on / off INFO/RESET+MEM Set all settings to factory defaults 7.17.2 Key combinations during operation INFO/RESET+MEM Switching Rocking Mode on / off (for 5062) TRIM/CUT+SLOW+PLUS...
Page 37
7.17.4 Long keystrokes INFO/RESET (2 s) Reset cut counter MEM (2 s) Save current position in active position memory (confirmation: both position LEDs go off) REV, FWD, SLOW Direct movement of specimen holder (as long as key is pressed), while SLOW key is pressed fast "typing"...
8 EXTERNAL CONTROL PANEL CUT 5062 Before connecting any electrical units, the device must be switched off. 8.1 EXTERNAL CONTROL PANEL CUT 5062 To provide a more comfortable usage of the microtome the optional external control panels can be used for CUT 5062. The control panel is equipped with a connection cable to the microtome with a length of 1.80 m so the placement of the control panel can be arranged individually.
Remove section waste from the waste tray. Please use supplied dust cover when device is not in use. Weekly Cleaning of blade holder. Yearly Annual inspection (performed by authorized Slee service technician) • Check of all functions • Lubricating of movable parts •...
If it is necessary to return the device, it must be cleaned and disinfected before delivery. It must be returned in its original packing. If the device or parts thereof are sent back in a dirty or non-disinfected condition, SLEE medical GmbH reserves the right to return the parts to the debit of the customer.
12 WARRANTY SLEE medical GmbH guarantees that the product delivered has been subjected to a comprehensive quality control procedure, and that the product is faultless and complies with all technical specifications and / or agreed characteristics warranted. SLEE medical GmbH guarantees that the device is manufactured under an ISO 9001:2015 and ISO 13485:2016 quality management system.