Abbott IonicRF Instructions For Use Manual
Abbott IonicRF Instructions For Use Manual

Abbott IonicRF Instructions For Use Manual

Generator pole mount

Advertisement

IonicRF™ Generator Pole Mount
Model RF-POLE
Instructions for Use

Advertisement

Table of Contents
loading

Summary of Contents for Abbott IonicRF

  • Page 1 IonicRF™ Generator Pole Mount Model RF-POLE Instructions for Use...
  • Page 2 CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. ™ Indicates a trademark of the Abbott group of companies. ‡ Indicates a third-party trademark, which is property of its respective owner. Pat. http://www.abbott.com/patents...
  • Page 3: Table Of Contents

    Contents Product Description........1 Intended Use........1 Indications for Use.
  • Page 4: Product Description

    Product Description The IonicRF™ generator pole mount is a bracket that provides a method of mounting the IonicRF™ generator on a compatible equipment pole (roll stand). Intended Use The IonicRF™ generator pole mount (hereafter referred to as the "pole mount") is an accessory that is intended to be used to secure the IonicRF™...
  • Page 5: Precautions

    Precautions The following precautions apply to these components: ▪ ▪ Only use the IonicRF™ pole mount with the IonicRF™ generator model RFG‑IONIC and the roll stand model RD‑0856. ▪ ▪ Before moving the roll stand with the generator attached or...
  • Page 6: Adverse Events

    Adverse Events Superficial tissue injury may occur if the pole mount is not installed and used according to this manual. Compatibility Guidelines Only the following products may be used with the IonicRF™ generator pole mount (model RF‑POLE): Product Model IonicRF™ generator RFG‑IONIC...
  • Page 7: Preparing To Use The Pole Mount

    The IonicRF™ generator is portable and can be placed on a level surface using the countertop stand or mounted to a compatible roll stand using the IonicRF™ generator pole mount. To use the pole mount with the IonicRF generator, follow the steps in this section. NOTE: This assembly requires two people.
  • Page 8 Figure 3. Placing the pole mount on the pole Slide the pole mount onto the pole until the top of the pole mount clamp is below the bottom edge of the black pole cap. NOTE: The top of the pole mount clamp should be at least 1.2 cm below the top of the pole.
  • Page 9: Step 2: Remove The Countertop Stand

    Step 2: Remove the Countertop Stand NOTE: The IonicRF™ generator is supplied with an attached countertop stand. Before attaching the generator to the pole mount, the countertop stand must be removed. Always place the generator on a level surface when adding or removing the countertop stand.
  • Page 10: Step 3: Attach The Generator To The Pole Mount

    Step 3: Attach the Generator to the Pole Mount NOTE: Attaching the generator requires two people—one to hold the generator in position and one to attach the generator. Person 1: Lock the wheels on the roll stand. Person 1: Ensure that the pole mount is fastened securely to the pole.
  • Page 11: Cleaning Guidelines

    Figure 7. Fastening the generator Person 2: Thread each of the four captive thumb screws until all are engaged. Person 2: Hand tighten each of the four captive thumb screws until the generator is fastened securely. Cleaning Guidelines To clean the pole mount, use any of the following: ▪...
  • Page 12: Symbols And Definitions

    FDA if the medical device manufacturer is unknown. Symbols and Definitions The symbols below and harmonized symbols may be found on the product or product label. For harmonized symbols, refer to the Universal Symbols Glossary at medical.abbott/manuals. Symbol Definition Caution Refer to instruction manual/booklet...
  • Page 13 Prescription use only Authorized representative in the European Community Importer European conformity, affixed in accordance with the relevant provisions of European Council Regulation 2017/745. Hereby, Abbott Medical declares that this device is in compliance with the relevant provisions of this regulation.
  • Page 14: Additional Symbols For Product Labels

    ▪ ▪ +1 855 478 5833 (toll-free within North America) ▪ ▪ +1 651 756 5833 For additional assistance, call your local Abbott Medical representative. Warranty Information See product packaging and relevant commercial agreements for specific warranty information.
  • Page 16 Abbott Medical Abbott Medical 5050 Nathan Lane North The Corporate Village Plymouth, MN 55442 USA Da Vincilaan 11 Box F1 +1 855 478 5833 1935 Zaventem +1 651 756 5833 Belgium +32 2 774 68 11 2021-08 ARTEN600218257 A *600218257*...

This manual is also suitable for:

Rf-pole