Page 1
- ORANGE 9020BT, ULTRA 90BT, ANYPLX 3-BT Document No. : EUM-BT-01 Issued Date : 26.May, 2016 Revision : 1.2 For better performance and safety, this manual may be changed without any prior notice. The original version of ULTRA 9020BT is being written in English.
Diagnostic X-ray Unit USER MANUAL Table of Contents Section 1. (General Information) Complied Standard ...................... Contact .......................... 5 Guarantee Condition & Contact ................. Safety Information ......................Radiation Information ....................Warning Information ..................... 10 Warning Label Information ..................Label Information ......................Intended Use ........................
(Diagnostic X-ray Unit with radiation protection in accordance with EN 60601-1-3:2008) ☞ EN 60601-2-28 : 2010(Second Edition) (X-ray Tube Assembly- [ULTRA 9020BT HV Tank] EN 60601-2-28:2010) ☞ EN 60601-2-54 : 2009(First Edition) (X-ray Equipment for Radiography: [ULTRA 9020BT] EN 60601-2-54: 2009)
The guarantee does not cover additional and indirect damages related with system operation. Service may be requested by callings +82 70 7510 3400 (or email to sales@ecoray.kr) customer service department for R.A numbers. The product name, serial number, date of purchase and the details of the problem should be provided.
Diagnostic X-ray Unit USER MANUAL Safety Information This equipment is designed and manufactured with due consideration given to the safety of the operator and subject, and also to the reliability of equipment. The following precaution must be observed for additional safety. Caution [ General ] Only an experienced expert should operate the unit.
Page 8
Diagnostic X-ray Unit USER MANUAL Caution [ Battery ] When you discover strange smell or liquid leakage, turn off the power Ø immediately and report to authorized dealer or manufacturer. If any liquids from the battery get into eyes, clean it with pure water Ø...
Diagnostic X-ray Unit USER MANUAL Radiation Information This equipment is manufactured with the necessary devices and it protects operator and patient from the ionized radiation that is radiated form the X-ray source assembly Caution User and operator must use a protective devices (X-ray protection partition) and Ø...
Diagnostic X-ray Unit USER MANUAL Warning Information This is the important information for the reader or user. User or reader should pay attention to this important information. If not, serious damage to people or equipment can be occurred. Warning [ General ] Do not modify this equipment without authorization of the manufacturer.
Diagnostic X-ray Unit USER MANUAL Warning Label Information Warning of ionized Radiation This label placed on the top of top case. The ionized radiation is dangerous for the operator if the appropriate safety measures are not strictly observed. Warning of dangerous voltage This label placed on the top of HV Tank.
Diagnostic X-ray Unit USER MANUAL Label Information Beam Limiting Device This label placed on the side of Identification collimator. This label explained collimator information. This label placed on the bottom case. This label explained product information. Hand Switch This label placed on the side Hand switch.
Page 13
Diagnostic X-ray Unit USER MANUAL Identification Label ① ② ③ Explanation Basic Information ① Electric Current Direct Current User must be fully aware of the manual before User Manual using. ② Disposal of this product must be handled according Disposal to local laws and regulations Caution General caution...
Page 14
Diagnostic X-ray Unit USER MANUAL Package Symbol & Label Symbol Explanation Note Up/Down Load in correct direction to prevent the damage of equipment. Wetting Careful about wetting when the equipment is being loaded. attention Because equipment can be damaged, be careful about shock Fragile when loading.
This x-ray unit for veterinary only is equipment to diagnose the fracture of equine’s joint especially. Product Name Designation Main product name : ULTRA 9020BT Addition product name : ORANGE 9020BT, ULTRA 90BT, ANYPLX 3-BT Notes to User To ensure safe operation and long term performance stability, it is essential that you fully understand the functions, operating and maintenance instruction reading this manual before operating equipment.
Guidance and manufacturer’'s declaration – electromagnetic emissions The ULTRA 9020BT is intended for use in the electromagnetic environment specified below. The customer or the user of the ULTRA 9020BT should assure that it is used in such an environment. Electromagnetic environment – guidance...
Page 17
Guidance and manufacturer’'s declaration – electromagnetic immunity The ULTRA 9020BT Image Intensifier is intended for use in the electromagnetic environment specified below. The customer or the user of the ULTRA 9020BT Image Intensifier should assure that it is used in such an environment.
Page 18
Guidance and manufacturer’'s declaration – electromagnetic immunity The ULTRA 9020BT is intended for use in the electromagnetic environment specified below. The customer or the user of the ULTRA 9020BT should assure that it is used in such an environment. IEC 60601 Electromagnetic environment –...
Diagnostic X-ray Unit USER MANUAL Proper Environment Caution : Avoid the following environments for operation or storage ; Where the equipment is exposed to water vapor. Where the equipment is exposed to direct sunlight. Where the equipment is exposed to dust. Where the equipment is exposed to high humidity.
Diagnostic X-ray Unit USER MANUAL Name of Each Part ④ ⑤ ① ② ③ ⑦ ⑥ ⑧ ⑨ ⑩ ⑪ ⑫ ① ⑦ Handle Laser Beam Outlet ② ⑧ Operation Panel Hand Switch Connector ③ ⑨ Main Body Angle Meter ④...
Diagnostic X-ray Unit USER MANUAL Operation Panel ④ ① ⑤ ② ⑥ ⑦ ③ ⑧ Name Performance This green lamp indicates unit ready exposure. This yellow lamp indicates the exposure is being made. ① System Indicate This red lamp indicates 15 sec wait time after exposure. This red lamp indicates the system problem.
Diagnostic X-ray Unit USER MANUAL Specifications ∎ Classification ∙Electrical classification External electrical source Class II ∙Electrical type No applied parts ∙IPX specification IPX 0 : Ordinary equipment Continuous operation with intermittent ∙Mode of operation (Time interval between 1 exposure : rest time 1 min) ∎...
Diagnostic X-ray Unit USER MANUAL X-ray Tube Specification ● Manufacture Toshiba (D-0814) ● Type Stationary ● Focal spot 0.8 mm ● Target angle 16 degrees ● Target material Tungsten ● Inherent filtration At least 0.8 mm Al at 45 kV Oil immersed (60℃...
Page 26
Diagnostic X-ray Unit USER MANUAL Rating Chart...
Diagnostic X-ray Unit USER MANUAL Operation Conditions Power Source Adapter input power : 100 – 240 V~, 50/60 Hz, 80 VA Battery pack input power : 58 VDC / 1 A Battery charging time : About 3.5 hrs Power consumption : Approximately 1.35 kW (Max.) The diagnostic x-ray unit has the technique selection function as follows.
Diagnostic X-ray Unit USER MANUAL Operation Instruction Collimation Practices 1) Positioning the patient with the collimator. 2) Place the loaded cassette on the patient’s backside. 3) Set SID using with the scale that is located the side of the collimator. 4) Turn on the collimator lamp and laser pointer by pushing the collimator/laser "ON"...
Page 31
Diagnostic X-ray Unit USER MANUAL Normal Operation Check 1) Please wear a lead apron while you make exposures. 2) Check the display of operation panel for confirming the battery's state. 3) If battery charging is completed, turn the main switch "ON". Then default value(kV : 40, mAs : 0.32) will be displayed on the panel with beep sound.
Page 32
Diagnostic X-ray Unit USER MANUAL APR Mode The 'APR Mode' is a function for user's convenience, that user can store an irradiation values of each exposure part in advance. In 'APR Mode', User can store twelve values at '12 Mode' and seven values at '7 Mode'. Mode APR Key 1+5, 1+6, 1+7 / 2+5, 2+6, 2+7...
Page 33
Diagnostic X-ray Unit USER MANUAL n Implementation of Radiographs 1) Please wear a lead apron while performing an x-ray 2) Turn on the main switch. 3) If you need to warm up this unit, expose 3 times each following the 5) chart in page.31.
Diagnostic X-ray Unit USER MANUAL Error Code and Corrective Actions Code Meanings of Error Corrective Action The ready contactor of the hand Reset main switch. If proper operation is Err01 switch is closed even it is not unavailable, change the hand switch. pushed.
Diagnostic X-ray Unit USER MANUAL Designated Significant Zones of Occupancy In case the operator brings x-ray unit out of designated place to animal's location, he must take free-radiation space before x-ray examination as following diagram indicates any significant zone of occupancy designated in accordance with this sub clause shall have a floor area of dimensions not less than 60 cm x 60 cm and a height of not less than 200 cm and wear appropriate lead apron(protective device) to protect himself from leakage radiation.
Diagnostic X-ray Unit USER MANUAL User Mode In this mode, user can select the mode which they want among below table. 1. Press the 'Laser Key' for 2 second to access the User Mode. 2. Set the mode by pressing kV key(UP/DOWN). 3.
Page 39
Diagnostic X-ray Unit USER MANUAL This is a function that Mode : U05 automatically switches to Set the mAs to "01", power-saving mode if the then user can choose the operating of equipment is time by unit of 1 min, not work above the from 1 to 10.
Diagnostic X-ray Unit USER MANUAL Calibration Mode Because this equipment uses battery power source, long-time use of battery can make some deviation of kV or mA value. Therefore once in year, should be measure the kV or mA periodically by kV or mA Meter. If deviation is found, calibrate it by using one of below Mode.
Diagnostic X-ray Unit USER MANUAL Default Mode This Mode will initialize the User Mode and APR Mode that user saved. 1. The way to access the Default Mode is same with Calibration Mode. 2. The saved values of User Mode or APR Mode will be initialized if you press the “APR Store Key”...
Diagnostic X-ray Unit USER MANUAL Diagnostic X-Ray Unit Stand Fixing 1. When installing diagnostic x-ray unit to the Stand, insert captive bolt (8 mm) through middle hole (8.5 mm) as below picture and make it stable by tighten the bolt. 2.
Page 43
Diagnostic X-ray Unit USER MANUAL Section 4. Schematics...