ACIST CVi User Manual Addendum
Hide thumbs Also See for CVi:

Advertisement

Quick Links

ACIST CVi Contrast Delivery System
(including CMS2000 and E2000 Voyager)
User Manual Addendum
Bracco Group
901082-001, 03 2012-08

Advertisement

Table of Contents
loading

Summary of Contents for ACIST CVi

  • Page 1 ACIST CVi Contrast Delivery System (including CMS2000 and E2000 Voyager) User Manual Addendum Bracco Group 901082-001, 03 2012-08...
  • Page 2: Contact Information

    Copyright © 2010,2011, 2012 ACIST Medical Systems, Inc. All rights reserved. The written and graphic product descriptions in this manual were effective at the time of printing. ACIST Medical Systems, Inc. reserves the right to change specifications and designs without prior notification.
  • Page 3: Table Of Contents

    How to Transfer the Injector Head to/from the Pedestal Cart ..15 Injection Parameter Ranges and Accuracies Supported Imaging Systems Hemodynamic Transducer Specifications EMC Tables 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page iii...
  • Page 4 This page intentionally left blank. Page iv | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 5: Introduction

    Introduction Introduction Purpose The purpose of this addendum for the ACIST CVi User Manual (900468) is to provide updated information regarding: of this additional warnings and symbol definitions • Addendum adjusting the tubing guides on the saline pump, • an optional Universal Power Supply unit, •...
  • Page 6 Introduction This page intentionally left blank. Page 2 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 7: Additional Warnings And Symbol Definitions

    To avoid the risk of electric shock, this equipment must only be connected • to a supply mains with protective earth. The ACIST CVi Contrast Delivery System may only be interfaced with • X-Ray equipment certified IEC 60601-1 2nd or 3rd edition.
  • Page 8 CAUTION: Federal law (USA) restricts this device to sale, distribution, and use by, or on the order of, a physician. Dangerous voltage. Explosion hazard. Do not tip European representative. Contact for service Page 4 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 9 Fragile, handle with care. Sterilized using ethylene oxide. Environmental enclosure rating, protected against objects greater than 12 mm and protected against dripping water, Injector Head. General warning sign. 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page 5...
  • Page 10 Additional Warnings and Symbol Definitions This page intentionally left blank. Page 6 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 11: Saline Pump Setup And Adjustment

    To reduce the need for tubing guide adjustment, the following steps are Saline Pump recommended in place of steps 20-23 on page 28 of the ACIST Contrast Delivery System User Manual. 1. Open the saline pump door by rotating the pump handle counterclockwise.
  • Page 12 Move Tubing Guide Downward Tubing Secure in Saline Pump If the tubing advances through the rear tubing guide, lower the rear tubing guide one notch. Page 8 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 13: Universal Power Supply

    Universal Power Supply Universal Power Supply Optional An optional Universal Power Supply unit is available for the CVi Contrast Delivery System: Universal Thumb Screw Power Supply Equipotential Lug Unit Input Cable Connection from Injector Head Power Siemens X-Ray Table Cable Connection...
  • Page 14: Setup Instructions

    Power Supply to Siemens Scanner Connection If connecting the CVi system to a Siemens scanner, follow these steps: 1. Connect cable part number 700709 from the circular Siemens Output Cable Connection on the Universal Power Supply to the connector on the lower left corner of the injector head, as shown: 2.
  • Page 15: Power Supply To Mains Power Connection

    Connect the AC line power cable to the power entry module on the power supply. The appropriate medical grade power cable is supplied with the ACIST injection system. This cable may only be substituted with an equivalent medical-grade power cable. The power supply must be positioned to provide easy access to either or both ends of the line power cable for disconnection from mains power.
  • Page 16 Universal Power Supply This page intentionally left blank. Page 12 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 17: Pedestal Cart Specifications

    Pedestal Cart Pedestal Cart Pedestal A Pedestal Cart for the CVi Contrast Delivery System (illustrated below) is available in some world-wide markets. Cart This section provides: Pedestal Cart specifications; • Universal power supply mounting instructions; and • Injector Head cart/bed rail transferring instructions.
  • Page 18: How To Mount The Power Supply On The Pedestal Cart

    Make sure the power supply is locked and secure. NOTE: You can re-position the handle by pulling it away from the cart and then resetting its orientation. Page 14 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 19: How To Transfer The Injector Head To/From The Pedestal Cart

    (bed rail or cart). 5. Engage lock knob (release and tighten clockwise). 6. Make sure unit is locked and secure. 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page 15...
  • Page 20 Pedestal Cart This page intentionally left blank. Page 16 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 21: Injection Parameter Ranges And Accuracies

    2. In 0.1 ml increments. Accuracy = SelectedVolume ± ∗ SelectedVo lume 3. In 1 psi (or 1 kPa) increments. 4. In 0.1 sec. increments. Accuracy: ± 0.1 seconds. 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page 17...
  • Page 22 Injection Parameter Ranges and Accuracies This page intentionally left blank. Page 18 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 23 Heart Speed Series This table reflects current interface capabilities as of this revision. Please contact the manufacturer for any updates if your system is not identified here. 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page 19...
  • Page 24 Supported Imaging Systems This page intentionally left blank. Page 20 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 25 <5µA at 115 Volt (alternating current), 60 Hz Overpressure withstand -400 to +4000 mmHg Defibrillation capacity 5 discharges in 5 min. at 360 joule 15% pressure transducer band width >200 Hz 901082-001,03 ENG 2012-08 ACIST CVi Contrast Delivery System User Manual Addendum | Page 21...
  • Page 26 Hemodynamic Transducer Specifications This page intentionally left blank. Page 22 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 27 The ACIST CVi Contrast Delivery System, the CMS 2000 System, and the E2000 Voyager Systems are intended for use in the electromagnetic environment specified in the following EMC Tables. The customer or the user of the ACIST CVi Contrast Delivery System should assure that it is used in such an environment.
  • Page 28 2:2007/Ac2010 typical commercial or hospital environment. Power frequency (50/60 Hz) magnetic field Note: Ut is the A.C. mains voltage prior to application of the test level. Page 24 | ACIST CVi Contrast Delivery System User Manual Addendum 901082-001,03 ENG 2012-08...
  • Page 29 If the measured field strength in the location in which the ACIST CVi Contrast Delivery System is used exceeds the applicable RF compliance level above, the ACIST CVi Contrast Delivery System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ACIST CVi Contrast Delivery System.
  • Page 30 The ACIST CVi Contrast Delivery System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ACIST CVi Contrast Delivery System can help prevent electromagnetic interference by maintaining a minimum distance...
  • Page 32 ACIST Medical Systems, Inc 7905 Fuller Road Eden Prairie, MN 55344 USA Authorized European Representative: Medical Product Service GmbH Borngasse 20 35619 Braunfels, Germany 901082-001,03 2012-08...