Pepperl+Fuchs K-LB Series Manual page 23

Surge protection barriers
Table of Contents

Advertisement

Functional Safety K-LB-*.*, F*-LB-I, P-LB-*.*.*
Operation
Current and Voltage Measurement
Device
Step no.
K-LB-1.30
1
K-LB-1.30G
K-LB-2.30
1
2
K-LB-2.30G
1
2
3
4
Table 5.1
Device
Step no.
K-LB-1.6
1
K-LB-1.6G
K-LB-2.6
1
2
K-LB-2.6G
1
2
3
4
Table 5.2
Resistance Measurement
Device
Step no.
K-LB-1.6
1
K-LB-1.30
2
K-LB-1.6G
K-LB-1.30G
K-LB-2.6
1
K-LB-2.30
2
K-LB-2.6G
3
K-LB-2.30G
4
Table 5.3
Rated voltage
30 V between Pins 7, 8
30 V between Pins 7, 8
30 V between Pins 5, 6
30 V between Pins 7, 8
30 V between Pins 5, 6
30 V between Pins 5, 7
30 V between Pins 5, 8
Rated voltage
6 V between Pins 7, 8
6 V between Pins 7, 8
6 V between Pins 5, 6
6 V between Pins 7, 8
6 V between Pins 5, 6
6 V between Pins 5, 7
6 V between Pins 5, 8
Measurement
Resistance between Pins 1, 7
Resistance between Pins 2, 8
Resistance between Pins 1, 7
Resistance between Pins 2, 8
Resistance between Pins 3, 5
Resistance between Pins 4, 6
Leakage current
Comment
< 50 µA
Test passed if leakage
current is below 200 µA.
< 50 µA
Test passed if leakage
current is below 200 µA.
< 50 µA
Test passed if leakage
current is below 200 µA.
Leakage current
Comment
< 50 µA
Test passed if leakage
current is below 200 µA.
< 50 µA
Test passed if leakage
current is below 200 µA.
< 50 µA
Test passed if leakage
current is below 200 µA.
Comment
Test passed if resistance
is below 2 .
Test passed if resistance
is below 2 .