® ® ® ® MICRO-TEMP LT Operation and Technical Manual Model 749 Localized Therapy Units Cincinnati Sub-Zero Products, Inc. • 12011 Mosteller Road • Cincinnati, Ohio 45241, U.S.A. www.cszmedical.com In Europe: • CEpartner4U • Esdoornlaan 13, 3951 DB Maarn. The Netherlands...
Page 2
MICRO-TEMP LT Symbols / Definitions Attention: Read Operation Instructions Temperature Set Water Temperature Set and Manual Before Point Indicator Temperature Operating O / I Increasing Decreasing OFF / ON Switch Low Water Temperature Temperature Level Fuse AC Voltage CAUTION Power Failure Risk of Explosion: Separate collection for Do not use in the...
Page 3
MICRO-TEMP LT OPERATION AND TECHNICAL MANUAL Cincinnati Sub-Zero Products, Inc., reserves the right to make equipment changes and improvements, which may not be reflected in this manual. WARNING • Read Operation Manual before operating. • A physician's order is required for setting pad temperature and use of equipment. As directed by physician, check patient's temperature and skin condition of areas in contact with pad.
Page 4
WARNING • Proper sanitation procedures must be practiced and hygienic safety must be maintained, to prevent contamination. Contamination can affect patient’s health, i.e. skin irritation/rash may result. • Any time water is found leaking into or around the unit, connecting hose, and/or pad, turn the unit off, disconnect the power cord from its power source, and correct the problem before proceeding.
Page 5
Electrostatic Discharge procedure should be followed during replacement of any electronic board. • Use only CSZ provided hospital grade power cord assembly and used in conjunction with a properly grounded receptacle or electric shock may result. • Do not operate pump without a pad connected or with pad clamps in the “CLOSED”...
TABLE OF CONTENTS TECHNICAL HELP ..............................1 AUTHORIZED EUROPEAN REPRESENTATIVE: ....................1 BEFORE YOU CALL FOR SERVICE..........................1 IN-WARRANTY REPAIR AND PARTS ........................1 RECEIVING INSPECTION............................1 IMPORTANT SAFETY INFORMATION .......................1 SECTION 1. INTRODUCTION..........................2 1-0. GENERAL SAFETY PRECAUTIONS ......................2 1-1. GENERAL DESCRIPTION OF THIS MANUAL ..................2 1-2.
Page 7
SECTION 5. FIELD REPAIR/SERVICE OF THE MICRO-TEMP LT UNIT ..........34 5-1. INTRODUCTION............................34 5-2. ACCESS TO THE INTERIOR OF THE MICRO-TEMP LT UNIT ..............36 5-2.1. SEPARATING THE FRONT AND REAR ENCLOSURES............36 5-2.2. RECONNECTING THE FRONT AND REAR ENCLOSURES .............36 5-2.3. DISCONNECTING THE CABLES FROM THE CONTROL BOARD ..........37 5-2.4.
All parts on your MICRO-TEMP LT unit are covered by a one year warranty (parts and labor if returned to CSZ). To return defective parts or units, first obtain a Returned Materials Authorization (RMA) number from our Medical Technical Service department.
1-2. GENERAL DESCRIPTION OF THE MICRO-TEMP LT SYSTEM The CSZ MICRO-TEMP LT (Model 749) Localized Heat Therapy System is used to raise a patient's body part temperature and/or maintain a desired patient body part temperature through conductive heat transfer. The CSZ MICRO-TEMP LT unit is intended for continuous operation and is composed of a heater, a circulating pump, and a microprocessor.
INTRODUCTION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL The Micro-Temp LT is intended for use in ambient temperatures of 20° C – 35° C (68° F – 95° F). The maximum contact surface temperature at the pad is 42° C (107° F). 1-3.
INTRODUCTION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 1-4.2. EXTERNAL FEATURES - REAR VIEW The external features in Figure (1-2) of the MICRO-TEMP LT unit are described as follows: The power switch is a beveled rocker switch labeled “I” for ON and “O” for OFF. The fuse housing holds replaceable fuses that are integrated in the switch assembly to protect against overload conditions.
INTRODUCTION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 1-4.3. EXTERNAL FEATURES – MEMBRANE CONTROL PANEL The external features in Figure (1-3) and Figure (1-4) of the MICRO-TEMP LT unit are described as follows: The digital temperature display shows the actual temperature of the circulating water and the set point temperature when the “SET”...
INTRODUCTION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 1-5. REQUIRED ACCESSORIES The only essential accessory for the MICRO-TEMP LT unit is a thermal pad. Thermal pads are available with or without a layer of foam bonded to one side and with a non-woven fabric on both sides to allow moist heat therapy when the foam or non-woven fabric is wetted.
INTRODUCTION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL The 0.24-gallon (0.9 liters) capacity reservoir is composed of a single circulating unit. The circulating water flows over and around the heating element and water temperature sensor located in the circulating reservoir. The heated water then flows out of the reservoir, to the circulating pump, through the pump housing and connecting hoses, and then to the localized heat therapy pad.
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL SECTION 2. GENERAL PREPARATION OF THE MICRO-TEMP LT SYSTEM 2-1. INTRODUCTION This section describes the procedures to prepare the MICRO-TEMP LT unit for general use. This entails unpacking the shipment, arranging all the equipment for the first time, and completing a test routine.
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Review Section (1-4.1), (1-4.2), and (1-4.3) to identify the features of the MICRO- TEMP LT unit. Collect and arrange a localized heat therapy pad described in Section (1-5). CAUTION • Use distilled water only.
Page 20
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL a. Upon initial power-up, the unit flashes the 42° C default set point four times, and then flashes the actual water temperature two times before constantly displaying the water temperature. The pump activates and begins circulating water.
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL The set point display changes and the numbers move up the scale. The longer the button is pressed the faster the digits change. When the button is released and repressed, the digits once again change slowly and then increase in speed.
Page 22
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL WARNING • A physician's order is required for setting pad temperature and use of equipment. As directed by physician, check the patient's temperature and skin condition of areas in contact with the pad. Check the pad water temperature. Pediatric patients, temperature- sensitive patients with vascular disease, surgical patients, diabetics, and Raynauds disease patients should be checked more frequently.
Page 23
GENERAL PREPARATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL WARNING • Any time water is found leaking into or around the unit, connecting hose, and/or pad, turn the unit off, disconnect the power cord from its power source, and correct the problem before proceeding.
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL SECTION 3. OPERATING THE MICRO-TEMP LT UNIT 3-1. INTRODUCTION This section describes how to operate the MICRO-TEMP LT unit in order to provide maximum therapy to a patient. First, collect the equipment and prepare the patient. Second, set the appropriate controls.
Page 25
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Inspect the power plug for bent or missing prongs. Connect the detachable power cord to the unit. WARNING • Do not by-pass ground on power cord plug. Electrical Hazards may result. Insert the plug into a properly grounded receptacle.
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 3-3. OPERATING THE MICRO-TEMP LT WARNING • A physician's order is required for setting pad temperature and use of equipment. As directed by physician, check patient's temperature and skin condition of areas in contact with pad;...
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL When the desired set temperature is displayed, release the Up or Down arrow and the “SET” button. NOTE: If water temperature is >1° C (2˚F) over the set po int temperature, the display will flash and the unit will sound an alarm.
Page 28
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL During normal operation, the status display shows the following message: Status Display Message Function The water temperature is displayed when the Micro-Temp LT is operating. If the “SET” button is being depressed, the set point temperature is being displayed.
Page 29
OPERATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL This alarm condition occurs when the actual water temperature is 1° C (2˚F) greater than the set point temperature. W hen this occurs, the display flashes the actual water temperature, an audible alarm sounds 2 beeps while flashing the red ”Warning”...
Test Tilt Switch Alarm 4-1.1. TEST EQUIPMENT REQUIRED The following test equipment is required to perform the preventive maintenance/functional check- out procedures: • CSZ's model TF100 86173 Dual-Matic (Temperature Tester, Flow Meter) • Calibrated temperature measuring device • Digital Safety Analyzer •...
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL SUGGESTED PREVENTIVE MAINTENANCE CHECKLIST MICRO-TEMP LT Model 749 Hospital Control No. Serial Number Check When Completed External cabinet in good condition. (No unusual dents or missing parts) All labels properly affixed. Quick disconnect couplings are tight, straight, and not leaking.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-2. MAINTENANCE OF THE WATER RESERVOIR The reservoir holds approximately 0.24 gallons (0.9 liters) of distilled water that remains in the unit between periods of use. Quarterly, the water reservoir should be drained and replenished. NOTE: DO NOT USE DE-IONIZED WATER, TAP WATER, OR ALCOHOL.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-2.3. DISINFECTION/DRY STORAGE PROCEDURE FOR CIRCULATING WATER UNITS Quarterly, at a minimum, the FLUID CIRCUIT DISINFECTION/DRY STORAGE PROCEDURE, listed below, should be conducted. CAUTION • For safe handling and use of chemicals follow manufacturer guidelines. •...
Page 34
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL UNIT U.S.P. Grade Propylene Glycol Per Unit MICRO-TEMP LT 2 ounces (60 cc) Continue to fill the water reservoir with distilled water. If unit is being placed in dry storage, continue with procedure. When all fluid has been removed from the unit, disconnect the drain hose(s) and wipe unit clean.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-3. MAINTENANCE OF THE MICRO-TEMP LT EXTERIOR – CLEANING INSTRUCTIONS The MICRO-TEMP LT unit is constructed of a medical grade plastic. For cleaning and disinfecting, always use conventional hospital-approved topical equipment cleaners and disinfectants that do not contain alcohol.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-6. GROUND CONTINUITY TEST WARNING • The repair, calibration, and servicing of the MICRO-TEMP LT should be performed by qualified Medical Equipment Service Technicians, Certified Biomedical Electronics Technicians, or Certified Clinical Engineers familiar with good repair practices for servicing medical devices, and in accordance with instructions contained in this manual.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-8. >1° C (2° F) OVER SET POINT CHECK WARNING • The repair, calibration, and servicing of the MICRO-TEMP LT should be performed by qualified Medical Equipment Service Technicians, Certified Biomedical Electronics Technicians, or Certified Clinical Engineers familiar with good repair practices for servicing medical devices, and in accordance with instructions contained in this manual.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-10. HIGH LIMIT SAFETY CHECK WARNING • The repair, calibration, and servicing of the MICRO-TEMP LT should be performed by qualified Medical Equipment Service Technicians, Certified Biomedical Electronics Technicians, or Certified Clinical Engineers familiar with good repair practices for servicing medical devices, and in accordance with instructions contained in this manual.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL CAUTION • Always drain the MICRO-TEMP LT to a sanitary drain because bio-contaminants may be present in the unit’s water supply. Twist off reservoir cap. Turn unit upside down and pour water into sanitary drain until the water level is to the bottom of the sight window.
MAINTENANCE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 4-13. TILT SWITCH ALARM CHECK WARNING • The repair, calibration, and servicing of the MICRO-TEMP LT should be performed by qualified Medical Equipment Service Technicians, Certified Biomedical Electronics Technicians, or Certified Clinical Engineers familiar with good repair practices for servicing medical devices, and in accordance with instructions contained in this manual.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL SECTION 5. FIELD REPAIR/SERVICE OF THE MICRO-TEMP LT UNIT WARNING • Always unplug the unit before accessing internal components during service. Failure to unplug the unit could result in electric shock. •...
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Figures (5-1), (6-1), (6-3), and (6-5) highlight the interior components of the MICRO-TEMP LT unit. The internal components referenced in Figure (5-1) are as follows: A. Water Temperature Sensor B. Heater C.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 5-2. ACCESS TO THE INTERIOR OF THE MICRO-TEMP LT UNIT All internal operating components are readily accessible by separating the front and rear enclosure. NOTE: DRAIN THE RESERVOIR AND DISCONNECT THE POWER CORD FROM THE POWER SOURCE BEFORE REMOVING ANY PART FROM THE UNIT.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 5-2.3. DISCONNECTING THE CABLES FROM THE CONTROL BOARD WARNING • Always unplug the unit before accessing internal components during service. Failure to unplug the unit could result in electric shock. •...
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Where white ribbon cable attaches to display board, pull out the tabs on the side of the connector. Pull white ribbon cable out of connector. 5-3. REPLACEMENT OF THE HEATER WARNING •...
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Insert and lock the 2 white wires from the heater into the 2 open slots on the 6- position connector (position number 3 and 5). Insert and reposition reservoir back into rear enclosure. Tighten hose clamp on the connecting hose between the reservoir outlet and the pump inlet.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Disconnect elbow from pump outlet. Disconnect white and black pump lead wires. Unscrew all four 8-32” Phillips head pump mounting screws from base. Remove pump (with brackets and connecting hose between pump and reservoir still attached) from enclosure.
Place the black rubber skirt that was set aside in Step H, around the new reservoirs water fill opening, just underneath the threads. Refill the reservoir as described in Section (4-2.2). 5-6. REPLACEMENT OF THE TWIN BLACK TUBING Contact CSZ Technical Service. 5-7. REPLACEMENT OF THE WATER TEMPERATURE SENSOR (THERMISTOR) WARNING •...
Page 49
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Obtain replacement water temperature sensor. CAUTION • Always drain the MICRO-TEMP LT to a sanitary drain because bio-contaminants may be present in the unit’s water supply. Drain the reservoir as described in Section (4-2.1). Separate the front and rear enclosure as described in Section (5-2.1).
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 5-8. REPLACEMENT OF THE I/O POWER SWITCH WARNING • Always unplug the unit before accessing internal components during service. Failure to unplug the unit could result in electric shock. • For continued protection against risk of fire, replace only with same type and rating of fuse.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Insert replacement switch with attached cables into rear enclosure. Using 7mm nut driver, secure screws with two kepnuts on the inside of the rear enclosure. Reattach two white heater wires to red 6-position connector. Reconnect power switch cables to control board.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Close power switch tab. 5-10. REPLACEMENT OF THE WATER LEVEL SENSOR ASSEMBLY WARNING • Always unplug the unit before accessing internal components during service. Failure to unplug the unit could result in electric shock. Obtain a replacement water level sensor.
Page 53
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL WARNING • Always unplug the unit before accessing internal components during service. Failure to unplug the unit could result in electric shock. • Working with electronic boards, plugs, and cables requires delicate handling. Proper Electrostatic Discharge procedure should be followed during replacement of any electronic board.
Page 54
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Using 7 mm nut driver, remove plastic nut, nylon locknut, internal tooth washer and ground wire from display board. Pull out display board. Install new display board. Reinstall the left side of the board with internal tooth washer, ground wire, and nylon locknut.
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Reinstall the left side of the board with internal tooth washer, ground wire, and nylon locknut. Secure right side of board with plastic nut. Reinstall control board. Secure control board to front enclosure with seven 6-32” Phillips head screws. Note: Insert the short screw into the hole at the bottom right of the control board (near the J6 Position of the board).
Page 56
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL TROUBLESHOOTING GUIDE OBSERVATION POSSIBLE ACTION TO BE TAKEN PROBLEM Turn off unit. Replace/ Repair the water level Water level sensor fault. sensor. Refer to Section (5-9). Turn off unit. Replace Control board failure.
Page 57
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL TROUBLESHOOTING GUIDE OBSERVATION POSSIBLE ACTION TO BE TAKEN PROBLEM Blue ribbon cable Correct connection to the disconnected from control board. Refer to control board. Section (5-2.3). E. Press a button and nothing happens.
Page 58
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL TROUBLESHOOTING GUIDE OBSERVATION POSSIBLE ACTION TO BE TAKEN PROBLEM Temperature readout Unit has reached high limit. indicates 111˚F / 44˚C or Turn unit off and allow unit higher. to cool. Service required. Power failure.
Page 59
FIELD REPAIR/SERVICE MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL TROUBLESHOOTING GUIDE OBSERVATION POSSIBLE ACTION TO BE TAKEN PROBLEM Turn OFF unit and make repairs based on the Water level sensor fault. display message. Refer to Section (3-5). Power failure. Restore power to the unit.
Section (6-3) outlines the recommended replacement parts inventory. It is strongly recommended that all parts be replaced with parts purchased from CSZ or CSZ International GmbH. Use of other parts could void the warranty on the unit and possibly damage the unit.
PARTS INFORMATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL 6-4. RETURNING PARTS UNDER WARRANTY All parts are covered by a one year warranty. To replace parts during the warranty period, send the part prepaid to: Cincinnati Sub-Zero Products, Inc. 12011 Mosteller Road Cincinnati, Ohio 45241 Tel: (513) 772-8810...
PARTS INFORMATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL FIGURE 6-1. MICRO-TEMP LT INTERNAL EXPLODED - FRONT VIEW INDEX NUMBER I.D. NUMBER DESCRIPTION 78057 Membrane Control Panel LT (100/115V) 78058 Membrane Control Panel (230/240V) 91090 Pump (100V) 91091 Pump (115V) 91092 Pump (230/240V) 70817...
PARTS INFORMATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL FIGURE 6-3. MICRO-TEMP LT INTERNAL EXPLODED - REAR VIEW INDEX NUMBER I.D. NUMBER DESCRIPTION 28352 Display Board Ribbon Cable 78043 Front Enclosure 78032 Display Board 23026 Pump Mounting Bracket 78039 Control Board 65150 Power Switch...
PARTS INFORMATION MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL FIGURE 6-5. RESERVOIR – SIDE VIEW INDEX NUMBER I.D. NUMBER DESCRIPTION 62491 Reservoir Cap 62490 Reservoir 65149 Water Level Sensor 62492 Heater Plate Cap 91090 Heater (100V) 91091 Heater (115V) 91092 Heater (230/240V) 62494...
SPECIFICATIONS AND CERTIFICATIONS MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL SECTION 7. SPECIFICATIONS AND CERTIFICATIONS OF THE MICRO- TEMP LT MICRO-TEMP LT MODEL 749 FEATURES CLASSIFICATION SAFETY SYSTEM Equipment is Class I. Maximum High Control Setting: 42° C (107° F) Equipment is Type B.
Page 69
Water Temp. Display Range: WARRANTY Water Temp.: 20° C - 44° C (68° F – 111° F) 1 yr. parts (Parts and Labor if returned to CSZ) Temp. Settings: Water Temp.: 1° C (2° F) CERTIFICATIONS MODEL 749 MEDICAL ELECTRICAL...
Page 70
SPECIFICATIONS AND CERTIFICATIONS MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL The following tables are presented in fulfillment of the requirements of IEC 60601-1-2 Guidance and manufacturer’s declaration – electromagnetic emissions The Micro-Temp LT, Model 749 is intended for use in the electromagnetic environment specified below. The customer or the user of the Micro-Temp LT, Model 749 should assure that it is used in such an environment.
Page 71
SPECIFICATIONS AND CERTIFICATIONS MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Guidance and manufacturer’s declaration – electromagnetic immunity The Micro-Temp LT, Model 749 is intended for use in the electromagnetic environment specified below. The customer or the user of the Micro-Temp LT, Model 749 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level...
Page 72
SPECIFICATIONS AND CERTIFICATIONS MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Recommended separation distances between portable and mobile RF communications equipment and the Micro-Temp LT, Model 749 The Micro-Temp LT, Model 749 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Micro-Temp LT, Model 749 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Micro-Temp LT, Model 749 as recommended below, according to the maximum output power of the communications equipment.
Page 73
SPECIFICATIONS AND CERTIFICATIONS MICRO-TEMP LT, Model 749 OPERATION AND TECHNICAL MANUAL Cincinnati Sub-Zero Products, Inc. 12011 Mosteller Road Cincinnati, OH 45241 Telephone: (513) 772-8810 Fax: (513) 772-9119 E-Mail: cszinc@cszinc.com...