Apera 201DJ-C User Manual

Conductivity electrode

Advertisement

Brief Introduction

Apera Instruments 2301-C Conductivity Electrode is featured with a proprietary platinum black sensor
for accurate readings in a wide conductivity range up to 200 mS/cm.

Technical Specifications

Range
Housing
Constant
Connector
Cable length
Dimension
Temperature sensor
Operating temperature

How to Install the Electrode

1.
Find the conductivity BNC socket (where it shows Cond) on the conductivity meter; Open the
rubber cap; Insert the BNC connector of the electrode to the BNC socket while twisting clockwise
until it's locked.
2.
After connecting the new electrode to your meter, adjust the meter's constant K to 1.0 according
to the instruction of your meter's manual. Then perform a 2 to 3-point calibration to ensure the
accuracy.

How to Use the Electrode

1.
There is a proper amount of deionized water in the cap on top of the electrode. The conductivity
sensor is soaked in it to keep its sensitivity.
2.
Before measuring, loosen the bottle cap, then pull out the electrode while twisting
counterclockwise. Place the storage bottle at a safe position.
3.
Rinse the electrode with distilled or deionized water and blot-dry it with clean tissue or filter paper.
Rinse the electrode again with your test solution if possible for the best result.
4.
Insert the electrode into your sample solution and stir it for a few seconds in the solution to
remove potential air bubbles. Then wait for the stable reading and take the measurement.
2301-C Conductivity Electrode
User Manual
0 to 200 mS/cm
POM
K=1.0
BNC
3 Ft
ø12*155 mm
N/A
32 to 176°F (0 - 80°C)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 201DJ-C and is the answer not in the manual?

Questions and answers

Summary of Contents for Apera 201DJ-C

  • Page 1 2301-C Conductivity Electrode User Manual Brief Introduction Apera Instruments 2301-C Conductivity Electrode is featured with a proprietary platinum black sensor for accurate readings in a wide conductivity range up to 200 mS/cm. Technical Specifications Range 0 to 200 mS/cm Housing Constant K=1.0...
  • Page 2 We warrant this electrode to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS for a period of SIX MONTHS from the delivery.

Table of Contents