Introduction Dear Customer, Thank you for choosing Apera Instruments EC60-Z Smart Conductivity Tester. Please carefully read this manual before using the product in order to have a reliable testing experience. 1.1 This product is designed with a two-way control on both the tester and ZenTest Mobile App.
Battery Installation Please install batteries according to the following steps. *Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ① Pull the battery cap up ② Slide the battery cap along to the direction of arrow ③...
Complete Kit Things to Know Before Use 1) A few drops of destilliertes water are added to the probe cap to keep the conductivity electrode in an activated state before the tester leaves factory. Generally, users can start using the tester directly.
Page 6
and indication icon will appear at the bottom left of the LCD screen. 5) After calibration, place the probe in 12.88 mS/cm conductivity calibration solution. If the value is accurate, it is not necessary to conduct 2 point calibration. If it is inaccurate, follow the steps in 3) to 4) to complete the 2 point of calibration using 12.88 mS/cm calibration solution.
9) For the self-diagnosis information, please refer to the table below: Symbol Self-Diagnosis information How to fix 1. Make sure the probe is fully submerged in the solution. 2. Check if the standard solution is expired or contaminated. 3. Check if the conductivity electrode (two The meter cannot recognize the black rods) is damaged.
Parameter Setting Table of Settings Factory Symbol Parameter Setting Contents Content Default Temperature Unit ˚C – ˚F ˚F Select automatic lock 5-20 seconds – Off Automatic Backlight Off 1-8 minutes – Off Automatic Power Off 10-20 minutes – Off Conductivity Reference 15 ˚C to 30 ˚C 25 ˚C Temperature...
5) Conductivity Calibration Reminder (P07) – set X hours (H) Or X days (D) in ZenTest mobile app – settings – Parameter – pH – Calibration Reminder. On the meter, you can only check the values that’s been set up on ZenTest App. For example, if 3 days is set up, the Er6 icon (see Figure-4) will appear in the lower right corner of the LCD screen in 3 days to remind you to perform calibration, also in the ZenTest App there will be a pop-up reminder.
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.