Validating Calibration - 493K K-METRON User Instructions

Thickness measurement gauge (0-30mm) with radio link up and software recording
Hide thumbs Also See for K-METRON:
Table of Contents

Advertisement

4.3 Validating Calibration.

After calibration is complete the measured data should be validated. To do this position the
calibrated magnet once more on each step and ensure the K-METRON's measured and
displayed value is that of the actual step thickness (note the measurement tolerances as per
the table above apply).
10mm Cube and 10mm Sphere Validating Calibration
19mm Sphere Validating Calibration – use the provided +10mm disk to measure units
over 20mm.
10 |
P a g e
K-METRON Radio User Instructions (Firmware 3.0.4; Doc v4.0)
Magnet Used Mag Type Calibration Range
10mm Sphere Sphere
10mm Cube
Cube
19mm Sphere 10-30mm 10mm – 30mm
0mm - 20mm
0mm - 20mm
+10mm Calibration
Add-on