MODEL 5318B
Level Tek II, Capacitance Level Switch
"XXXXX" represents the current measurement.
If the low setpoint is to be adjust the press the
[DN] once. The display will show:
Low SP
XXXXX
"XXXXX" represents the current measurement.
4.6.2.

ADJUSTING THE SETPOINT

Once the desired setpoint is displayed, the
adjustment procedure is entered by pressing the
[RT] pushbutton once. For these instructions we
will assume the low setpoint is to be adjusted.
After a short period of time the bottom line of the
display will scroll the message:
"Press RT to adjust
Pressing the [RT] pushbutton will begin the
adjustment procedure. The display will show:
Low SP
DN 0.0%
After a short period of time the top line of the
display will scroll the following message:
"Press RT to save or LT to
At this point the [UP] and [DN] pushbuttons can
be used to determine the amount the setpoint is
to be changed. The amount of change will scroll
in increments of 0.1%.
The lower line of the display will change to show
the amount of change (YY X.X%). Where "YY"
is the direction of the change, "DN" or "UP" and
"X.X%" is the amount of change to be applied.
Pressing the [RT] pushbutton will apply the
change and the display will return to the display
of the adjusted setpoint value:
Low SP
XXXXX
"XXXXX" represents the adjusted setpoint value.
- Or -
Pressing the [LT] pushbutton will exit the
adjustment procedure without making any
setpoint"
quit"
Page 15
INSTRUCTION MANUAL
change and the display will return to the display
of the setpoint value:
Low SP
XXXXX
"XXXXX" represents the unadjusted setpoint
value.
4.6.3.
RETURNING TO THE NORMAL
DISPLAY MODE
Pressing the [LT] pushbutton twice will return the
display to the normal operating mode.
4.7.
RECOVERING THE FACTORY
DEFAULT SETTINGS
Under certain conditions it may be desirable to
restore the instrument to its factory default
settings.
Care should be taken when executing this
feature since it will erase all existing settings and
calibration.
This function can be found at the end of the
Setup Menu. It is accessed from the normal
mode by pressing the [DN] pushbutton once, the
[RT] pushbutton once and the [DN] pushbutton
eight times. The display will show:
CalibCLR
NO
After a short period of time the top line of the
display will scroll the message:
"Selecting RESET erase current settings
and
calibration"
Pressing the [RT] pushbutton will enter the
function. The display will show:
CalibCLR
Press RT
After a short period of time the bottom line of the
display will scroll the message:
"Press RT to erase calibration and set
default
setup"