Gait up GO User Manual page 27

Table of Contents

Advertisement

IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF immunity
IEC 61000-4-3
ENGLISH
N/A (DC supply)
30 A/m, 50 or 60 Hz
tested in X, Y and Z
axes.
10 V/m @
80MHz-2.7GHz
80% AM à 1kHz
30 A/m, 50 or 60 Hz
10 V/m @
80MHz-2.7GHz
80% AM à 1kHz
Power frequency
magnetic fields should
be at levels
characteristic of a
typical commercial or
hospital environment.
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Physilog6
sensor device,
including the USB
cable, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter
Minimum separation
distance shall be
calculated by following
equation:
E= (6/D)*sqrt(P)
Where:
● E is the
immunity test
level in V/m
● P is the
maximum
output power
rating of the
transmitter in
watts (W)
according to
the transmitter
76/84