Guidance And Manufacturer's Declaration - Electromagnetic Immunity; Electromagnetic Environment Guidance - Gait up GO User Manual

Table of Contents

Advertisement

RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ Flockers
emissions
IEC 61000-3-3
Table 11. Electromagnetic emissions

9.5. Guidance and manufacturer's declaration - Electromagnetic immunity

Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transients/burst
IEC 61000-4-4
Surges
ENGLISH
Class B
N/A
N/A
IEC 60601 test level
+/- 15 kV Air
+/- ± 2/4/8/15 kV
Contact
Supply: ± 2 kV
I/O lines : ± 1 kV
Frequency : 100 kHz
N/A (DC supply with cable < 3m)
very low and are not likely to cause any
interference in nearby electronic equipment.
Gait Up GO medical device is intended to be used
in home environment
N/A
N/A
Compliance level
+/- 15 kV Air
+/- ± 2/4/8/15 kV
Contact
Supply: ± 2 kV
I/O lines : ± 1 kV
Frequency : 100 kHz
Electromagnetic
environment
guidance
Floors should be
wood, concrete
or ceramic tile. If floors
are covered with
synthetic
material, the relative
humidity
should be at least 30%.
Note: Some ESD
events might cause the
unexpected reset of
the Physilog6 sensor
device. If after the
unexpected reset, the
device seems blocked,
please perform a
manual reset (see
section 6.1)
USB cable must be the
one provided by the
manufacturer
75/84