Page 3
USER MANUAL EARTH RESISTANCE METER MRU-11 SONEL S.A. Wokulskiego 11 58-100 Świdnica Version 1.05 22.07.2021...
Page 4
The MRU-11 meter is a modern, easy and safe measuring device. Please acquaint yourself with this manual in order to avoid measuring errors and prevent possible problems in operation of the meter. MRU-11 – USER MANUAL...
MRU-11 meter is designed to measure earth resistance values. Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the user.
ON and then OFF the screen backlight; To turn OFF the device, press and hold ON/OFF button for approx. 2 sec- onds. A screen indicating shut- ting off the device is shown briefly. MRU-11 – USER MANUAL...
Short press START button to accept the selected value. You will enter the screen with settings of audio messages - bEEP. Use UP and DOWN buttons to turn the audio messages ON ( ) or OFF ( MRU-11 – USER MANUAL...
Page 9
You will enter the screen with measuring voltage selection Un. Use UP and DOWN buttons to set the measuring voltage at 25 V or 50 V. The measuring voltage set relates to all measurement functions in the meter. MRU-11 – USER MANUAL...
Page 10
Long press (> 2 s) ON/OFF button to turn off the meter without accepting the changes introduced at the current setting position. Short press (> 2 s) ON/OFF button to enter the measurement screen R 3P without accepting the changes introduced at the current setting position. MRU-11 – USER MANUAL...
24 V but lower than 40 V, the xxV!, >24V, NOISE! and measurement is blocked. The interference voltage is lower than 24 V, but with a high NOISE! value - the measurement result may be affected by addi- tional uncertainty. MRU-11 – USER MANUAL...
→ meter → H current electrode → earth → tested earth electrode; • S electrode (voltage electrode) for measuring the voltage drop across the resistance of the meas- ured earthing as a result of current flow. Disconnect the tested earth electrode from the installation of the facility. MRU-11 – USER MANUAL...
Page 13
The earth electrode being tested and the current electrode and the voltage electrode should be located in one line. Press START. The progress of the measurement is indi- cated by horizontal lines on the screen. MRU-11 – USER MANUAL...
Page 14
24 V. Voltage of interferences is measured up to the level of 100 V, but above 40 V it is signalled as dangerous. The meter must not be connected to voltages exceeding 100 V. MRU-11 – USER MANUAL...
Page 15
Flashing edges of symbols: E or H or S, two or all three at Flashing edges the same time: disconnected one, two or three leads to the terminals, or the resistance of the auxiliary auxiliary elec- trode/s is outside the measuring range. MRU-11 – USER MANUAL...
Press DOWN button until the screen of R measurement appears. The meter is in the mode of measuring the in- terference voltage between the test terminals. The measuring voltage is compatible with the voltage selected when setting up the device. MRU-11 – USER MANUAL...
Page 17
R whereas the upper part of the screen shows additional results for R . The re- sult is displayed for 20 seconds. The result can be recalled by pressing UP button. MRU-11 – USER MANUAL...
Page 18
In majority of cases the achieved measurement ac- curacy is satisfactory. However, you should always be aware of the uncertainty in- cluded in the measurement. MRU-11 – USER MANUAL...
Page 19
Ω value Flashing symbols: Flashing symbols E or H or S, or both of them, or all three at the same time, one or two or three test leads are dis- connected from the measurement sockets. MRU-11 – USER MANUAL...
Test leads should be connected to the measurement terminals in the device, as shown above. In order to start the measurement, press START button. MRU-11 – USER MANUAL...
Page 21
24 V but lower than 40 V, the NOISE! and measurement is blocked. The interfering signal (noise signal) is below 24 V, but with NOISE! a high value - the measurement result may be affected by additional uncertainty. MRU-11 – USER MANUAL...
Connect the current electrode driven into ground to the E socket of the meter. Turn on the meter using ON/OFF but- ton. Set measurement unit and distance L between electrodes according to sec. 3 … steps Press DOWN button until the screen of ρ measurement appears. MRU-11 – USER MANUAL...
Page 23
(the Wenner’s method). If this is not the case the earthing resistance measurement must be carried out by means of the four-pole method and calculations must be performed individually. Pay particular attention to the quality of the connection of the tested object with the MRU-11 – USER MANUAL...
Page 24
Flashing symbols E or ES or H or S, or two of them, or Flashing symbols: three, or all of them at the same time – one or two or three or four test leads are disconnected from the measurement sockets. MRU-11 – USER MANUAL...
Message bAtt indicates that batter- ies/rechargeable batter- ies are totally discharged, all measurements are blocked. The meter switches off automatically after 5 sec. MRU-11 – USER MANUAL...
5.2 Replacing (rechargeable) batteries MRU-11 is powered by four AA alkaline LR6 batteries or rechargeable batteries of NiMH type. The (rechargeable) batteries are placed in the compartment at the bottom of the enclosure. The de- vice is not equipped with an internal battery charger. Rechargeable batteries must be recharged in an external charger.
If meter is to be stored for a prolonged period of time, the batteries must be removed from the device. In order to prevent a total discharge of the battery pack in the case of a prolonged storage, charge it from time to time. MRU-11 – USER MANUAL...
Measuring current at the short circuit of > 20 mA. Measuring frequency of 125 Hz or 150 Hz. Selected test voltage: 25 V or 50 V. Maximum interference voltage for the R measurement is 24 V. MRU-11 – USER MANUAL...
Page 29
(above sea level) ....................≤2000 m* x) quality standard ..........design and manufacturing are ISO 9001 compliant y) measurement method ............technical, conforming to EN 61557-5 z) the device meets the EMC requirements according to: ....EN 61326-1 and EN 61326-2-2 MRU-11 – USER MANUAL...
±0.2 digit/°C for R<1 kΩ Temperature ±0.07%/°C ±0.2 digit/°C for R≥1 kΩ According to the formulas shown Serial interference voltage in par. 9.2.1 (U =3V 50/60Hz) Resistance of earth contact According to the formula in probes par. 9.2.2 MRU-11 – USER MANUAL...
Cable on the reel, 50 m, yellow, for measuring earthing values WAPRZ050YEBBSZ (banana plugs) Cable on the reel, 100 m, red, for measuring WAPRZ100REBBSZ earthing values Cable on the reel, 200 m, yellow, for measur- WAPRZ200YEBBSZ ing earthing values MRU-11 – USER MANUAL...
The manufacturer of the device and provider of guarantee and post-guarantee service: SONEL S.A. Wokulskiego 11 58-100 Świdnica Poland tel. +48 74 858 38 60 fax +48 74 858 38 09 E-mail: export@sonel.pl Web page: www.sonel.pl NOTE! Service repairs must be performed only by the manufacturer. MRU-11 – USER MANUAL...
National Metrological Institute. According to ILAC-G24 „Guidelines for determination of calibration intervals of measuring instru- ments”, SONEL S.A. recommends periodical metrological inspection of the instruments it manufac- tures no less frequently than once every 12 months.
Need help?
Do you have a question about the MRU-11 and is the answer not in the manual?
Questions and answers