Page 1
PT2 E-1 4 9 3 OPERATION AND M AINTENANCE M ANUAL FOR RIKEN OX YGEN GAS DETECTOR HEAD GD-F3 A-1 Req uest for the Custom ers • Read and und erstand this op erating m anual b efore using the d etector head . •...
Page 2
[CAUTIONS ON OPERATION] This is a fixed typ e g as d etector (head ) and alarm instrum ent to d etect oxyg en g as in air. The g as alarm instrum ent is a safety instrum ent, not an analyzer nor m easuring instrum ent to m ake q uantitative and q ualitative analysis or m easurem ent. Please und erstand follow ing p oints and use this d etector head correctly. ...
TABLE OF CONTENTS 1 . PRODUCT OUTLINE 1 −1 . Preface ……………………………………………………………………… 3 1 −2 . Ap p lication for use ………………………………………………………… 3 1 −3 . Id entification of each sig nal w ord ………………………………………… 3 2 . IM PORTANT INSTRUCTIONS FOR THE SAFETY 2...
1 . PR O D U C T O U TLIN E 1−1.Pr e f a ce Thank you for p urchasing our fixed typ e Oxyg en Detector Head GD-F3 A-1 . First of all, verify that the m od el num b er of this p rod uct is id entical with sp ecifications d escrib ed in this m anual. This m anual is a g uid e to op eration of the GD-F3 A-1 . ...
3−2.N a m e o f e a ch p a r t a n d f u n ctio n s ⑧ ④ ⑥ ③ ⑦ ...
Page 9
! W A R N IN G The oxyg en d etector shall b e used as follow ing com p osition. N on-hazard ous H azard ous area area Li Lo Ci Co Lw...
Page 10
(2 ) Non-exp losion p roof system N on-hazard ous area N on-hazard ous area Detector head CVVS1 .2 5 sq ・ 2 C or eq uivalent (M AX 6 0 0 m ) ...
4 . H O W TO U SE 4 −1 . Before op eration All p ersons w ho use this d etector for the first tim e and w ho has even used the d etector are req uested to follow all the instructions for use carefully. If not, the instrum ent m ay b e d efective and correct g as d etection m ay not b e p erform ed . ...
6 . M A IN TEN A N C E A N D IN SPEC TIO N Gas d etection and alarm instrum ent is kep t in continuous op eration over a long tim e and m ust p erform a vital role as a safety d evice. For this p urp ose, p eriod ical insp ection m ust b e m ad e. The Hig h-p ressure Gas Safety Act in Jap an sets forth the ob lig ation of p eriod ical insp ection of ...
6 −2 . C a lib r a tio n Perform g as calib ration at new installation, 6 m onths insp ection and sensor rep lacem ent. 6 −2 −1 . R e q u ir e d k it a n d to o ls f o r ca lib r a tio n ...
7 . M EA SU R ES A T A BN O R M A L C A SE 7 −1 . R e s p o n d in g to tr o u b le a la r m ...
Page 20
9 . PR O D U C T SPEC IFIC A TIO N S 9 −1 . Sta n d a r d s p e cif ica tio n s M od el : GD-F3 A-1 Detection p rincip le ...
Page 21
1 0 . D ETEC TIO N PR IN C IPLE The GD-F3 A-1 em p loys a g alvanic cell as oxyg en d etection sensor that consists of nob le m etal and b ase m etal electrod es in an electrolyte. A m em b rane covers the cell and allow s g as in the sam p led g as to d iffuse into the electrolyte at a rate p rop ortional to the p artial p ressure of oxyg en. The oxyg en reacts in the ...
Need help?
Do you have a question about the GD-F3A-1 and is the answer not in the manual?